Implantable Cardioverter-Defibrillator Lead Extraction by Conventional Traction and Counter-Traction Technique

A 46-year-old man presented to our institution with inappropriate implantable cardioverter-defibrillator (ICD) shock delivery. The ICD (single chamber, dual shock coils) was implanted for sustained monomorphic ventricular tachycardia with unstable hemodynamics and underlying systolic left ventricula...

Descripción completa

Detalles Bibliográficos
Autores principales: Park, Jong Sung, Pak, Hui-Nam, Lee, Moon-Hyoung, Kim, Sung Soon, Joung, Boyoung
Formato: Texto
Lenguaje:English
Publicado: The Korean Society of Cardiology 2010
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC2933469/
https://www.ncbi.nlm.nih.gov/pubmed/20830258
http://dx.doi.org/10.4070/kcj.2010.40.8.418
Descripción
Sumario:A 46-year-old man presented to our institution with inappropriate implantable cardioverter-defibrillator (ICD) shock delivery. The ICD (single chamber, dual shock coils) was implanted for sustained monomorphic ventricular tachycardia with unstable hemodynamics and underlying systolic left ventricular dysfunction. ICD interrogation revealed recurrent episodes of ICD shock due to noise sensing and increased impedance of right ventricular (RV)-lead. With the impression of lead fracture, ICD lead extraction was performed. The fractured ICD lead was completely removed by traction of locking stylet and counter-traction of polypropylene dilator sheath. A new lead was inserted and the patient was discharged without complications after 2 days. To our knowledge, this is the first report on ICD lead extraction by conventional traction and counter-traction technique in Korea.