Compensation of motion artifacts in catheter-based optical frequency domain imaging

A novel heterodyne Doppler interferometer method for compensating motion artifacts caused by cardiac motion in intracoronary optical frequency domain imaging (OFDI) is demonstrated. To track the relative motion of a catheter with regard to the vessel, a motion tracking system is incorporated with a...

Descripción completa

Detalles Bibliográficos
Autores principales: Ha, J. Y., Shishkov, M., Colice, M., Oh, W. Y., Yoo, H., Liu, L., Tearney, G. J., Bouma, B. E.
Formato: Texto
Lenguaje:English
Publicado: Optical Society of America 2010
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3003228/
https://www.ncbi.nlm.nih.gov/pubmed/20589002
http://dx.doi.org/10.1364/OE.18.011418
Descripción
Sumario:A novel heterodyne Doppler interferometer method for compensating motion artifacts caused by cardiac motion in intracoronary optical frequency domain imaging (OFDI) is demonstrated. To track the relative motion of a catheter with regard to the vessel, a motion tracking system is incorporated with a standard OFDI system by using wavelength division multiplexing (WDM) techniques. Without affecting the imaging beam, dual WDM monochromatic beams are utilized for tracking the relative radial and longitudinal velocities of a catheter-based fiber probe. Our results demonstrate that tracking instantaneous velocity can be used to compensate for distortion in the images due to motion artifacts, thus leading to accurate reconstruction and volumetric measurements with catheter-based imaging.