Performance Evaluation of PETbox: A Low Cost Bench Top Preclinical PET Scanner

PURPOSE: PETbox is a low cost bench top preclinical PET scanner dedicated to pharmacokinetic and pharmacodynamic mouse studies. A prototype system was developed at our institute, and this manuscript characterizes the performance of the prototype system. PROCEDURES: The PETbox detector consists of a...

Descripción completa

Detalles Bibliográficos
Autores principales: Zhang, Hui, Bao, Qinan, Vu, Nam T., Silverman, Robert W., Taschereau, Richard, Berry-Pusey, Brittany N., Douraghy, Ali, Rannou, Fernando R., Stout, David B., Chatziioannou, Arion F.
Formato: Texto
Lenguaje:English
Publicado: Springer-Verlag 2010
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3053444/
https://www.ncbi.nlm.nih.gov/pubmed/20812031
http://dx.doi.org/10.1007/s11307-010-0413-y
_version_ 1782199742385618944
author Zhang, Hui
Bao, Qinan
Vu, Nam T.
Silverman, Robert W.
Taschereau, Richard
Berry-Pusey, Brittany N.
Douraghy, Ali
Rannou, Fernando R.
Stout, David B.
Chatziioannou, Arion F.
author_facet Zhang, Hui
Bao, Qinan
Vu, Nam T.
Silverman, Robert W.
Taschereau, Richard
Berry-Pusey, Brittany N.
Douraghy, Ali
Rannou, Fernando R.
Stout, David B.
Chatziioannou, Arion F.
author_sort Zhang, Hui
collection PubMed
description PURPOSE: PETbox is a low cost bench top preclinical PET scanner dedicated to pharmacokinetic and pharmacodynamic mouse studies. A prototype system was developed at our institute, and this manuscript characterizes the performance of the prototype system. PROCEDURES: The PETbox detector consists of a 20 × 44 bismuth germanate crystal array with a thickness of 5 mm and cross-section size of 2.05 × 2.05 mm. Two such detectors are placed facing each other at a spacing of 5 cm, forming a dual-head geometry optimized for imaging mice. The detectors are kept stationary during the scan, making PETbox a limited angle tomography system. 3D images are reconstructed using a maximum likelihood and expectation maximization (ML–EM) method. The performance of the prototype system was characterized based on a modified set of the NEMA NU 4-2008 standards. RESULTS: In-plane image spatial resolution was measured to be an average of 1.53 mm full width at half maximum for coronal images and 2.65 mm for the anterior–posterior direction. The volumetric reconstructed resolution was below 8 mm(3) at most locations in the field of view (FOV). The sensitivity, scatter fraction, and noise equivalent count rate (NECR) were measured for different energy windows. With an energy window of 150 - 650 keV and a timing window of 20 ns optimized for mouse imaging, the peak absolute sensitivity was 3.99% at the center of FOV and a peak NECR of 20 kcps was achieved for a total activity of 3.2 MBq (86.8 μCi). Phantom and in vivo imaging studies were performed and demonstrated the utility of the system at low activity levels. The quantitation capabilities of the system were also characterized showing that despite the limited angle tomography, reasonably good quantification accuracy was achieved over a large dynamic range of activity levels. CONCLUSIONS: The presented results demonstrate the potential of this new tomograph for small animal imaging.
format Text
id pubmed-3053444
institution National Center for Biotechnology Information
language English
publishDate 2010
publisher Springer-Verlag
record_format MEDLINE/PubMed
