Evaluation of performance, safety, subject acceptance, and compliance of a disposable autoinjector for subcutaneous injections in healthy volunteers
OBJECTIVE: A disposable autoinjector was developed for subcutaneous (SC) self-injection by patients with chronic diseases. To verify its performance and evaluate its acceptance, a clinical study was conducted in healthy volunteers, comparing SC injections performed by subjects using the autoinjector...
Autores principales: | , , , , , , , , |
---|---|
Formato: | Texto |
Lenguaje: | English |
Publicado: |
Dove Medical Press
2010
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC2962403/ https://www.ncbi.nlm.nih.gov/pubmed/21049090 |
_version_ | 1782189191200768000 |
---|---|
author | Berteau, Cecile Schwarzenbach, Florence Donazzolo, Yves Latreille, Mathilde Berube, Julie Abry, Herve Cotten, Joël Feger, Celine Laurent, Philippe E |
author_facet | Berteau, Cecile Schwarzenbach, Florence Donazzolo, Yves Latreille, Mathilde Berube, Julie Abry, Herve Cotten, Joël Feger, Celine Laurent, Philippe E |
author_sort | Berteau, Cecile |
collection | PubMed |
description | OBJECTIVE: A disposable autoinjector was developed for subcutaneous (SC) self-injection by patients with chronic diseases. To verify its performance and evaluate its acceptance, a clinical study was conducted in healthy volunteers, comparing SC injections performed by subjects using the autoinjector with SC injections performed by nurses using a syringe. METHODS: This was a randomized, single-center, crossover study comparing SC self-injection using an autoinjector with SC nurse-administered injection using a syringe. Two volumes (0.2 mL and 1 mL) were injected into healthy volunteers. Study objectives included assessment of the accuracy and consistency of the volume injected by the injection systems, and skin reaction and pain associated with the injection. The fluid depot in the SC tissue layer was evaluated by ultrasound. Subject acceptance was evaluated using questionnaires on attitudes and emotions towards the injection technique, and challenged by seeking the subjects’ preferred system for a final study injection or future treatment. RESULTS: A total of 960 injections (480 with autoinjector, 480 with syringe) were performed in 40 subjects. There were no significant differences in mean fluid leakage and injected volumes between the systems. Pain associated with the injection was significantly lower with the auto-injector than with the syringe. Local skin reaction at the injection site was overall satisfactory. Injections were appropriately performed by all subjects. At study end, all 40 subjects preferred the autoinjector for a final study injection and for future treatment. CONCLUSION: This study indicated that the autoinjector used by the subject was similar to a syringe used by a nurse in terms of performance and safety in administering the injections, and better in terms of pain, overall acceptance, and preference. |
format | Text |
id | pubmed-2962403 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2010 |
publisher | Dove Medical Press |
record_format | MEDLINE/PubMed |
spelling | pubmed-29624032010-11-03 Evaluation of performance, safety, subject acceptance, and compliance of a disposable autoinjector for subcutaneous injections in healthy volunteers Berteau, Cecile Schwarzenbach, Florence Donazzolo, Yves Latreille, Mathilde Berube, Julie Abry, Herve Cotten, Joël Feger, Celine Laurent, Philippe E Patient Prefer Adherence Original Research OBJECTIVE: A disposable autoinjector was developed for subcutaneous (SC) self-injection by patients with chronic diseases. To verify its performance and evaluate its acceptance, a clinical study was conducted in healthy volunteers, comparing SC injections performed by subjects using the autoinjector with SC injections performed by nurses using a syringe. METHODS: This was a randomized, single-center, crossover study comparing SC self-injection using an autoinjector with SC nurse-administered injection using a syringe. Two volumes (0.2 mL and 1 mL) were injected into healthy volunteers. Study objectives included assessment of the accuracy and consistency of the volume injected by the injection