Cargando…

In-house validation of a liquid chromatography tandem mass spectrometry method for the analysis of lipophilic marine toxins in shellfish using matrix-matched calibration

A liquid chromatography tandem mass spectrometry (LC-MS/MS) method for the quantitative analysis of lipophilic marine toxins in shellfish extracts (mussel, oyster, cockle and clam) was validated in-house using European Union (EU) Commission Decision 2002/657/EC as a guideline. The validation include...

Descripción completa

Detalles Bibliográficos
Autores principales: Gerssen, Arjen, van Olst, Erik H. W., Mulder, Patrick P. J., de Boer, Jacob
Formato: Texto
Lenguaje:English
Publicado: Springer-Verlag 2010
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC2906727/
https://www.ncbi.nlm.nih.gov/pubmed/20552174
http://dx.doi.org/10.1007/s00216-010-3886-2
_version_ 1782184047254962176
author Gerssen, Arjen
van Olst, Erik H. W.
Mulder, Patrick P. J.
de Boer, Jacob
author_facet Gerssen, Arjen
van Olst, Erik H. W.
Mulder, Patrick P. J.
de Boer, Jacob
author_sort Gerssen, Arjen
collection PubMed
description A liquid chromatography tandem mass spectrometry (LC-MS/MS) method for the quantitative analysis of lipophilic marine toxins in shellfish extracts (mussel, oyster, cockle and clam) was validated in-house using European Union (EU) Commission Decision 2002/657/EC as a guideline. The validation included the toxins okadaic acid (OA), yessotoxin (YTX), azaspiracid-1 (AZA1), pectenotoxin-2 (PTX2) and 13-desmethyl spirolide-C (SPX1). Validation was performed at 0.5, 1 and 1.5 times the current EU permitted levels, which are 160 µg kg(-1) for OA, AZA1 and PTX2 and 1,000 µg kg(-1) for YTX. For SPX1, 400 µg kg(-1) was chosen as the target level as no legislation has been established yet for this compound. The method was validated for determination in crude methanolic shellfish extracts and for extracts purified by solid-phase extraction (SPE). Extracts were also subjected to hydrolysis conditions to determine the performance of the method for OA and dinophysistoxin esters. The toxins were quantified against a set of matrix-matched standards instead of standard solutions in methanol. To save valuable standard, methanolic extract instead of the homogenate was spiked with the toxin standard. This was justified by the fact that the extraction efficiency is high for all relevant toxins (above 90%). The method performed very well with respect to accuracy, intraday precision (repeatability), interday precision (within-laboratory reproducibility), linearity, decision limit, specificity and ruggedness. At the permitted level the accuracy ranged from 102 to 111%, the repeatability from 2.6 to 6.7% and the reproducibility from 4.7 to 14.2% in crude methanolic extracts. The crude extracts performed less satisfactorily with respect to the linearity (less than 0.990) and the change in LC-MS/MS sensitivity during the series (more than 25%). SPE purification resulted in greatly improved linearity and signal stability during the series. Recently the European Food Safety Authority (EFSA) has suggested that to not exceed the acute reference dose the levels should be below 45 µg kg(-1) OA equivalents and 30 µg kg(-1) AZA1 equivalents. A single-day validation was successfully conducted at these levels. If the regulatory levels are lowered towards the EFSA suggested values, the official methods prescribed in legislation (mouse and rat bioassay) will no longer be sensitive enough. The validated LC-MS/MS method presented has the potential to replace these animal tests. ELECTRONIC SUPPLEMENTARY MATERIAL: The online version of this article (doi:10.1007/s00216-010-3886-2) contains supplementary material, which is available to authorized users.
format Text
id pubmed-2906727
institution National Center for Biotechnology Information
language English
publishDate 2010
publisher Springer-Verlag
record_format MEDLINE/PubMed
spelling pubmed-29067272010-08-06 In-house validation of a liquid chromatography tandem mass spectrometry method for the analysis of lipophilic marine toxins in shellfish using matrix-matched calibration Gerssen, Arjen van Olst, Erik H. W. Mulder, Patrick P. J. de Boer, Jacob Anal Bioanal Chem Original Paper A liquid chromatography tandem mass spectrometry (LC-MS/MS) method for the quantitative analysis of lipophilic marine toxins in shellfish extracts (mussel, oyster, cockle and clam) was validated in-house using European Union (EU) Commission Decision 2002/657/EC as a guideline. The validation included the toxins okadaic acid (OA), yessotoxin (YTX), azaspiracid-1 (AZA1), pectenotoxin-2 (PTX2) and 13-desmethyl spirolide-C (SPX1). Validation was performed at 0.5, 1 and 1.5 times the current EU permitted levels, which are 160 µg