Comparison of lens- and fiber-coupled CCD detectors for X-ray computed tomography

X-ray imaging detectors with an identical phosphor and a CCD chip but employing lens- and fiber-coupling between them have been compared. These are designed for X-ray imaging experiments, especially computed tomography, at the medium-length beamline at the SPring-8 synchrotron radiation facility. It...

Descripción completa

Detalles Bibliográficos
Autores principales: Uesugi, K., Hoshino, M., Yagi, N.
Formato: Texto
Lenguaje:English
Publicado: International Union of Crystallography 2011
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3042329/
https://www.ncbi.nlm.nih.gov/pubmed/21335908
http://dx.doi.org/10.1107/S0909049510044523
_version_ 1782198530490761216
author Uesugi, K.
Hoshino, M.
Yagi, N.
author_facet Uesugi, K.
Hoshino, M.
Yagi, N.
author_sort Uesugi, K.
collection PubMed
description X-ray imaging detectors with an identical phosphor and a CCD chip but employing lens- and fiber-coupling between them have been compared. These are designed for X-ray imaging experiments, especially computed tomography, at the medium-length beamline at the SPring-8 synchrotron radiation facility. It was found that the transmittance of light to the CCD is about four times higher in the fiber-coupled detector. The uniformity of response in the lens-coupled detector has a global shading of up to 40%, while pixel-to-pixel variation owing to a chicken-wire pattern was dominant in the fiber-coupled detector. Apart from the higher transmittance, the fiber-coupled detector has a few characteristics that require attention when it is used for computed tomography, which are browning of the fiber, discontinuity in the image, image distortion, and dark spots in the chicken-wire pattern. Thus, it is most suitable for high-speed tomography of samples that tend to deform, for example biological and soft materials.
format Text
id pubmed-3042329
institution National Center for Biotechnology Information
language English
publishDate 2011
publisher International Union of Crystallography
record_format MEDLINE/PubMed
spelling pubmed-30423292011-03-10 Comparison of lens- and fiber-coupled CCD detectors for X-ray computed tomography Uesugi, K. Hoshino, M. Yagi, N. J Synchrotron Radiat Research Papers X-ray imaging detectors with an identical phosphor and a CCD chip but employing lens- and fiber-coupling between them have been compared. These are designed for X-ray imaging experiments, especially computed tomography, at the medium-length beamline at the SPring-8 synchrotron radiation facility. It was found that the transmittance of light to the CCD is about four times higher in the fiber-coupled detector. The uniformity of response in the lens-coupled detector has a global shading of up to 40%, while pixel-to-pixel variation owing to a chicken-wire pattern was dominant in the fiber-coupled detector. Apart from the higher transmittance, the fiber-coupled detector has a few characteristics that require attention when it is used for computed tomography, which are browning of the fiber, discontinuity in the image, image distortion, and dark spots in the chicken-wire pattern. Thus, it is most suitable for high-speed tomography of samples that tend to deform, for example biological and soft materials. International Union of Crystallography 2011-03-01 2010-12-02 /pmc/articles/PMC3042329/ /pubmed/21335908 http://dx.doi.org/10.1107/S0909049510044523 Text en © K. Uesugi et al. 2011 http://creativecommons.org/licenses/by/2.0/uk/ This is an open-access article distributed under the terms of the Creative Commons Attribution Licence, which permits unrestricted use, distribution, and reproduction in any medium, provided the original authors and source are cited.
spellingShingle Research Papers
Uesugi, K.
Hoshino, M.
Yagi, N.
Comparison of lens- and fiber-coupled CCD detectors for X-ray computed tomography
title Comparison of lens- and fiber-coupled CCD detectors for X-ray computed tomography
title_full Comparison of lens- and fiber-coupled CCD detectors for X-ray computed tomography
title_fullStr Comparison of lens- and fiber-coupled CCD detectors for X-ray computed tomography
title_full_unstemmed Comparison of lens- and fiber-coupled CCD detectors for X-ray computed tomography
title_short Comparison of lens- and fiber-coupled CCD detectors for X-ray computed tomography
title_sort comparison of lens- and fiber-coupled ccd detectors for x-ray computed tomography
topic Research Papers
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3042329/
https://www.ncbi.nlm.nih.gov/pubmed/21335908
http://dx.doi.org/10.1107/S0909049510044523
work_keys_str_mv AT uesugik comparisonoflensandfibercoupledccddetectorsforxraycomputedtomography
AT hoshinom comparisonoflensandfibercoupledccddetectorsforxraycomputedtomography
AT yagin comparisonoflensandfibercoupledccddetectorsforxraycomputedtomography