Comparison of CDE data in phacoemulsification between an open hospital-based ambulatory surgical center and a free-standing ambulatory surgical center

Mean CDE (cumulative dissipated energy) values were compared for an open hospital- based surgical center and a free-standing surgical center. The same model of phacoemulsifier (Alcon Infiniti Ozil) was used. Mean CDE values showed that surgeons (individual private practice) at the free-standing surg...

Descripción completa

Detalles Bibliográficos
Autores principales: Chen, Ming, Chen, Mindy
Formato: Texto
Lenguaje:English
Publicado: Dove Medical Press 2010
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC2993129/
https://www.ncbi.nlm.nih.gov/pubmed/21151334
http://dx.doi.org/10.2147/OPTH.S15076
Descripción
Sumario:Mean CDE (cumulative dissipated energy) values were compared for an open hospital- based surgical center and a free-standing surgical center. The same model of phacoemulsifier (Alcon Infiniti Ozil) was used. Mean CDE values showed that surgeons (individual private practice) at the free-standing surgical center were more efficient than surgeons (individual private practice) at the open hospital-based surgical center (mean CDE at the hospital-based surgical center 18.96 seconds [SD = 12.51]; mean CDE at the free-standing surgical center 13.2 seconds [SD = 9.5]). CDE can be used to monitor the efficiency of a cataract surgeon and surgical center in phacoemulsification. The CDE value may be used by institutions as one of the indicators for quality control and audit in phacoemulsification.