Optimized point dose measurement for monitor unit verification in intensity modulated radiation therapy using 6 MV photons by three different methodologies with different detector-phantom combinations: A comparative study

The study was aimed to compare accuracy of monitor unit verification in intensity modulated radiation therapy (IMRT) using 6 MV photons by three different methodologies with different detector phantom combinations. Sixty patients were randomly chosen. Zero degree couch and gantry angle plans were ge...

Descripción completa

Detalles Bibliográficos
Autores principales: Sarkar, Biplab, Ghosh, Bhaswar, Sriramprasath, Mahendramohan, Sukumaran, Basu, Ayan, Goswami, Jyotirup, Ray, Amitabh
Formato: Texto
Lenguaje:English
Publicado: Medknow Publications 2010
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC2936183/
https://www.ncbi.nlm.nih.gov/pubmed/20927221
http://dx.doi.org/10.4103/0971-6203.62129
_version_ 1782186454692134912
author Sarkar, Biplab
Ghosh, Bhaswar
Sriramprasath,
Mahendramohan, Sukumaran
Basu, Ayan
Goswami, Jyotirup
Ray, Amitabh
author_facet Sarkar, Biplab
Ghosh, Bhaswar
Sriramprasath,
Mahendramohan, Sukumaran
Basu, Ayan
Goswami, Jyotirup
Ray, Amitabh
author_sort Sarkar, Biplab
collection PubMed
description The study was aimed to compare accuracy of monitor unit verification in intensity modulated radiation therapy (IMRT) using 6 MV photons by three different methodologies with different detector phantom combinations. Sixty patients were randomly chosen. Zero degree couch and gantry angle plans were generated in a plastic universal IMRT verification phantom and 30×30×30 cc water phantom and measured using 0.125 cc and 0.6 cc chambers, respectively. Actual gantry and couch angle plans were also measured in water phantom using 0.6 cc chamber. A suitable point of measurement was chosen from the beam profile for each field. When the zero-degree gantry, couch angle plans and actual gantry, couch angle plans were measured by 0.6 cc chamber in water phantom, the percentage mean difference (MD) was 1.35%, 2.94 % and Standard Deviation (SD) was 2.99%, 5.22%, respectively. The plastic phantom measurements with 0.125 cc chamber Semiflex ionisation chamber (SIC) showed an MD=4.21% and SD=2.73 %, but when corrected for chamber-medium response, they showed an improvement, with MD=3.38 % and SD=2.59 %. It was found that measurements with water phantom and 0.6cc chamber at gantry angle zero degree showed better conformity than other measurements of medium-detector combinations. Correction in plastic phantom measurement improved the result only marginally, and actual gantry angle measurement in a flat- water phantom showed higher deviation.
format Text
id pubmed-2936183
institution National Center for Biotechnology Information
language English
publishDate 2010
publisher Medknow Publications
record_format MEDLINE/PubMed
spelling pubmed-29361832010-10-06 Optimized point dose measurement for monitor unit verification in intensity modulated radiation therapy using 6 MV photons by three different methodologies with different detector-phantom combinations: A comparative study Sarkar, Biplab Ghosh, Bhaswar Sriramprasath, Mahendramohan, Sukumaran Basu, Ayan Goswami, Jyotirup Ray, Amitabh J Med Phys Original Article The study was aimed to compare accuracy of monitor unit verification in intensity modulated radiation therapy (IMRT) using 6 MV photons by three different methodologies with different detector phantom combinations. Sixty patients were randomly chosen. Zero degree couch and gantry angle plans were generated in a plastic universal IMRT verification phantom and 30×30×30 cc water phantom and measured using 0.125 cc and 0.6 cc chambers, respectively. Actual gantry and couch angle plans were also measured in water phantom using 0.6 cc chamber. A suitable point of measurement was chosen from the beam profile for each field. When the zero-degree gantry, couch angle plans and actual gantry, couch angle plans were measured by 0.6 cc chamber in water phantom, the percentage mean difference (MD) was 1.35%, 2.94 % and Standard Deviation (SD) was 2.99%, 5.22%, respectively. The plastic phantom measurements with 0.125 cc chamber