Cargando…
Iliac Vein Injury Due to a Damaged Hot Shears™ Tip Cover During Robot Assisted Radical Prostatectomy
We report a rare case of vascular injury secondary to a damaged Hot Shears™ tip cover. Two 1 mm holes in the tip cover resulted in perforations in the obturator and external iliac veins during pelvic node dissection. Bleeding was controlled with bipolar coagulation and a 5 mm metal clip in the obtur...