Initial quality performance results using a phantom to simulate chest computed radiography

The aim of this study was to develop a homemade phantom for quantitative quality control in chest computed radiography (CR). The phantom was constructed from copper, aluminium, and polymenthylmethacrylate (PMMA) plates as well as Styrofoam materials. Depending on combinations, the literature suggest...

Descripción completa

Detalles Bibliográficos
Autores principales: Muhogora, Wilbroad, Padovani, Renato, Msaki, Peter
Formato: Texto
Lenguaje:English
Publicado: Medknow Publications & Media Pvt Ltd 2011
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3048950/
https://www.ncbi.nlm.nih.gov/pubmed/21430855
http://dx.doi.org/10.4103/0971-6203.75468
_version_ 1782199200746831872
author Muhogora, Wilbroad
Padovani, Renato
Msaki, Peter
author_facet Muhogora, Wilbroad
Padovani, Renato
Msaki, Peter
author_sort Muhogora, Wilbroad
collection PubMed
description The aim of this study was to develop a homemade phantom for quantitative quality control in chest computed radiography (CR). The phantom was constructed from copper, aluminium, and polymenthylmethacrylate (PMMA) plates as well as Styrofoam materials. Depending on combinations, the literature suggests that these materials can simulate the attenuation and scattering characteristics of lung, heart, and mediastinum. The lung, heart, and mediastinum regions were simulated by 10 mm x 10 mm x 0.5 mm, 10 mm x 10 mm x 0.5 mm and 10 mm x 10 mm x 1 mm copper plates, respectively. A test object of 100 mm x 100 mm and 0.2 mm thick copper was positioned to each region for CNR measurements. The phantom was exposed to x-rays generated by different tube potentials that covered settings in clinical use: 110-120 kVp (HVL=4.26-4.66 mm Al) at a source image distance (SID) of 180 cm. An approach similar to the recommended method in digital mammography was applied to determine the CNR values of phantom images produced by a Kodak CR 850A system with post-processing turned off. Subjective contrast-detail studies were also carried out by using images of Leeds TOR CDR test object acquired under similar exposure conditions as during CNR measurements. For clinical kVp conditions relevant to chest radiography, the CNR was highest over 90-100 kVp range. The CNR data correlated with the results of contrast detail observations. The values of clinical tube potentials at which CNR is the highest are regarded to be optimal kVp settings. The simplicity in phantom construction can offer easy implementation of related quality control program.
format Text
id pubmed-3048950
institution National Center for Biotechnology Information
language English
publishDate 2011
publisher Medknow Publications & Media Pvt Ltd
record_format MEDLINE/PubMed
spelling pubmed-30489502011-03-22 Initial quality performance results using a phantom to simulate chest computed radiography Muhogora, Wilbroad Padovani, Renato Msaki, Peter J Med Phys Original Article The aim of this study was to develop a homemade phantom for quantitative quality control in chest computed radiography (CR). The phantom was constructed from copper, aluminium, and polymenthylmethacrylate (PMMA) plates as well as Styrofoam materials. Depending on combinations, the literature suggests that these materials can simulate the attenuation and scattering characteristics of lung, heart, and mediastinum. The lung, heart, and mediastinum regions were simulated by 10 mm x 10 mm x 0.5 mm, 10 mm x 10 mm x 0.5 mm and 10 mm x 10 mm x 1 mm copper plates, respectively. A test object of 100 mm x 100 mm and 0.2 mm thick copper was positioned to each region for CNR measurements. The phantom was exposed to x-rays generated by different tube potentials that covered settings in clinical use: 110-120 kVp (HVL=4.26-4.66 mm Al) at a source image distance (SID) of 180 cm. An approach similar to the recommended method in digital mammography was applied to determine the CNR values of phantom images produced by a Kodak CR 850A system with post-processing turned off. Subjective contrast-detail studies were also carried out by using images of Leeds TOR CDR test object acquired under similar exposure conditions as during CNR measurements. For clinical kVp conditions relevant to chest radiography, the CNR was highest over 90-100 kVp range. The CNR data correlated with the results of contrast detail observations. The values of clinical tube potentials at which CNR is the highest are regarded to be optimal kVp settings. The simplicity in phantom construction can offer easy implementation of related quality control program. Medknow Publications & Media Pvt Ltd 2011 /pmc/articles/PMC3048950/ /pubmed/21430855 http://dx.doi.org/10.4103/0971-6203.75468 Text en Copyright: © Journal of Medical Physics http://creativecommons.org/licenses/by-nc-sa/3.0 This is an open-access article distributed under the terms of the Creative Commons Attribution-Noncommercial-Share Alike 3.0 Unported, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
spellingShingle Original Article
Muhogora, Wilbroad
Padovani, Renato
Msaki, Peter
Initial quality performance results using a phantom to simulate chest computed radiography
title Initial quality performance results using a phantom to simulate chest computed radiography
title_full Initial quality performance results using a phantom to simulate chest computed radiography
title_fullStr Initial quality performance results using a phantom to simulate chest computed radiography
title_full_unstemmed Initial quality performance results using a phantom to simulate chest computed radiography
title_short Initial quality performance results using a phantom to simulate chest computed radiography
title_sort initial quality performance results using a phantom to simulate chest computed radiography
topic Original Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3048950/
https://www.ncbi.nlm.nih.gov/pubmed/21430855
http://dx.doi.org/10.4103/0971-6203.75468
work_keys_str_mv AT muhogorawilbroad initialqualityperformanceresultsusingaphantomtosimulatechestcomputedradiography
AT padovanirenato initialqualityperformanceresultsusingaphantomtosimulatechestcomputedradiography
AT msakipeter initialqualityperformanceresultsusingaphantomtosimulatechestcomputedradiography