Optimization of the imaging protocol of an X-ray CT scanner for evaluation of normoxic polymer gel dosimeters

X-ray computer tomography (CT) has previously been reported as an evaluation tool for polymer gel (PAG) dosimeters. In this study, the imaging protocol of a Siemens Emotion X-ray CT scanner was optimized to evaluate PAGAT normoxic gel dosimeters. The scan parameters were optimized as 130 kV and 150...

Descripción completa

Detalles Bibliográficos
Autores principales: Subramanian, Brindha, Ravindran, Paul B., Baldock, Clive
Formato: Texto
Lenguaje:English
Publicado: Medknow Publication 2006
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3004139/
https://www.ncbi.nlm.nih.gov/pubmed/21206668
http://dx.doi.org/10.4103/0971-6203.26692
_version_ 1782193955732979712
author Subramanian, Brindha
Ravindran, Paul B.
Baldock, Clive
author_facet Subramanian, Brindha
Ravindran, Paul B.
Baldock, Clive
author_sort Subramanian, Brindha
collection PubMed
description X-ray computer tomography (CT) has previously been reported as an evaluation tool for polymer gel (PAG) dosimeters. In this study, the imaging protocol of a Siemens Emotion X-ray CT scanner was optimized to evaluate PAGAT normoxic gel dosimeters. The scan parameters were optimized as 130 kV and 150 mA with a slice thickness of 3 mm for smaller fields and 5 mm for larger fields of irradiation. The number of images to be averaged to reduce noise to an acceptable level was concluded to be 25. It was also concluded that the total monomer concentration required is 7% with 10 mM THP to obtain a maximum change in CT number at dose levels up to 15 Gy for evaluation with X-ray CT. Optimal scan parameters may vary with X-ray CT scanner. Hence the imaging protocol of each scanner to be used for evaluating polymer gels requires individual optimization for the purposes of gel dosimetry evaluation.
format Text
id pubmed-3004139
institution National Center for Biotechnology Information
language English
publishDate 2006
publisher Medknow Publication
record_format MEDLINE/PubMed
spelling pubmed-30041392011-01-04 Optimization of the imaging protocol of an X-ray CT scanner for evaluation of normoxic polymer gel dosimeters Subramanian, Brindha Ravindran, Paul B. Baldock, Clive J Med Phys Original Article X-ray computer tomography (CT) has previously been reported as an evaluation tool for polymer gel (PAG) dosimeters. In this study, the imaging protocol of a Siemens Emotion X-ray CT scanner was optimized to evaluate PAGAT normoxic gel dosimeters. The scan parameters were optimized as 130 kV and 150 mA with a slice thickness of 3 mm for smaller fields and 5 mm for larger fields of irradiation. The number of images to be averaged to reduce noise to an acceptable level was concluded to be 25. It was also concluded that the total monomer concentration required is 7% with 10 mM THP to obtain a maximum change in CT number at dose levels up to 15 Gy for evaluation with X-ray CT. Optimal scan parameters may vary with X-ray CT scanner. Hence the imaging protocol of each scanner to be used for evaluating polymer gels requires individual optimization for the purposes of gel dosimetry evaluation. Medknow Publication 2006 /pmc/articles/PMC3004139/ /pubmed/21206668 http://dx.doi.org/10.4103/0971-6203.26692 Text en © Journal of Medical Physics http://creativecommons.org/licenses/by/2.0/ This is an open-access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
spellingShingle Original Article
Subramanian, Brindha
Ravindran, Paul B.
Baldock, Clive
Optimization of the imaging protocol of an X-ray CT scanner for evaluation of normoxic polymer gel dosimeters
title Optimization of the imaging protocol of an X-ray CT scanner for evaluation of normoxic polymer gel dosimeters
title_full Optimization of the imaging protocol of an X-ray CT scanner for evaluation of normoxic polymer gel dosimeters
title_fullStr Optimization of the imaging protocol of an X-ray CT scanner for evaluation of normoxic polymer gel dosimeters
title_full_unstemmed Optimization of the imaging protocol of an X-ray CT scanner for evaluation of normoxic polymer gel dosimeters
title_short Optimization of the imaging protocol of an X-ray CT scanner for evaluation of normoxic polymer gel dosimeters
title_sort optimization of the imaging protocol of an x-ray ct scanner for evaluation of normoxic polymer gel dosimeters
topic Original Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3004139/
https://www.ncbi.nlm.nih.gov/pubmed/21206668
http://dx.doi.org/10.4103/0971-6203.26692
work_keys_str_mv AT subramanianbrindha optimizationoftheimagingprotocolofanxrayctscannerforevaluationofnormoxicpolymergeldosimeters
AT ravindranpaulb optimizationoftheimagingprotocolofanxrayctscannerforevaluationofnormoxicpolymergeldosimeters
AT baldockclive optimizationoftheimagingprotocolofanxrayctscannerforevaluationofnormoxicpolymergeldosimeters