spelling pubmed-30534442011-09-30 Performance Evaluation of PETbox: A Low Cost Bench Top Preclinical PET Scanner Zhang, Hui Bao, Qinan Vu, Nam T. Silverman, Robert W. Taschereau, Richard Berry-Pusey, Brittany N. Douraghy, Ali Rannou, Fernando R. Stout, David B. Chatziioannou, Arion F. Mol Imaging Biol Research Article PURPOSE: PETbox is a low cost bench top preclinical PET scanner dedicated to pharmacokinetic and pharmacodynamic mouse studies. A prototype system was developed at our institute, and this manuscript characterizes the performance of the prototype system. PROCEDURES: The PETbox detector consists of a 20 × 44 bismuth germanate crystal array with a thickness of 5 mm and cross-section size of 2.05 × 2.05 mm. Two such detectors are placed facing each other at a spacing of 5 cm, forming a dual-head geometry optimized for imaging mice. The detectors are kept stationary during the scan, making PETbox a limited angle tomography system. 3D images are reconstructed using a maximum likelihood and expectation maximization (ML–EM) method. The performance of the prototype system was characterized based on a modified set of the NEMA NU 4-2008 standards. RESULTS: In-plane image spatial resolution was measured to be an average of 1.53 mm full width at half maximum for coronal images and 2.65 mm for the anterior–posterior direction. The volumetric reconstructed resolution was below 8 mm(3) at most locations in the field of view (FOV). The sensitivity, scatter fraction, and noise equivalent count rate (NECR) were measured for different energy windows. With an energy window of 150 - 650 keV and a timing window of 20 ns optimized for mouse imaging, the peak absolute sensitivity was 3.99% at the center of FOV and a peak NECR of 20 kcps was achieved for a total activity of 3.2 MBq (86.8 μCi). Phantom and in vivo imaging studies were performed and demonstrated the utility of the system at low activity levels. The quantitation capabilities of the system were also characterized showing that despite the limited angle tomography, reasonably good quantification accuracy was achieved over a large dynamic range of activity levels. CONCLUSIONS: The presented results demonstrate the potential of this new tomograph for small animal imaging. Springer-Verlag 2010-09-02 2011 /pmc/articles/PMC3053444/ /pubmed/20812031 http://dx.doi.org/10.1007/s11307-010-0413-y Text en © The Author(s) 2010 https://creativecommons.org/licenses/by-nc/4.0/This article is distributed under the terms of the Creative Commons Attribution Non-Commercial License which permits any non-commercial use, distribution, and reproduction in any medium, provided the original author(s) and source are credited.
spellingShingle Research Article
Zhang, Hui
Bao, Qinan
Vu, Nam T.
Silverman, Robert W.
Taschereau, Richard
Berry-Pusey, Brittany N.
Douraghy, Ali
Rannou, Fernando R.
Stout, David B.
Chatziioannou, Arion F.
Performance Evaluation of PETbox: A Low Cost Bench Top Preclinical PET Scanner
title Performance Evaluation of PETbox: A Low Cost Bench Top Preclinical PET Scanner
title_full Performance Evaluation of PETbox: A Low Cost Bench Top Preclinical PET Scanner
title_fullStr Performance Evaluation of PETbox: A Low Cost Bench Top Preclinical PET Scanner
title_full_unstemmed Performance Evaluation of PETbox: A Low Cost Bench Top Preclinical PET Scanner
title_short Performance Evaluation of PETbox: A Low Cost Bench Top Preclinical PET Scanner
title_sort performance evaluation of petbox: a low cost bench top preclinical pet scanner
topic Research Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3053444/
https://www.ncbi.nlm.nih.gov/pubmed/20812031
http://dx.doi.org/10.1007/s11307-010-0413-y
work_keys_str_mv AT zhanghui performanceevaluationofpetboxalowcostbenchtoppreclinicalpetscanner
AT baoqinan performanceevaluationofpetboxalowcostbenchtoppreclinicalpetscanner
AT vunamt performanceevaluationofpetboxalowcostbenchtoppreclinicalpetscanner
AT silvermanrobertw performanceevaluationofpetboxalowcostbenchtoppreclinicalpetscanner
AT taschereaurichard performanceevaluationofpetboxalowcostbenchtoppreclinicalpetscanner
AT berrypuseybrittanyn performanceevaluationofpetboxalowcostbenchtoppreclinicalpetscanner
AT douraghyali performanceevaluationofpetboxalowcostbenchtoppreclinicalpetscanner
AT rannoufernandor performanceevaluationofpetboxalowcostbenchtoppreclinicalpetscanner
AT stoutdavidb performanceevaluationofpetboxalowcostbenchtoppreclinicalpetscanner
AT chatziioannouarionf performanceevaluationofpetboxalowcostbenchtoppreclinicalpetscanner