Suboptimal Performance of Blood Glucose Meters in an Antenatal Diabetes Clinic

OBJECTIVE: The objective of this study was to evaluate the performance of blood glucose meters in diabetes associated with pregnancy (DP). RESEARCH DESIGN AND METHODS: Finger-prick blood glucose levels measured using six different glucose meters on 102 patients with DP attending an antenatal clinic...

Descripción completa

Detalles Bibliográficos
Autores principales: Perera, Nimalie Jacintha, Molyneaux, Lynda, Constantino, Maria Ines, McGill, Marg, Yue, Dennis Koon-See, Twigg, Stephen Morris, Ross, Glynis Pauline
Formato: Texto
Lenguaje:English
Publicado: American Diabetes Association 2011
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3024344/
https://www.ncbi.nlm.nih.gov/pubmed/21216849
http://dx.doi.org/10.2337/dc10-1457
_version_ 1782196767895322624
author Perera, Nimalie Jacintha
Molyneaux, Lynda
Constantino, Maria Ines
McGill, Marg
Yue, Dennis Koon-See
Twigg, Stephen Morris
Ross, Glynis Pauline
author_facet Perera, Nimalie Jacintha
Molyneaux, Lynda
Constantino, Maria Ines
McGill, Marg
Yue, Dennis Koon-See
Twigg, Stephen Morris
Ross, Glynis Pauline
author_sort Perera, Nimalie Jacintha
collection PubMed
description OBJECTIVE: The objective of this study was to evaluate the performance of blood glucose meters in diabetes associated with pregnancy (DP). RESEARCH DESIGN AND METHODS: Finger-prick blood glucose levels measured using six different glucose meters on 102 patients with DP attending an antenatal clinic were compared with laboratory plasma glucose results. HbA(1c) and hematocrit were also measured. RESULTS: The plasma glucose range was 2.2–9.4 mmol/L with hematocrit 33–37% and mean HbA(1c) 5.5% ± 0.56 (SD). All meters provided plasma equivalent results except one, which reported whole blood glucose that was adjusted to plasma equivalent values. The absolute glucose difference [meter − plasma glucose] was 0.232 ± 0.69 to 0.725 ± 0.62 mmol/L mean ± SD and bias ranged from 6.1 to 15.8%. Two meters were affected by hematocrit <36% (P < 0.05). CONCLUSIONS: Blood glucose meters in current use are not optimally accurate when compared with plasma glucose measurement in DP. Recognition of this deviation is essential to prevent inappropriate treatment of DP.
format Text
id pubmed-3024344
institution National Center for Biotechnology Information
language English
publishDate 2011
publisher American Diabetes Association
record_format MEDLINE/PubMed
spelling pubmed-30243442012-02-01 Suboptimal Performance of Blood Glucose Meters in an Antenatal Diabetes Clinic Perera, Nimalie Jacintha Molyneaux, Lynda Constantino, Maria Ines McGill, Marg Yue, Dennis Koon-See Twigg, Stephen Morris Ross, Glynis Pauline Diabetes Care Original Research OBJECTIVE: The objective of this study was to evaluate the performance of blood glucose meters in diabetes associated with pregnancy (DP). RESEARCH DESIGN AND METHODS: Finger-prick blood glucose levels measured using six different glucose meters on 102 patients with DP attending an antenatal clinic were compared with laboratory plasma glucose results. HbA(1c) and hematocrit were also measured. RESULTS: The plasma glucose range was 2.2–9.4 mmol/L with hematocrit 33–37% and mean HbA(1c) 5.5% ± 0.56 (SD). All meters provided plasma equivalent results except one, which reported whole blood glucose that was adjusted to plasma equivalent values. The absolute glucose difference [meter − plasma glucose] was 0.232 ± 0.69 to 0.725 ± 0.62 mmol/L mean ± SD and bias ranged from 6.1 to 15.8%. Two meters were affected by hematocrit <36% (P < 0.05). CONCLUSIONS: Blood glucose meters in current use are not optimally accurate when compared with plasma glucose measurement in DP. Recognition of this deviation is essential to prevent inappropriate treatment of DP. American Diabetes Association 2011-02 2011-01-20 /pmc/articles/PMC3024344/ /pubmed/21216849 http://dx.doi.org/10.2337/dc10-1457 Text en © 2011 by the American Diabetes Association. Readers may use this article as long as the work is properly cited, the use is educational and not for profit, and the work is not altered. See http://creativecommons.org/licenses/by-nc-nd/3.0/ for details.
spellingShingle Original Research
Perera, Nimalie Jacintha
Molyneaux, Lynda
Constantino, Maria Ines
McGill, Marg
Yue, Dennis Koon-See
Twigg, Stephen Morris
Ross, Glynis Pauline
Suboptimal Performance of Blood Glucose Meters in an Antenatal Diabetes Clinic
title Suboptimal Performance of Blood Glucose Meters in an Antenatal Diabetes Clinic
title_full Suboptimal Performance of Blood Glucose Meters in an Antenatal Diabetes Clinic
title_fullStr Suboptimal Performance of Blood Glucose Meters in an Antenatal Diabetes Clinic
title_full_unstemmed Suboptimal Performance of Blood Glucose Meters in an Antenatal Diabetes Clinic
title_short Suboptimal Performance of Blood Glucose Meters in an Antenatal Diabetes Clinic
title_sort suboptimal performance of blood glucose meters in an antenatal diabetes clinic
topic Original Research
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3024344/
https://www.ncbi.nlm.nih.gov/pubmed/21216849
http://dx.doi.org/10.2337/dc10-1457
work_keys_str_mv AT pereranimaliejacintha suboptimalperformanceofbloodglucosemetersinanantenataldiabetesclinic
AT molyneauxlynda suboptimalperformanceofbloodglucosemetersinanantenataldiabetesclinic
AT constantinomariaines suboptimalperformanceofbloodglucosemetersinanantenataldiabetesclinic
AT mcgillmarg suboptimalperformanceofbloodglucosemetersinanantenataldiabetesclinic
AT yuedenniskoonsee suboptimalperformanceofbloodglucosemetersinanantenataldiabetesclinic
AT twiggstephenmorris suboptimalperformanceofbloodglucosemetersinanantenataldiabetesclinic
AT rossglynispauline suboptimalperformanceofbloodglucosemetersinanantenataldiabetesclinic