Cargando…

Application of headspace solid-phase microextraction followed by gas chromatography–mass spectrometry to determine short-chain alkane monocarboxylic acids in aqueous samples

In this study, a procedure was developed to determine short-chain alkane monocarboxylic acids (SCMAs) in aqueous samples using headspace solid-phase microextraction (HS-SPME) followed by gas chromatography (GC) coupled with mass spectrometry (MS). A Stabilwax-DA capillary column (30 m × 0.32-mm inne...

Descripción completa

Detalles Bibliográficos
Autores principales: Banel, Anna, Wasielewska, Marta, Zygmunt, Bogdan
Formato: Texto
Lenguaje:English
Publicado: Springer-Verlag 2010
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3044222/
https://www.ncbi.nlm.nih.gov/pubmed/21107976
http://dx.doi.org/10.1007/s00216-010-4415-z
Descripción
Sumario:In this study, a procedure was developed to determine short-chain alkane monocarboxylic acids (SCMAs) in aqueous samples using headspace solid-phase microextraction (HS-SPME) followed by gas chromatography (GC) coupled with mass spectrometry (MS). A Stabilwax-DA capillary column (30 m × 0.32-mm inner diameter, 0.50-μm film thickness) was used for GC separation and a 60-μm poly(ethylene glycol) fiber was used to isolate SCMAs from water and introduce them into the gas chromatograph. Parameters of HS-SPME, analyte desorption, and GC-MS analysis were selected and an analytical procedure was proposed. Limits of quantitation were on the order of about 0.2 mg L(-1). As an example of the application of the procedure, SCAMs were determined in municipal wastewater at different steps of treatment.