Semiflex ionisation chamber (SIC) showed an MD=4.21% and SD=2.73 %, but when corrected for chamber-medium response, they showed an improvement, with MD=3.38 % and SD=2.59 %. It was found that measurements with water phantom and 0.6cc chamber at gantry angle zero degree showed better conformity than other measurements of medium-detector combinations. Correction in plastic phantom measurement improved the result only marginally, and actual gantry angle measurement in a flat- water phantom showed higher deviation. Medknow Publications 2010 /pmc/articles/PMC2936183/ /pubmed/20927221 http://dx.doi.org/10.4103/0971-6203.62129 Text en © Journal of Medical Physics http://creativecommons.org/licenses/by/2.0/ This is an open-access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
spellingShingle Original Article
Sarkar, Biplab
Ghosh, Bhaswar
Sriramprasath,
Mahendramohan, Sukumaran
Basu, Ayan
Goswami, Jyotirup
Ray, Amitabh
Optimized point dose measurement for monitor unit verification in intensity modulated radiation therapy using 6 MV photons by three different methodologies with different detector-phantom combinations: A comparative study
title Optimized point dose measurement for monitor unit verification in intensity modulated radiation therapy using 6 MV photons by three different methodologies with different detector-phantom combinations: A comparative study
title_full Optimized point dose measurement for monitor unit verification in intensity modulated radiation therapy using 6 MV photons by three different methodologies with different detector-phantom combinations: A comparative study
title_fullStr Optimized point dose measurement for monitor unit verification in intensity modulated radiation therapy using 6 MV photons by three different methodologies with different detector-phantom combinations: A comparative study
title_full_unstemmed Optimized point dose measurement for monitor unit verification in intensity modulated radiation therapy using 6 MV photons by three different methodologies with different detector-phantom combinations: A comparative study
title_short Optimized point dose measurement for monitor unit verification in intensity modulated radiation therapy using 6 MV photons by three different methodologies with different detector-phantom combinations: A comparative study
title_sort optimized point dose measurement for monitor unit verification in intensity modulated radiation therapy using 6 mv photons by three different methodologies with different detector-phantom combinations: a comparative study
topic Original Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC2936183/
https://www.ncbi.nlm.nih.gov/pubmed/20927221
http://dx.doi.org/10.4103/0971-6203.62129
work_keys_str_mv AT sarkarbiplab optimizedpointdosemeasurementformonitorunitverificationinintensitymodulatedradiationtherapyusing6mvphotonsbythreedifferentmethodologieswithdifferentdetectorphantomcombinationsacomparativestudy
AT ghoshbhaswar optimizedpointdosemeasurementformonitorunitverificationinintensitymodulatedradiationtherapyusing6mvphotonsbythreedifferentmethodologieswithdifferentdetectorphantomcombinationsacomparativestudy
AT sriramprasath optimizedpointdosemeasurementformonitorunitverificationinintensitymodulatedradiationtherapyusing6mvphotonsbythreedifferentmethodologieswithdifferentdetectorphantomcombinationsacomparativestudy
AT mahendramohansukumaran optimizedpointdosemeasurementformonitorunitverificationinintensitymodulatedradiationtherapyusing6mvphotonsbythreedifferentmethodologieswithdifferentdetectorphantomcombinationsacomparativestudy
AT basuayan optimizedpointdosemeasurementformonitorunitverificationinintensitymodulatedradiationtherapyusing6mvphotonsbythreedifferentmethodologieswithdifferentdetectorphantomcombinationsacomparativestudy
AT goswamijyotirup optimizedpointdosemeasurementformonitorunitverificationinintensitymodulatedradiationtherapyusing6mvphotonsbythreedifferentmethodologieswithdifferentdetectorphantomcombinationsacomparativestudy
AT rayamitabh optimizedpointdosemeasurementformonitorunitverificationinintensitymodulatedradiationtherapyusing6mvphotonsbythreedifferentmethodologieswithdifferentdetectorphantomcombinationsacomparativestudy