Evaluation methods for detecting changes in beam output and energy in radiation beams from high-energy linear accelerators

There is need for simple methods for checking consistency of beam outputs and energy in linear accelerators used for radiotherapy. A method was designed by the department using perspex phantom with which the dosimetric data of two medical linear accelerators (Clinac 600 CD, Clinac 2300 CD) were eval...

Descripción completa

Detalles Bibliográficos
Autores principales: Ravichandran, R., Binukumar, J. P., Davis, C. A., Krishnamurthy, K., Sivakumar, S. S.
Formato: Texto
Lenguaje:English
Publicado: Medknow Publications 2007
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3000505/
https://www.ncbi.nlm.nih.gov/pubmed/21157527
http://dx.doi.org/10.4103/0971-6203.35720
_version_ 1782193549697089536
author Ravichandran, R.
Binukumar, J. P.
Davis, C. A.
Krishnamurthy, K.
Sivakumar, S. S.
author_facet Ravichandran, R.
Binukumar, J. P.
Davis, C. A.
Krishnamurthy, K.
Sivakumar, S. S.
author_sort Ravichandran, R.
collection PubMed
description There is need for simple methods for checking consistency of beam outputs and energy in linear accelerators used for radiotherapy. A method was designed by the department using perspex phantom with which the dosimetric data of two medical linear accelerators (Clinac 600 CD, Clinac 2300 CD) were evaluated over a period of 30 months. The efficacy of methods followed was checked. Routine beam consistency checks were designed for photon beams with 15 cm/ 5 cm depth ionizations in perspex phantom and variable depth combinations for electron beams. Calculated ionization ratios were compared with measured values to show their significance. The dose/MU for all radiation beams was maintained within 2% accuracy over the period of 30 months. Clinac 600 CD machine showed decreasing trend of cGy/MU, while Clinac 2300 CD showed increasing trend of cGy/MU over a period, which needed tuning of monitor chamber two times each. Tuning of output to achieve standard value was carried out once, for all electron energies when the output dose/MU exceeded 3%. During one week (June 2005), there were slight changes in electron energy detected using the ratio method, which did not recur anytime afterwards. The methods designed are adequate to find the consistency in the beam output and energies in the radiotherapy linacs.
format Text
id pubmed-3000505
institution National Center for Biotechnology Information
language English
publishDate 2007
publisher Medknow Publications
record_format MEDLINE/PubMed
spelling pubmed-30005052010-12-14 Evaluation methods for detecting changes in beam output and energy in radiation beams from high-energy linear accelerators Ravichandran, R. Binukumar, J. P. Davis, C. A. Krishnamurthy, K. Sivakumar, S. S. J Med Phys Original Article There is need for simple methods for checking consistency of beam outputs and energy in linear accelerators used for radiotherapy. A method was designed by the department using perspex phantom with which the dosimetric data of two medical linear accelerators (Clinac 600 CD, Clinac 2300 CD) were evaluated over a period of 30 months. The efficacy of methods followed was checked. Routine beam consistency checks were designed for photon beams with 15 cm/ 5 cm depth ionizations in perspex phantom and variable depth combinations for electron beams. Calculated ionization ratios were compared with measured values to show their significance. The dose/MU for all radiation beams was maintained within 2% accuracy over the period of 30 months. Clinac 600 CD machine showed decreasing trend of cGy/MU, while Clinac 2300 CD showed increasing trend of cGy/MU over a period, which needed tuning of monitor chamber two times each. Tuning of output to achieve standard value was carried out once, for all electron energies when the output dose/MU exceeded 3%. During one week (June 2005), there were slight changes in electron energy detected using the ratio method, which did not recur anytime afterwards. The methods designed are adequate to find the consistency in the beam output and energies in the radiotherapy linacs. Medknow Publications 2007 /pmc/articles/PMC3000505/ /pubmed/21157527 http://dx.doi.org/10.4103/0971-6203.35720 Text en © Journal of Medical Physics http://creativecommons.org/licenses/by/2.0/ This is an open-access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
spellingShingle Original Article
Ravichandran, R.
Binukumar, J. P.
Davis, C. A.
Krishnamurthy, K.
Sivakumar, S. S.
Evaluation methods for detecting changes in beam output and energy in radiation beams from high-energy linear accelerators
title Evaluation methods for detecting changes in beam output and energy in radiation beams from high-energy linear accelerators
title_full Evaluation methods for detecting changes in beam output and energy in radiation beams from high-energy linear accelerators
title_fullStr Evaluation methods for detecting changes in beam output and energy in radiation beams from high-energy linear accelerators
title_full_unstemmed Evaluation methods for detecting changes in beam output and energy in radiation beams from high-energy linear accelerators
title_short Evaluation methods for detecting changes in beam output and energy in radiation beams from high-energy linear accelerators
title_sort evaluation methods for detecting changes in beam output and energy in radiation beams from high-energy linear accelerators
topic Original Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3000505/
https://www.ncbi.nlm.nih.gov/pubmed/21157527
http://dx.doi.org/10.4103/0971-6203.35720
work_keys_str_mv AT ravichandranr evaluationmethodsfordetectingchangesinbeamoutputandenergyinradiationbeamsfromhighenergylinearaccelerators
AT binukumarjp evaluationmethodsfordetectingchangesinbeamoutputandenergyinradiationbeamsfromhighenergylinearaccelerators
AT davisca evaluationmethodsfordetectingchangesinbeamoutputandenergyinradiationbeamsfromhighenergylinearaccelerators
AT krishnamurthyk evaluationmethodsfordetectingchangesinbeamoutputandenergyinradiationbeamsfromhighenergylinearaccelerators
AT sivakumarss evaluationmethodsfordetectingchangesinbeamoutputandenergyinradiationbeamsfromhighenergylinearaccelerators