Technical validation of an autoantibody test for lung cancer
Background: Publications on autoantibodies to tumour-associated antigens (TAAs) have failed to show either calibration or reproducibility data. The validation of a panel of six TAAs to which autoantibodies have been described is reported here. Materials and methods: Three separate groups of patients...
Autores principales: | , , , , , , , , , |
---|---|
Formato: | Texto |
Lenguaje: | English |
Publicado: |
Oxford University Press
2010
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC2911202/ https://www.ncbi.nlm.nih.gov/pubmed/20124350 http://dx.doi.org/10.1093/annonc/mdp606 |
Sumario: | Background: Publications on autoantibodies to tumour-associated antigens (TAAs) have failed to show either calibration or reproducibility data. The validation of a panel of six TAAs to which autoantibodies have been described is reported here. Materials and methods: Three separate groups of patients with newly diagnosed lung cancer were identified, along with control individuals, and their samples used to validate an enzyme-linked immunosorbant assay. Precision, linearity, assay reproducibility and antigen batch reproducibility were all assessed. Results: For between-replicate error, samples with higher signals gave coefficients of variation (CVs) in the range 7%–15%. CVs for between-plate variation were only 1%–2% higher. For between-run error, CVs were in the range 15%–28%. In linearity studies, the slope was close to 1.0 and correlation coefficient values were generally >0.8. The sensitivity and specificity of individual batches of antigen varied slightly between groups of patients; however, the sensitivity and specificity of the panel of antigens as a whole remained constant. The validity of the calibration system was demonstrated. Conclusions: A calibrated six-panel assay of TAAs has been validated for identifying nearly 40% of primary lung cancers via a peripheral blood test. Levels of reproducibility, precision and linearity would be acceptable for an assay used in a regulated clinical setting. |
---|