Correlation between ORange(®) (Gen 1, pseudophakic) intraoperative refraction and 1-week postcataract surgery autorefraction

BACKGROUND: This study aimed to find the strength of the correlation between ORange(®) (intraoperative wavefront aberrometer) objective refraction during pseudophakic cataract surgery (Gen 1) and 1-week objective autorefraction postcataract surgery in order to establish the clinical degree of confid...

Descripción completa

Detalles Bibliográficos
Autor principal: Chen, Ming
Formato: Texto
Lenguaje:English
Publicado: Dove Medical Press 2011
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3045070/
https://www.ncbi.nlm.nih.gov/pubmed/21383949
http://dx.doi.org/10.2147/OPTH.S17489
Descripción
Sumario:BACKGROUND: This study aimed to find the strength of the correlation between ORange(®) (intraoperative wavefront aberrometer) objective refraction during pseudophakic cataract surgery (Gen 1) and 1-week objective autorefraction postcataract surgery in order to establish the clinical degree of confidence in the ORange. Gen 1 was used because Gen 2 was not available at the time of the study. However, Gen 1 can still be a good reference. METHODS: Thirty-two consecutive eyes were included in this prospective nonrandomized study. ORange was used after phacoemulsification and intraocular lens implantation. There were no complications and no withdrawals from the study. The refraction data from the ORange were recorded in spherical equivalents and put into Excel. One week postoperatively, autorefraction was done and the data were again recorded in spherical equivalents and put into Excel. Both sets of data were then transferred to SPSS for a correlation study. Because ORange is an objective refractor, it is necessary to use another objective refractor, such as the Zeiss Autorefractor, for comparison. Measurement at 1 week postoperatively is for cornea clearance for autorefraction and is the closest time to intraoperative assessment for comparison. RESULTS: The Pearson correlation was r = +0.56, P < 0.01, which qualifies for a moderate or good correlation (0.5–0.75). CONCLUSION: Intraoperative refraction using the ORange Gen 1 after cataract surgery and intraocular lens implantation is moderate to good when correlated with objective autorefraction 1 week after cataract surgery.