Initial performance evaluation of the UniCel® DxH 800 Coulter® cellular analysis system

The Beckman Coulter UniCel® DxH 800 is a hematology analyzer incorporating new electronic and mechanical design with advanced algorithm technology to perform CBC, white blood cell (WBC) differential, nucleated red blood cell (NRBC), and reticulocyte analysis. Evaluation of this instrument was perfor...

Descripción completa

Detalles Bibliográficos
Autores principales: HEDLEY, B D, KEENEY, M, CHIN-YEE, I, BROWN, W
Formato: Texto
Lenguaje:English
Publicado: Blackwell Publishing Ltd 2011
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3044820/
https://www.ncbi.nlm.nih.gov/pubmed/20491996
http://dx.doi.org/10.1111/j.1751-553X.2010.01239.x
_version_ 1782198781746348032
author HEDLEY, B D
KEENEY, M
CHIN-YEE, I
BROWN, W
author_facet HEDLEY, B D
KEENEY, M
CHIN-YEE, I
BROWN, W
author_sort HEDLEY, B D
collection PubMed
description The Beckman Coulter UniCel® DxH 800 is a hematology analyzer incorporating new electronic and mechanical design with advanced algorithm technology to perform CBC, white blood cell (WBC) differential, nucleated red blood cell (NRBC), and reticulocyte analysis. Evaluation of this instrument was performed in our 800-bed tertiary care hospital and specifically centered upon the correlation of WBC, NRBC, and platelet (PLT) enumeration when compared to a predicate analyzer, the Coulter® LH 780, and flow cytometry (FCM) reference methods. Of particular interest were those samples with morphologically confirmed interference and extreme leukocytosis (evaluated with respect to red blood cell parameter correction). The sample set (n = 272) consisted of morphologically normal and hematologically abnormal patients. Correlation of the WBC, PLT, and NRBC showed r(2) values of 0.994, 0.985, and 0.910 for the DxH 800 vs. FCM, respectively. The presence of interfering particles did not affect the accuracy of the DxH 800 with respect to WBC counts. The DxH 800 showed accurate PLT and NRBC counts in the clinically significant low range when compared to FCM. Compared to the LH 780, flagging rates were significantly reduced (NRBC flag), or equivalent (WBC, PLT flag) on the DxH 800. The DxH 800 demonstrated higher sensitivity and specificity for PLTs and NRBCs and achieved a lower NRBC false negative rate compared to the LH 780. The UniCel® DxH 800 represents a significant improvement to previous impedance analyzers in accurately detecting the presence of NRBCs at counts >1/100 WBC. Furthermore, it provides accurate PLT and WBC counts in the presence of interference and improved NRBC flagging efficiency when compared to the LH 780. Correction of red blood cell parameters is appropriate and accurate in cases of extreme leukocytosis.
format Text
id pubmed-3044820
institution National Center for Biotechnology Information
language English
publishDate 2011
publisher Blackwell Publishing Ltd
record_format MEDLINE/PubMed
spelling pubmed-30448202011-03-02 Initial performance evaluation of the UniCel® DxH 800 Coulter® cellular analysis system HEDLEY, B D KEENEY, M CHIN-YEE, I BROWN, W Int J Lab Hematol Original Articles The Beckman Coulter UniCel® DxH 800 is a hematology analyzer incorporating new electronic and mechanical design with advanced algorithm technology to perform CBC, white blood cell (WBC) differential, nucleated red blood cell (NRBC), and reticulocyte analysis. Evaluation of this instrument was performed in our 800-bed tertiary care hospital and specifically centered upon the correlation of WBC, NRBC, and platelet (PLT) enumeration when compared to a predicate analyzer, the Coulter® LH 780, and flow cytometry (FCM) reference methods. Of particular interest were those samples with morphologically confirmed interference and extreme leukocytosis (evaluated with respect to red blood cell parameter correction). The sample set (n = 272) consisted of morphologically normal and hematologically abnormal patients. Correlation of the WBC, PLT, and NRBC showed r(2) values of 0.994, 0.985, and 0.910 for the DxH 800 vs. FCM, respectively. The presence of interfering particles did not affect the accuracy of the DxH 800 with respect to WBC counts. The DxH 800 showed accurate PLT and NRBC counts in the clinically significant low range when compared to FCM. Compared to the LH 780, flagging rates were significantly reduced (NRBC flag), or equivalent (WBC, PLT flag) on the DxH 800. The DxH 800 demonstrated higher sensitivity and specificity for PLTs and NRBCs and achieved a lower NRBC false negative rate compared to the LH 780. The UniCel® DxH 800 represents a significant improvement to previous impedance analyzers in accurately detecting the presence of NRBCs at counts >1/100 WBC. Furthermore, it provides accurate PLT and WBC counts in the presence of interference and improved NRBC flagging efficiency when compared to the LH 780. Correction of red blood cell parameters is appropriate and accurate in cases of extreme leukocytosis. Blackwell Publishing Ltd 2011-02 /pmc/articles/PMC3044820/ /pubmed/20491996 http://dx.doi.org/10.1111/j.1751-553X.2010.01239.x Text en Copyright © 2011 Blackwell Publishing Ltd http://creativecommons.org/licenses/by/2.5/ Re-use of this article is permitted in accordance with the Creative Commons Deed, Attribution 2.5, which does not permit commercial exploitation.
spellingShingle Original Articles
HEDLEY, B D
KEENEY, M
CHIN-YEE, I
BROWN, W
Initial performance evaluation of the UniCel® DxH 800 Coulter® cellular analysis system
title Initial performance evaluation of the UniCel® DxH 800 Coulter® cellular analysis system
title_full Initial performance evaluation of the UniCel® DxH 800 Coulter® cellular analysis system
title_fullStr Initial performance evaluation of the UniCel® DxH 800 Coulter® cellular analysis system
title_full_unstemmed Initial performance evaluation of the UniCel® DxH 800 Coulter® cellular analysis system
title_short Initial performance evaluation of the UniCel® DxH 800 Coulter® cellular analysis system
title_sort initial performance evaluation of the unicel® dxh 800 coulter® cellular analysis system
topic Original Articles
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3044820/
https://www.ncbi.nlm.nih.gov/pubmed/20491996
http://dx.doi.org/10.1111/j.1751-553X.2010.01239.x
work_keys_str_mv AT hedleybd initialperformanceevaluationoftheuniceldxh800coultercellularanalysissystem
AT keeneym initialperformanceevaluationoftheuniceldxh800coultercellularanalysissystem
AT chinyeei initialperformanceevaluationoftheuniceldxh800coultercellularanalysissystem
AT brownw initialperformanceevaluationoftheuniceldxh800coultercellularanalysissystem