Evaluation of performance, safety, subject acceptance, and compliance of a disposable autoinjector for subcutaneous injections in healthy volunteers

OBJECTIVE: A disposable autoinjector was developed for subcutaneous (SC) self-injection by patients with chronic diseases. To verify its performance and evaluate its acceptance, a clinical study was conducted in healthy volunteers, comparing SC injections performed by subjects using the autoinjector...

Descripción completa

Detalles Bibliográficos
Autores principales: Berteau, Cecile, Schwarzenbach, Florence, Donazzolo, Yves, Latreille, Mathilde, Berube, Julie, Abry, Herve, Cotten, Joël, Feger, Celine, Laurent, Philippe E
Formato: Texto
Lenguaje:English
Publicado: Dove Medical Press 2010
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC2962403/
https://www.ncbi.nlm.nih.gov/pubmed/21049090
_version_ 1782189191200768000
author Berteau, Cecile
Schwarzenbach, Florence
Donazzolo, Yves
Latreille, Mathilde
Berube, Julie
Abry, Herve
Cotten, Joël
Feger, Celine
Laurent, Philippe E
author_facet Berteau, Cecile
Schwarzenbach, Florence
Donazzolo, Yves
Latreille, Mathilde
Berube, Julie
Abry, Herve
Cotten, Joël
Feger, Celine
Laurent, Philippe E
author_sort Berteau, Cecile
collection PubMed
description OBJECTIVE: A disposable autoinjector was developed for subcutaneous (SC) self-injection by patients with chronic diseases. To verify its performance and evaluate its acceptance, a clinical study was conducted in healthy volunteers, comparing SC injections performed by subjects using the autoinjector with SC injections performed by nurses using a syringe. METHODS: This was a randomized, single-center, crossover study comparing SC self-injection using an autoinjector with SC nurse-administered injection using a syringe. Two volumes (0.2 mL and 1 mL) were injected into healthy volunteers. Study objectives included assessment of the accuracy and consistency of the volume injected by the injection systems, and skin reaction and pain associated with the injection. The fluid depot in the SC tissue layer was evaluated by ultrasound. Subject acceptance was evaluated using questionnaires on attitudes and emotions towards the injection technique, and challenged by seeking the subjects’ preferred system for a final study injection or future treatment. RESULTS: A total of 960 injections (480 with autoinjector, 480 with syringe) were performed in 40 subjects. There were no significant differences in mean fluid leakage and injected volumes between the systems. Pain associated with the injection was significantly lower with the auto-injector than with the syringe. Local skin reaction at the injection site was overall satisfactory. Injections were appropriately performed by all subjects. At study end, all 40 subjects preferred the autoinjector for a final study injection and for future treatment. CONCLUSION: This study indicated that the autoinjector used by the subject was similar to a syringe used by a nurse in terms of performance and safety in administering the injections, and better in terms of pain, overall acceptance, and preference.
format Text
id pubmed-2962403
institution National Center for Biotechnology Information
language English
publishDate 2010
publisher Dove Medical Press
record_format MEDLINE/PubMed
spelling pubmed-29624032010-11-03 Evaluation of performance, safety, subject acceptance, and compliance of a disposable autoinjector for subcutaneous injections in healthy volunteers Berteau, Cecile Schwarzenbach, Florence Donazzolo, Yves Latreille, Mathilde Berube, Julie Abry, Herve Cotten, Joël Feger, Celine Laurent, Philippe E Patient Prefer Adherence Original Research OBJECTIVE: A disposable autoinjector was developed for subcutaneous (SC) self-injection by patients with chronic diseases. To verify its performance and evaluate its acceptance, a clinical study was conducted in healthy volunteers, comparing SC injections performed by subjects using the autoinjector with SC injections performed by nurses using a syringe. METHODS: This was a randomized, single-center, crossover study comparing SC self-injection using an autoinjector with SC nurse-administered injection using a syringe. Two volumes (0.2 mL and 1 mL) were injected into healthy volunteers. Study objectives