Development of a Multicommutated Flow System with Chemiluminometric Detection for Quantification of Gentamicin in Pharmaceuticals

A new flow procedure based on multicommutation with chemiluminometric detection was developed to quantify gentamicin sulphate in pharmaceutical formulations. This approach is based on gentamicin's ability to inhibit the chemiluminometric reaction between luminol and hypochlorite in alkaline med...

Descripción completa

Detalles Bibliográficos
Autores principales: Santos, Lúcia H. M. L. M., Araújo, A. N., Reis, Boaventura, Montenegro, M. C. B. S. M.
Formato: Texto
Lenguaje:English
Publicado: Hindawi Publishing Corporation 2010
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC2957854/
https://www.ncbi.nlm.nih.gov/pubmed/20981343
http://dx.doi.org/10.1155/2010/913207
Descripción
Sumario:A new flow procedure based on multicommutation with chemiluminometric detection was developed to quantify gentamicin sulphate in pharmaceutical formulations. This approach is based on gentamicin's ability to inhibit the chemiluminometric reaction between luminol and hypochlorite in alkaline medium, causing a decrease in the analytical signal. The inhibition of the analytical signal is proportional to the concentration of gentamicin sulphate, within a linear range of 1 to 4 μg mL(−1) with a coefficient variation <3%. A sample throughput of 55 samples h(−1) was obtained. The developed method is sensitive, simple, with low reagent consumption, reproducible, and inexpensive, and when applied to the analysis of pharmaceutical formulations (eye drops and injections) it gave results with RSD between 1.10 and 4.40%.