The impact of traditional and smart pump infusion technology on nurse medication administration performance in a simulated inpatient unit

OBJECTIVE: Assess the impact of infusion pump technologies (traditional pump vs smart pump vs smart pump with barcode) on nurses' ability to safely administer intravenous medications. DESIGN: Experimental study with a repeated measures design. SETTING: High-fidelity simulated inpatient unit. RE...

Descripción completa

Detalles Bibliográficos
Autores principales: Trbovich, P L, Pinkney, S, Cafazzo, J A, Easty, A C
Formato: Texto
Lenguaje:English
Publicado: BMJ Group 2010
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC2975961/
https://www.ncbi.nlm.nih.gov/pubmed/20427310
http://dx.doi.org/10.1136/qshc.2009.032839
_version_ 1782190969448300544
author Trbovich, P L
Pinkney, S
Cafazzo, J A
Easty, A C
author_facet Trbovich, P L
Pinkney, S
Cafazzo, J A
Easty, A C
author_sort Trbovich, P L
collection PubMed
description OBJECTIVE: Assess the impact of infusion pump technologies (traditional pump vs smart pump vs smart pump with barcode) on nurses' ability to safely administer intravenous medications. DESIGN: Experimental study with a repeated measures design. SETTING: High-fidelity simulated inpatient unit. RESULTS: The nurses remedied 60% of “wrong drug” errors. This rate did not vary as a function of pump type. The nurses remedied “wrong patient” errors more often when using the barcode pump (88%) than when using the traditional pump (46%) or the smart pump (58%) (Cochran Q=14.36; p<0.05). The number of nurses who remedied “wrong dose hard limit” errors was higher when using the smart pump (75%) and the barcode pump (79%) than when using the traditional pump (38%) (Cochran Q=12.13; p<0.003). Conversely, there was no difference in remediation of “wrong dose soft limit” errors across pump types. The nurses' pump programming was less accurate when mathematical conversions were required. Success rates on secondary infusions were low (55.6%) and did not vary as a function of pump type. CONCLUSIONS: These findings indicate that soft (changeable) limits in smart infusion pumps had no significant effect in preventing dosing errors. Provided that smart pumps are programmed with hard (unchangeable) limits, they can prevent dosing errors, thereby increasing patient safety. Until barcode pumps are integrated with other systems within the medication administration process, their role in enhancing patient safety will be limited. Further improvements to pump technologies are needed to mitigate risks associated with intravenous infusions, particularly secondary infusions.
format Text
id pubmed-2975961
institution National Center for Biotechnology Information
language English
publishDate 2010
publisher BMJ Group
record_format MEDLINE/PubMed
spelling pubmed-29759612010-11-26 The impact of traditional and smart pump infusion technology on nurse medication administration performance in a simulated inpatient unit Trbovich, P L Pinkney, S Cafazzo, J A Easty, A C Qual Saf Health Care Education and Training OBJECTIVE: Assess the impact of infusion pump technologies (traditional pump vs smart pump vs smart pump with barcode) on nurses' ability to safely administer intravenous medications. DESIGN: Experimental study with a repeated measures design. SETTING: High-fidelity simulated inpatient unit. RESULTS: The nurses remedied 60% of “wrong drug” errors. This rate did not vary as a function of pump type. The nurses remedied “wrong patient” errors more often when using the barcode pump (88%) than when using the traditional pump (46%) or the smart pump (58%) (Cochran Q=14.36; p<0.05). The number of nurses who remedied “wrong dose hard limit” errors was higher when using the smart pump (75%) and the barcode pump (79%) than when using the traditional pump (38%) (Cochran Q=12.13; p<0.003). Conversely, there was no difference in remediation of “wrong dose soft limit” errors across pump types. The nurses' pump programming was less accurate when mathematical conversions were required. Success rates on secondary infusions were low (55.6%) and did not vary as a function of pump type. CONCLUSIONS: These findings indicate that soft (changeable) limits in smart infusion pumps had no significant effect in preventing dosing errors. Provided that smart pumps are programmed with hard (unchangeable) limits, they can prevent dosing errors, thereby increasing patient safety. Until barcode pumps are integrated with other systems within the medication administration process, their role in enhancing patient safety will be limited. Further improvements to pump technologies are needed to mitigate risks associated with intravenous infusions, particularly secondary infusions. BMJ Group 2010-04-27 2010-10 /pmc/articles/PMC2975961/ /pubmed/20427310 http://dx.doi.org/10.1136/qshc.2009.032839 Text en © 2010, Published by the BMJ Publishing Group Limited. For permission to use (where not already granted under a licence) please go to http://group.bmj.com/group/rights-licensing/permissions. This is an open-access article distributed under the terms of the Creative Commons Attribution Non-commercial License, which permits use, distribution, and reproduction in any medium, provided the original work is properly cited, the use is non commercial and is otherwise in compliance with the license. See: http://creativecommons.org/licenses/by-nc/2.0/ and http://creativecommons.org/licenses/by-nc/2.0/legalcode.
spellingShingle Education and Training
Trbovich, P L
Pinkney, S
Cafazzo, J A
Easty, A C
The impact of traditional and smart pump infusion technology on nurse medication administration performance in a simulated inpatient unit
title The impact of traditional and smart pump infusion technology on nurse medication administration performance in a simulated inpatient unit
title_full The impact of traditional and smart pump infusion technology on nurse medication administration performance in a simulated inpatient unit
title_fullStr The impact of traditional and smart pump infusion technology on nurse medication administration performance in a simulated inpatient unit
title_full_unstemmed The impact of traditional and smart pump infusion technology on nurse medication administration performance in a simulated inpatient unit
title_short The impact of traditional and smart pump infusion technology on nurse medication administration performance in a simulated inpatient unit
title_sort impact of traditional and smart pump infusion technology on nurse medication administration performance in a simulated inpatient unit
topic Education and Training
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC2975961/
https://www.ncbi.nlm.nih.gov/pubmed/20427310
http://dx.doi.org/10.1136/qshc.2009.032839
work_keys_str_mv AT trbovichpl theimpactoftraditionalandsmartpumpinfusiontechnologyonnursemedicationadministrationperformanceinasimulatedinpatientunit
AT pinkneys theimpactoftraditionalandsmartpumpinfusiontechnologyonnursemedicationadministrationperformanceinasimulatedinpatientunit
AT cafazzoja theimpactoftraditionalandsmartpumpinfusiontechnologyonnursemedicationadministrationperformanceinasimulatedinpatientunit
AT eastyac theimpactoftraditionalandsmartpumpinfusiontechnologyonnursemedicationadministrationperformanceinasimulatedinpatientunit
AT trbovichpl impactoftraditionalandsmartpumpinfusiontechnologyonnursemedicationadministrationperformanceinasimulatedinpatientunit
AT pinkneys impactoftraditionalandsmartpumpinfusiontechnologyonnursemedicationadministrationperformanceinasimulatedinpatientunit
AT cafazzoja impactoftraditionalandsmartpumpinfusiontechnologyonnursemedicationadministrationperformanceinasimulatedinpatientunit
AT eastyac impactoftraditionalandsmartpumpinfusiontechnologyonnursemedicationadministrationperformanceinasimulatedinpatientunit