Biomechanical testing of rectangular humeral shaft prosthesis: higher torsional stability without increased fracture risk

BACKGROUND: Rectangular cementless femur shaft prostheses have a higher primary stability than round shafts. A novel rectangular humeral shaft design was tested with two questions: does the rectangular design cause a higher fracture risk during implantation than round designs, and does it increase t...

Descripción completa

Detalles Bibliográficos
Autores principales: Flury, Matthias P., Schmoelz, Werner, Schreiber, Ulrich, Goldhahn, Joerg
Formato: Texto
Lenguaje:English
Publicado: Springer-Verlag 2010
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3034038/
https://www.ncbi.nlm.nih.gov/pubmed/20857127
http://dx.doi.org/10.1007/s00402-010-1170-8
Descripción
Sumario:BACKGROUND: Rectangular cementless femur shaft prostheses have a higher primary stability than round shafts. A novel rectangular humeral shaft design was tested with two questions: does the rectangular design cause a higher fracture risk during implantation than round designs, and does it increase the torsional stiffness? MATERIALS AND METHODS: Two series with six paired human humeri (total 24) were tested on one side with the rectangular shaft and on the contralateral side with a round shaft. In the first series, the shaft implantation was carried out with a constant speed of 100 mm/min and the maximum force was measured when the fracture occurred. In the second series, the implants were preloaded with 50 N and then rotated at 2° per second with monitoring of the torsional torque. RESULTS: The maximum force at fracture showed no significant difference for the two designs (p = 0.34). Higher age and low bone density reduced the force required for fracture. The rectangular shaft showed significant higher torsional moments (p < 0.05). CONCLUSIONS: In biomechanical testing, the rectangular shaft had a significantly higher primary torsional stability than the round shaft without a higher risk of fracture during cementless implantation. Fracture risk and torsional stability are influenced by age and bone density.