Cargando…

Arterial pressure-based cardiac output monitoring: a multicenter validation of the third-generation software in septic patients

PURPOSE: Second-generation FloTrac software has been shown to reliably measure cardiac output (CO) in cardiac surgical patients. However, concerns have been raised regarding its accuracy in vasoplegic states. The aim of the present multicenter study was to investigate the accuracy of the third-gener...

Descripción completa

Detalles Bibliográficos
Autores principales: De Backer, Daniel, Marx, Gernot, Tan, Andrew, Junker, Christopher, Van Nuffelen, Marc, Hüter, Lars, Ching, Willy, Michard, Frédéric, Vincent, Jean-Louis
Formato: Texto
Lenguaje:English
Publicado: Springer-Verlag 2010
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3028067/
https://www.ncbi.nlm.nih.gov/pubmed/21153399
http://dx.doi.org/10.1007/s00134-010-2098-8
Descripción
Sumario:PURPOSE: Second-generation FloTrac software has been shown to reliably measure cardiac output (CO) in cardiac surgical patients. However, concerns have been raised regarding its accuracy in vasoplegic states. The aim of the present multicenter study was to investigate the accuracy of the third-generation software in patients with sepsis, particularly when total systemic vascular resistance (TSVR) is low. METHODS: Fifty-eight septic patients were included in this prospective observational study in four university-affiliated ICUs. Reference CO was measured by bolus pulmonary thermodilution (iCO) using 3–5 cold saline boluses. Simultaneously, CO was computed from the arterial pressure curve recorded on a computer using the second-generation (CO(G2)) and third-generation (CO(G3)) FloTrac software. CO was also measured by semi-continuous pulmonary thermodilution (CCO). RESULTS: A total of 401 simultaneous measurements of iCO, CO(G2), CO(G3), and CCO were recorded. The mean (95%CI) biases between CO(G2) and iCO, CO(G3) and iCO, and CCO and iCO were −10 (−15 to −5)% [−0.8 (−1.1 to −0.4) L/min], 0 (−4 to 4)% [0 (−0.3 to 0.3) L/min], and 9 (6–13)% [0.7 (0.5–1.0) L/min], respectively. The percentage errors were 29 (20–37)% for CO(G2), 30 (24–37)% for CO(G3), and 28 (22–34)% for CCO. The difference between iCO and CO(G2) was significantly correlated with TSVR (r (2) = 0.37, p < 0.0001). A very weak (r (2) = 0.05) relationship was also observed for the difference between iCO and CO(G3). CONCLUSIONS: In patients with sepsis, the third-generation FloTrac software is more accurate, as precise, and less influenced by TSVR than the second-generation software. ELECTRONIC SUPPLEMENTARY MATERIAL: The online version of this article (doi:10.1007/s00134-010-2098-8) contains supplementary material, which is available to authorized users.