Optimization of the imaging protocol of an X-ray CT scanner for evaluation of normoxic polymer gel dosimeters
X-ray computer tomography (CT) has previously been reported as an evaluation tool for polymer gel (PAG) dosimeters. In this study, the imaging protocol of a Siemens Emotion X-ray CT scanner was optimized to evaluate PAGAT normoxic gel dosimeters. The scan parameters were optimized as 130 kV and 150...
Autores principales: | , , |
---|---|
Formato: | Texto |
Lenguaje: | English |
Publicado: |
Medknow Publication
2006
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3004139/ https://www.ncbi.nlm.nih.gov/pubmed/21206668 http://dx.doi.org/10.4103/0971-6203.26692 |
_version_ | 1782193955732979712 |
---|---|
author | Subramanian, Brindha Ravindran, Paul B. Baldock, Clive |
author_facet | Subramanian, Brindha Ravindran, Paul B. Baldock, Clive |
author_sort | Subramanian, Brindha |
collection | PubMed |
description | X-ray computer tomography (CT) has previously been reported as an evaluation tool for polymer gel (PAG) dosimeters. In this study, the imaging protocol of a Siemens Emotion X-ray CT scanner was optimized to evaluate PAGAT normoxic gel dosimeters. The scan parameters were optimized as 130 kV and 150 mA with a slice thickness of 3 mm for smaller fields and 5 mm for larger fields of irradiation. The number of images to be averaged to reduce noise to an acceptable level was concluded to be 25. It was also concluded that the total monomer concentration required is 7% with 10 mM THP to obtain a maximum change in CT number at dose levels up to 15 Gy for evaluation with X-ray CT. Optimal scan parameters may vary with X-ray CT scanner. Hence the imaging protocol of each scanner to be used for evaluating polymer gels requires individual optimization for the purposes of gel dosimetry evaluation. |
format | Text |
id | pubmed-3004139 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2006 |
publisher | Medknow Publication |
record_format | MEDLINE/PubMed |
spelling | pubmed-30041392011-01-04 Optimization of the imaging protocol of an X-ray CT scanner for evaluation of normoxic polymer gel dosimeters Subramanian, Brindha Ravindran, Paul B. Baldock, Clive J Med Phys Original Article X-ray computer tomography (CT) has previously been reported as an evaluation tool for polymer gel (PAG) dosimeters. In this study, the imaging protocol of a Siemens Emotion X-ray CT scanner was optimized to evaluate PAGAT normoxic gel dosimeters. The scan parameters were optimized as 130 kV and 150 mA with a slice thickness of 3 mm for smaller fields and 5 mm for larger fields of irradiation. The number of images to be averaged to reduce noise to an acceptable level was concluded to be 25. It was also concluded that the total monomer concentration required is 7% with 10 mM THP to obtain a maximum change in CT number at dose levels up to 15 Gy for evaluation with X-ray CT. Optimal scan parameters may vary with X-ray CT scanner. Hence the imaging protocol of each scanner to be used for evaluating polymer gels requires individual optimization for the purposes of gel dosimetry evaluation. Medknow Publication 2006 /pmc/articles/PMC3004139/ /pubmed/21206668 http://dx.doi.org/10.4103/0971-6203.26692 Text en © Journal of Medical Physics http://creativecommons.org/licenses/by/2.0/ This is an open-access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited. |
spellingShingle | Original Article Subramanian, Brindha Ravindran, Paul B. Baldock, Clive Optimization of the imaging protocol of an X-ray CT scanner for evaluation of normoxic polymer gel dosimeters |
title | Optimization of the imaging protocol of an X-ray CT scanner for evaluation of normoxic polymer gel dosimeters |
title_full | Optimization of the imaging protocol of an X-ray CT scanner for evaluation of normoxic polymer gel dosimeters |
title_fullStr | Optimization of the imaging protocol of an X-ray CT scanner for evaluation of normoxic polymer gel dosimeters |
title_full_unstemmed | Optimization of the imaging protocol of an X-ray CT scanner for evaluation of normoxic polymer gel dosimeters |
title_short | Optimization of the imaging protocol of an X-ray CT scanner for evaluation of normoxic polymer gel dosimeters |
title_sort | optimization of the imaging protocol of an x-ray ct scanner for evaluation of normoxic polymer gel dosimeters |
topic | Original Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3004139/ https://www.ncbi.nlm.nih.gov/pubmed/21206668 http://dx.doi.org/10.4103/0971-6203.26692 |
work_keys_str_mv | AT subramanianbrindha optimizationoftheimagingprotocolofanxrayctscannerforevaluationofnormoxicpolymergeldosimeters AT ravindranpaulb optimizationoftheimagingprotocolofanxrayctscannerforevaluationofnormoxicpolymergeldosimeters AT baldockclive optimizationoftheimagingprotocolofanxrayctscannerforevaluationofnormoxicpolymergeldosimeters |