Comparison of temperature rise in the pulp chamber with different light curing units: An in-vitro study
AIMS/OBJECTIVES: This in vitro study was designed to measure and compare the temperature rise in the pulp chamber with different light curing units. MATERIALS AND METHODS: The study was done in two settings-in-vitro and in-vivo simulation. In in-vitro setting, 3mm and 6mm acrylic spacers with 4mm ti...
Autores principales: | , , , , |
---|---|
Formato: | Texto |
Lenguaje: | English |
Publicado: |
Medknow Publications
2010
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC2980608/ https://www.ncbi.nlm.nih.gov/pubmed/21116387 http://dx.doi.org/10.4103/0972-0707.71644 |
_version_ | 1782191631975317504 |
---|---|
author | Rajesh Ebenezar, A V Anilkumar, R Indira, R Ramachandran, S Srinivasan, M R |
author_facet | Rajesh Ebenezar, A V Anilkumar, R Indira, R Ramachandran, S Srinivasan, M R |
author_sort | Rajesh Ebenezar, A V |
collection | PubMed |
description | AIMS/OBJECTIVES: This in vitro study was designed to measure and compare the temperature rise in the pulp chamber with different light curing units. MATERIALS AND METHODS: The study was done in two settings-in-vitro and in-vivo simulation. In in-vitro setting, 3mm and 6mm acrylic spacers with 4mm tip diameter thermocouple was used and six groups were formed according to the light curing source- 3 Quartz-Tungsten-Halogen (QTH) units and 3 Light-Emitting-Diode (LED) units. For the LED units, three modes of curing like pulse-cure mode, fast mode and ramp mode were used. For in-vivo simulation, 12 caries free human third molar tooth with fused root were used. K-type thermocouple with 1 mm tip diameter was used. Occlusal cavity was prepared, etched, rinsed with water and blot dried; bonding agent was applied and incremental curing of composite was done. Thermal emission for each light curing agent was noted. RESULTS: Temperature rise was very minimal in LED light cure units than in QTH light cure units in both the settings. Temperature rise was minimal at 6mm distance when compared to 3 mm distance. Among the various modes, fast mode produces the less temperature rise. Temperature rise in all the light curing units was well within the normal range of pulpal physiology. CONCLUSION: Temperature rise caused due to light curing units does not result in irreversible pulpal damage. |
format | Text |
id | pubmed-2980608 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2010 |
publisher | Medknow Publications |
record_format | MEDLINE/PubMed |
spelling | pubmed-29806082010-11-29 Comparison of temperature rise in the pulp chamber with different light curing units: An in-vitro study Rajesh Ebenezar, A V Anilkumar, R Indira, R Ramachandran, S Srinivasan, M R J Conserv Dent Original Article AIMS/OBJECTIVES: This in vitro study was designed to measure and compare the temperature rise in the pulp chamber with different light curing units. MATERIALS AND METHODS: The study was done in two settings-in-vitro and in-vivo simulation. In in-vitro setting, 3mm and 6mm acrylic spacers with 4mm tip diameter thermocouple was used and six groups were formed according to the light curing source- 3 Quartz-Tungsten-Halogen (QTH) units and 3 Light-Emitting-Diode (LED) units. For the LED units, three modes of curing like pulse-cure mode, fast mode and ramp mode were used. For in-vivo simulation, 12 caries free human third molar tooth with fused root were used. K-type thermocouple with 1 mm tip diameter was used. Occlusal cavity was prepared, etched, rinsed with water and blot dried; bonding agent was applied and incremental curing of composite was done. Thermal emission for each light curing agent was noted. RESULTS: Temperature rise was very minimal in LED light cure units than in QTH light cure units in both the settings. Temperature rise was minimal at 6mm distance when compared to 3 mm distance. Among the various modes, fast mode produces the less temperature rise. Temperature rise in all the light curing units was well within the normal range of pulpal physiology. CONCLUSION: Temperature rise caused due to light curing units does not result in irreversible pulpal damage. Medknow Publications 2010 /pmc/articles/PMC2980608/ /pubmed/21116387 http://dx.doi.org/10.4103/0972-0707.71644 Text en © Journal of Conservative Dentistry http://creativecommons.org/licenses/by/2.0/ This is an open-access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited. |
spellingShingle | Original Article Rajesh Ebenezar, A V Anilkumar, R Indira, R Ramachandran, S Srinivasan, M R Comparison of temperature rise in the pulp chamber with different light curing units: An in-vitro study |
title | Comparison of temperature rise in the pulp chamber with different light curing units: An in-vitro study |
title_full | Comparison of temperature rise in the pulp chamber with different light curing units: An in-vitro study |
title_fullStr | Comparison of temperature rise in the pulp chamber with different light curing units: An in-vitro study |
title_full_unstemmed | Comparison of temperature rise in the pulp chamber with different light curing units: An in-vitro study |
title_short | Comparison of temperature rise in the pulp chamber with different light curing units: An in-vitro study |
title_sort | comparison of temperature rise in the pulp chamber with different light curing units: an in-vitro study |
topic | Original Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC2980608/ https://www.ncbi.nlm.nih.gov/pubmed/21116387 http://dx.doi.org/10.4103/0972-0707.71644 |
work_keys_str_mv | AT rajeshebenezarav comparisonoftemperatureriseinthepulpchamberwithdifferentlightcuringunitsaninvitrostudy AT anilkumarr comparisonoftemperatureriseinthepulpchamberwithdifferentlightcuringunitsaninvitrostudy AT indirar comparisonoftemperatureriseinthepulpchamberwithdifferentlightcuringunitsaninvitrostudy AT ramachandrans comparisonoftemperatureriseinthepulpchamberwithdifferentlightcuringunitsaninvitrostudy AT srinivasanmr comparisonoftemperatureriseinthepulpchamberwithdifferentlightcuringunitsaninvitrostudy |