kg(-1) for OA, AZA1 and PTX2 and 1,000 µg kg(-1) for YTX. For SPX1, 400 µg kg(-1) was chosen as the target level as no legislation has been established yet for this compound. The method was validated for determination in crude methanolic shellfish extracts and for extracts purified by solid-phase extraction (SPE). Extracts were also subjected to hydrolysis conditions to determine the performance of the method for OA and dinophysistoxin esters. The toxins were quantified against a set of matrix-matched standards instead of standard solutions in methanol. To save valuable standard, methanolic extract instead of the homogenate was spiked with the toxin standard. This was justified by the fact that the extraction efficiency is high for all relevant toxins (above 90%). The method performed very well with respect to accuracy, intraday precision (repeatability), interday precision (within-laboratory reproducibility), linearity, decision limit, specificity and ruggedness. At the permitted level the accuracy ranged from 102 to 111%, the repeatability from 2.6 to 6.7% and the reproducibility from 4.7 to 14.2% in crude methanolic extracts. The crude extracts performed less satisfactorily with respect to the linearity (less than 0.990) and the change in LC-MS/MS sensitivity during the series (more than 25%). SPE purification resulted in greatly improved linearity and signal stability during the series. Recently the European Food Safety Authority (EFSA) has suggested that to not exceed the acute reference dose the levels should be below 45 µg kg(-1) OA equivalents and 30 µg kg(-1) AZA1 equivalents. A single-day validation was successfully conducted at these levels. If the regulatory levels are lowered towards the EFSA suggested values, the official methods prescribed in legislation (mouse and rat bioassay) will no longer be sensitive enough. The validated LC-MS/MS method presented has the potential to replace these animal tests. ELECTRONIC SUPPLEMENTARY MATERIAL: The online version of this article (doi:10.1007/s00216-010-3886-2) contains supplementary material, which is available to authorized users. Springer-Verlag 2010-06-16 2010 /pmc/articles/PMC2906727/ /pubmed/20552174 http://dx.doi.org/10.1007/s00216-010-3886-2 Text en © The Author(s) 2010 https://creativecommons.org/licenses/by-nc/4.0/ This article is distributed under the terms of the Creative Commons Attribution Noncommercial License which permits any noncommercial use, distribution, and reproduction in any medium, provided the original author(s) and source are credited.
spellingShingle Original Paper
Gerssen, Arjen
van Olst, Erik H. W.
Mulder, Patrick P. J.
de Boer, Jacob
In-house validation of a liquid chromatography tandem mass spectrometry method for the analysis of lipophilic marine toxins in shellfish using matrix-matched calibration
title In-house validation of a liquid chromatography tandem mass spectrometry method for the analysis of lipophilic marine toxins in shellfish using matrix-matched calibration
title_full In-house validation of a liquid chromatography tandem mass spectrometry method for the analysis of lipophilic marine toxins in shellfish using matrix-matched calibration
title_fullStr In-house validation of a liquid chromatography tandem mass spectrometry method for the analysis of lipophilic marine toxins in shellfish using matrix-matched calibration
title_full_unstemmed In-house validation of a liquid chromatography tandem mass spectrometry method for the analysis of lipophilic marine toxins in shellfish using matrix-matched calibration
title_short In-house validation of a liquid chromatography tandem mass spectrometry method for the analysis of lipophilic marine toxins in shellfish using matrix-matched calibration
title_sort in-house validation of a liquid chromatography tandem mass spectrometry method for the analysis of lipophilic marine toxins in shellfish using matrix-matched calibration
topic Original Paper
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC2906727/
https://www.ncbi.nlm.nih.gov/pubmed/20552174
http://dx.doi.org/10.1007/s00216-010-3886-2
work_keys_str_mv AT gerssenarjen inhousevalidationofaliquidchromatographytandemmassspectrometrymethodfortheanalysisoflipophilicmarinetoxinsinshellfishusingmatrixmatchedcalibration
AT vanolsterikhw inhousevalidationofaliquidchromatographytandemmassspectrometrymethodfortheanalysisoflipophilicmarinetoxinsinshellfishusingmatrixmatchedcalibration
AT mulderpatrickpj inhousevalidationofaliquidchromatographytandemmassspectrometrymethodfortheanalysisoflipophilicmarinetoxinsinshellfishusingmatrixmatchedcalibration
AT deboerjacob inhousevalidationofaliquidchromatographytandemmassspectrometrymethodfortheanalysisoflipophilicmarinetoxinsinshellfishusingmatrixmatchedcalibration