Performance evaluation of a dedicated computed tomography scanner used for virtual simulation using in-house fabricated CT phantoms
Comprehensive tests on single slice CT scanner was carried out using in-house fabricated phantoms/test tools following AAPM recommended methods to independently validate the auto-performance test (APT) results. Test results of all the electromechanical parameters were found within the specified limi...
Autores principales: | Sharma, D. S., Sharma, S. D., Sanu, K. K., Saju, S., Deshpande, D. D., Kannan, S. |
---|---|
Formato: | Texto |
Lenguaje: | English |
Publicado: |
Medknow Publications
2006
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3003891/ https://www.ncbi.nlm.nih.gov/pubmed/21206637 http://dx.doi.org/10.4103/0971-6203.25667 |