systems, and skin reaction and pain associated with the injection. The fluid depot in the SC tissue layer was evaluated by ultrasound. Subject acceptance was evaluated using questionnaires on attitudes and emotions towards the injection technique, and challenged by seeking the subjects’ preferred system for a final study injection or future treatment. RESULTS: A total of 960 injections (480 with autoinjector, 480 with syringe) were performed in 40 subjects. There were no significant differences in mean fluid leakage and injected volumes between the systems. Pain associated with the injection was significantly lower with the auto-injector than with the syringe. Local skin reaction at the injection site was overall satisfactory. Injections were appropriately performed by all subjects. At study end, all 40 subjects preferred the autoinjector for a final study injection and for future treatment. CONCLUSION: This study indicated that the autoinjector used by the subject was similar to a syringe used by a nurse in terms of performance and safety in administering the injections, and better in terms of pain, overall acceptance, and preference. Dove Medical Press 2010-10-05 /pmc/articles/PMC2962403/ /pubmed/21049090 Text en © 2010 Berteau et al, publisher and licensee Dove Medical Press Ltd. This is an Open Access article which permits unrestricted noncommercial use, provided the original work is properly cited. |
spellingShingle | Original Research Berteau, Cecile Schwarzenbach, Florence Donazzolo, Yves Latreille, Mathilde Berube, Julie Abry, Herve Cotten, Joël Feger, Celine Laurent, Philippe E Evaluation of performance, safety, subject acceptance, and compliance of a disposable autoinjector for subcutaneous injections in healthy volunteers |
title | Evaluation of performance, safety, subject acceptance, and compliance of a disposable autoinjector for subcutaneous injections in healthy volunteers |
title_full | Evaluation of performance, safety, subject acceptance, and compliance of a disposable autoinjector for subcutaneous injections in healthy volunteers |
title_fullStr | Evaluation of performance, safety, subject acceptance, and compliance of a disposable autoinjector for subcutaneous injections in healthy volunteers |
title_full_unstemmed | Evaluation of performance, safety, subject acceptance, and compliance of a disposable autoinjector for subcutaneous injections in healthy volunteers |
title_short | Evaluation of performance, safety, subject acceptance, and compliance of a disposable autoinjector for subcutaneous injections in healthy volunteers |
title_sort | evaluation of performance, safety, subject acceptance, and compliance of a disposable autoinjector for subcutaneous injections in healthy volunteers |
topic | Original Research |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC2962403/ https://www.ncbi.nlm.nih.gov/pubmed/21049090 |
work_keys_str_mv | AT berteaucecile evaluationofperformancesafetysubjectacceptanceandcomplianceofadisposableautoinjectorforsubcutaneousinjectionsinhealthyvolunteers AT schwarzenbachflorence evaluationofperformancesafetysubjectacceptanceandcomplianceofadisposableautoinjectorforsubcutaneousinjectionsinhealthyvolunteers AT donazzoloyves evaluationofperformancesafetysubjectacceptanceandcomplianceofadisposableautoinjectorforsubcutaneousinjectionsinhealthyvolunteers AT latreillemathilde evaluationofperformancesafetysubjectacceptanceandcomplianceofadisposableautoinjectorforsubcutaneousinjectionsinhealthyvolunteers AT berubejulie evaluationofperformancesafetysubjectacceptanceandcomplianceofadisposableautoinjectorforsubcutaneousinjectionsinhealthyvolunteers AT abryherve evaluationofperformancesafetysubjectacceptanceandcomplianceofadisposableautoinjectorforsubcutaneousinjectionsinhealthyvolunteers AT cottenjoel evaluationofperformancesafetysubjectacceptanceandcomplianceofadisposableautoinjectorforsubcutaneousinjectionsinhealthyvolunteers AT fegerceline evaluationofperformancesafetysubjectacceptanceandcomplianceofadisposableautoinjectorforsubcutaneousinjectionsinhealthyvolunteers AT laurentphilippee evaluationofperformancesafetysubjectacceptanceandcomplianceofadisposableautoinjectorforsubcutaneousinjectionsinhealthyvolunteers |