Electronic tissue compensation achieved with both dynamic and static multileaf collimator in eclipse treatment planning system for Clinac 6 EX and 2100 CD Varian linear accelerators: Feasibility and dosimetric study

Dynamic multileaf collimator (DMLC) and static multileaf collimator (SMLC), along with three-dimensional treatment planning system (3-D TPS), open the possibility of tissue compensation. A method using electronic tissue compensator (ETC) has been implemented in Eclipse 3-D TPS (V 7.3, Varian Medical...

Descripción completa

Detalles Bibliográficos
Autores principales: Kinhikar, Rajesh A., Sharma, Pramod K., Patkar, Sachin, Tambe, Chandrashekhar M., Deshpande, Deepak D.
Formato: Texto
Lenguaje:English
Publicado: Medknow Publications 2007
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3000531/
https://www.ncbi.nlm.nih.gov/pubmed/21157535
http://dx.doi.org/10.4103/0971-6203.33241
_version_ 1782193551171387392
author Kinhikar, Rajesh A.
Sharma, Pramod K.
Patkar, Sachin
Tambe, Chandrashekhar M.
Deshpande, Deepak D.
author_facet Kinhikar, Rajesh A.
Sharma, Pramod K.
Patkar, Sachin
Tambe, Chandrashekhar M.
Deshpande, Deepak D.
author_sort Kinhikar, Rajesh A.
collection PubMed
description Dynamic multileaf collimator (DMLC) and static multileaf collimator (SMLC), along with three-dimensional treatment planning system (3-D TPS), open the possibility of tissue compensation. A method using electronic tissue compensator (ETC) has been implemented in Eclipse 3-D TPS (V 7.3, Varian Medical Systems, Palo Alto, USA) at our center. The ETC was tested for head and neck conformal radiotherapy planning. The purpose of this study was to verify the feasibility of DMLC and SMLC in head and neck field irradiation for delivering homogeneous dose in the midplane at a pre-defined depth. In addition, emphasis was given to the dosimetric aspects in commissioning ETC in Eclipse. A Head and Neck Phantom (The Phantom Laboratory, USA) was used for the dosimetric verification. Planning was carried out for both DMLC and SMLC ETC plans. The dose calculated at central axis by eclipse with DMLC and SMLC was noted. This was compared with the doses measured on machine with ion chamber and thermoluminescence dosimetry (TLD). The calculated isodose curves and profiles were compared with the measured ones. The dose profiles along the two major axes from Eclipse were also compared with the profiles obtained from Amorphous Silicon (AS500) Electronic portal imaging device (EPID) on Clinac 6 EX machine. In uniform dose regions, measured dose values agreed with the calculated doses within 3%. Agreement between calculated and measured isodoses in the dose gradient zone was within 3 mm. The isodose curves and the profiles were found to be in good agreement with the measured curves and profiles. The measured and the calculated dose profiles along the two major axes were flat for both DMLC and SMLC. The dosimetric verification of ETC for both the linacs demonstrated the feasibility and the accuracy of the ETC treatment modality for achieving uniform dose distributions. Therefore, ETC can be used as a tool in head and neck treatment planning optimization for improved dose uniformity.
format Text
id pubmed-3000531
institution National Center for Biotechnology Information
language English
publishDate 2007
publisher Medknow Publications
record_format MEDLINE/PubMed
spelling pubmed-30005312010-12-14 Electronic tissue compensation achieved with both dynamic and static multileaf collimator in eclipse treatment planning system for Clinac 6 EX and 2100 CD Varian linear accelerators: Feasibility and dosimetric study Kinhikar, Rajesh A. Sharma, Pramod K. Patkar, Sachin Tambe, Chandrashekhar M. Deshpande, Deepak D. J Med Phys Original Article Dynamic multileaf collimator (DMLC) and static multileaf collimator (SMLC), along with three-dimensional treatment planning system (3-D TPS), open the possibility of tissue compensation. A method using electronic tissue compensator (ETC) has been implemented in Eclipse 3-D TPS (V 7.3, Varian Medical Systems, Palo Alto, USA) at our center. The ETC was tested for head and neck conformal radiotherapy planning. The purpose of this study was to verify the feasibility of DMLC and SMLC in head and neck field irradiation for delivering homogeneous dose in the midplane at a pre-defined depth. In addition, emphasis was given to the dosimetric aspects in commissioning ETC in Eclipse. A Head and Neck Phantom (The Phantom Laboratory, USA) was used for the dosimetric verification. Planning was carried out for both DMLC and SMLC ETC plans. The dose calculated at central axis by eclipse with DMLC and SMLC was noted. This was compared with the doses measured on machine with ion chamber and thermoluminescence dosimetry (TLD). The calculated isodose curves and profiles were compared with the measured ones. The dose profiles along the two major axes from Eclipse were also compared with the profiles obtained from Amorphous Silicon (AS500) Electronic portal imaging device (EPID) on Clinac 6 EX machine. In uniform dose regions, measured dose values agreed with the calculated doses within 3%. Agreement between calculated and measured isodoses in the dose gradient zone was within 3 mm. The isodose curves and the profiles were found to be in good agreement with the measured curves and profiles. The measured and the calculated dose profiles along the two major axes were flat for both DMLC and SMLC. The dosimetric verification of ETC for both the linacs demonstrated the feasibility and the accuracy of the ETC treatment modality for achieving uniform dose distributions. Therefore, ETC can be used as a tool in head and neck treatment planning optimization for improved dose uniformity. Medknow Publications 2007 /pmc/articles/PMC3000531/ /pubmed/21157535 http://dx.doi.org/10.4103/0971-6203.33241 Text en © Journal of Medical Physics http://creativecommons.org/licenses/by/2.0/ This is an open-access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
spellingShingle Original Article
Kinhikar, Rajesh A.
Sharma, Pramod K.
Patkar, Sachin
Tambe, Chandrashekhar M.
Deshpande, Deepak D.
Electronic tissue compensation achieved with both dynamic and static multileaf collimator in eclipse treatment planning system for Clinac 6 EX and 2100 CD Varian linear accelerators: Feasibility and dosimetric study
title Electronic tissue compensation achieved with both dynamic and static multileaf collimator in eclipse treatment planning system for Clinac 6 EX and 2100 CD Varian linear accelerators: Feasibility and dosimetric study
title_full Electronic tissue compensation achieved with both dynamic and static multileaf collimator in eclipse treatment planning system for Clinac 6 EX and 2100 CD Varian linear accelerators: Feasibility and dosimetric study
title_fullStr Electronic tissue compensation achieved with both dynamic and static multileaf collimator in eclipse treatment planning system for Clinac 6 EX and 2100 CD Varian linear accelerators: Feasibility and dosimetric study
title_full_unstemmed Electronic tissue compensation achieved with both dynamic and static multileaf collimator in eclipse treatment planning system for Clinac 6 EX and 2100 CD Varian linear accelerators: Feasibility and dosimetric study
title_short Electronic tissue compensation achieved with both dynamic and static multileaf collimator in eclipse treatment planning system for Clinac 6 EX and 2100 CD Varian linear accelerators: Feasibility and dosimetric study
title_sort electronic tissue compensation achieved with both dynamic and static multileaf collimator in eclipse treatment planning system for clinac 6 ex and 2100 cd varian linear accelerators: feasibility and dosimetric study
topic Original Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3000531/
https://www.ncbi.nlm.nih.gov/pubmed/21157535
http://dx.doi.org/10.4103/0971-6203.33241
work_keys_str_mv AT kinhikarrajesha electronictissuecompensationachievedwithbothdynamicandstaticmultileafcollimatorineclipsetreatmentplanningsystemforclinac6exand2100cdvarianlinearacceleratorsfeasibilityanddosimetricstudy
AT sharmapramodk electronictissuecompensationachievedwithbothdynamicandstaticmultileafcollimatorineclipsetreatmentplanningsystemforclinac6exand2100cdvarianlinearacceleratorsfeasibilityanddosimetricstudy
AT patkarsachin electronictissuecompensationachievedwithbothdynamicandstaticmultileafcollimatorineclipsetreatmentplanningsystemforclinac6exand2100cdvarianlinearacceleratorsfeasibilityanddosimetricstudy
AT tambechandrashekharm electronictissuecompensationachievedwithbothdynamicandstaticmultileafcollimatorineclipsetreatmentplanningsystemforclinac6exand2100cdvarianlinearacceleratorsfeasibilityanddosimetricstudy
AT deshpandedeepakd electronictissuecompensationachievedwithbothdynamicandstaticmultileafcollimatorineclipsetreatmentplanningsystemforclinac6exand2100cdvarianlinearacceleratorsfeasibilityanddosimetricstudy