included assessment of the accuracy and consistency of the volume injected by the injection systems, and skin reaction and pain associated with the injection. The fluid depot in the SC tissue layer was evaluated by ultrasound. Subject acceptance was evaluated using questionnaires on attitudes and emotions towards the injection technique, and challenged by seeking the subjects’ preferred system for a final study injection or future treatment. RESULTS: A total of 960 injections (480 with autoinjector, 480 with syringe) were performed in 40 subjects. There were no significant differences in mean fluid leakage and injected volumes between the systems. Pain associated with the injection was significantly lower with the auto-injector than with the syringe. Local skin reaction at the injection site was overall satisfactory. Injections were appropriately performed by all subjects. At study end, all 40 subjects preferred the autoinjector for a final study injection and for future treatment. CONCLUSION: This study indicated that the autoinjector used by the subject was similar to a syringe used by a nurse in terms of performance and safety in administering the injections, and better in terms of pain, overall acceptance, and preference. Dove Medical Press 2010-10-05 /pmc/articles/PMC2962403/ /pubmed/21049090 Text en © 2010 Berteau et al, publisher and licensee Dove Medical Press Ltd. This is an Open Access article which permits unrestricted noncommercial use, provided the original work is properly cited.
spellingShingle Original Research
Berteau, Cecile
Schwarzenbach, Florence
Donazzolo, Yves
Latreille, Mathilde
Berube, Julie
Abry, Herve
Cotten, Joël
Feger, Celine
Laurent, Philippe E
Evaluation of performance, safety, subject acceptance, and compliance of a disposable autoinjector for subcutaneous injections in healthy volunteers
title Evaluation of performance, safety, subject acceptance, and compliance of a disposable autoinjector for subcutaneous injections in healthy volunteers
title_full Evaluation of performance, safety, subject acceptance, and compliance of a disposable autoinjector for subcutaneous injections in healthy volunteers
title_fullStr Evaluation of performance, safety, subject acceptance, and compliance of a disposable autoinjector for subcutaneous injections in healthy volunteers
title_full_unstemmed Evaluation of performance, safety, subject acceptance, and compliance of a disposable autoinjector for subcutaneous injections in healthy volunteers
title_short Evaluation of performance, safety, subject acceptance, and compliance of a disposable autoinjector for subcutaneous injections in healthy volunteers
title_sort evaluation of performance, safety, subject acceptance, and compliance of a disposable autoinjector for subcutaneous injections in healthy volunteers
topic Original Research
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC2962403/
https://www.ncbi.nlm.nih.gov/pubmed/21049090
work_keys_str_mv AT berteaucecile evaluationofperformancesafetysubjectacceptanceandcomplianceofadisposableautoinjectorforsubcutaneousinjectionsinhealthyvolunteers
AT schwarzenbachflorence evaluationofperformancesafetysubjectacceptanceandcomplianceofadisposableautoinjectorforsubcutaneousinjectionsinhealthyvolunteers
AT donazzoloyves evaluationofperformancesafetysubjectacceptanceandcomplianceofadisposableautoinjectorforsubcutaneousinjectionsinhealthyvolunteers
AT latreillemathilde evaluationofperformancesafetysubjectacceptanceandcomplianceofadisposableautoinjectorforsubcutaneousinjectionsinhealthyvolunteers
AT berubejulie evaluationofperformancesafetysubjectacceptanceandcomplianceofadisposableautoinjectorforsubcutaneousinjectionsinhealthyvolunteers
AT abryherve evaluationofperformancesafetysubjectacceptanceandcomplianceofadisposableautoinjectorforsubcutaneousinjectionsinhealthyvolunteers
AT cottenjoel evaluationofperformancesafetysubjectacceptanceandcomplianceofadisposableautoinjectorforsubcutaneousinjectionsinhealthyvolunteers
AT fegerceline evaluationofperformancesafetysubjectacceptanceandcomplianceofadisposableautoinjectorforsubcutaneousinjectionsinhealthyvolunteers
AT laurentphilippee evaluationofperformancesafetysubjectacceptanceandcomplianceofadisposableautoinjectorforsubcutaneousinjectionsinhealthyvolunteers