Development and Validation of HPLC Method for the Determination of Pregabalin in Capsules
A simple, precise, specific, and accurate reverse phase HPLC method has been developed for the determination of pregabalin in capsule dosage form. The chromatography was set on Hypersil BDS, C8, 150×4.6 mm, 5 μm column using photodiode array detector. The mobile phase consisting of phosphate buffer...
Autores principales: | , |
---|---|
Formato: | Texto |
Lenguaje: | English |
Publicado: |
Medknow Publications
2010
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3013579/ https://www.ncbi.nlm.nih.gov/pubmed/21218069 http://dx.doi.org/10.4103/0250-474X.73935 |
Sumario: | A simple, precise, specific, and accurate reverse phase HPLC method has been developed for the determination of pregabalin in capsule dosage form. The chromatography was set on Hypersil BDS, C8, 150×4.6 mm, 5 μm column using photodiode array detector. The mobile phase consisting of phosphate buffer pH 6.9 and acetonitrile in the ratio of 95:05 with flow rate of 1 ml/min. The method was validated according to ICH guidelines with respect to specificity, linearity, accuracy, precision and robustness. Lower limit of quantification is 0.6 mg/l. The pregabalin sample solution was found to be stable at room temperature for about 26 h. |
---|