Patient-Specific Unicompartmental Knee Resurfacing Arthroplasty: Use of a Novel Interference Lock to Reduce Tibial Insert Micromotion and Backside Wear

Micromotion has long been associated with wear of polyethylene tibial inserts, potentially causing failure of unicompartmental knee replacement systems. One cause of micromotion is the locking mechanism between the undersurface of the polyethylene and the tibial tray. The objective of this study was...

Descripción completa

Detalles Bibliográficos
Autores principales: Steklov, Nick, Chao, Nam, Srivastav, Sudesh
Formato: Texto
Lenguaje:English
Publicado: Bentham Open 2010
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3069356/
https://www.ncbi.nlm.nih.gov/pubmed/21464910
http://dx.doi.org/10.2174/1874120701004010156
_version_ 1782201339419295744
author Steklov, Nick
Chao, Nam
Srivastav, Sudesh
author_facet Steklov, Nick
Chao, Nam
Srivastav, Sudesh
author_sort Steklov, Nick
collection PubMed
description Micromotion has long been associated with wear of polyethylene tibial inserts, potentially causing failure of unicompartmental knee replacement systems. One cause of micromotion is the locking mechanism between the undersurface of the polyethylene and the tibial tray. The objective of this study was to investigate the use of new novel lock designs for reducing the micromotion associated with a patient specific tibial implant. Micromotion occurring between the tibial insert and tibial tray was measured using the DVRT method for two new lock designs and compared to the micromotion measured for the prior generation lock design. In total, 18 samples were tested, six in each of the three designs: prior generation CLEARANCE lock design, new LINE-TO-LINE design, new INTERFERENCE design. At the lowest loading of 100N, the motion index of the new interference lock design (4.9 ±2.6 microns) was less than the motion index of both the new line-to-line lock design (11.9±4.5 microns) and the prior generation clearance lock design (154.9±84.4 microns). Likewise, at the higher loading of 400N, the motion index of the new interference lock design (58.7 ±7.2 microns) was again less than the motion index of the two alternative lock designs (new line-to-line: 135.5±57.6 microns; prior generation clearance: 371.8±151.4 microns). When compared to previously published micromotion data for TKR systems, the results demonstrate that a novel interference design locking mechanism as part of a patient-specific unicompartmental knee arthroplasty system can significantly reduce micromotion that is believed to be one of the causes of osteolysis.
format Text
id pubmed-3069356
institution National Center for Biotechnology Information
language English
publishDate 2010
publisher Bentham Open
record_format MEDLINE/PubMed
spelling pubmed-30693562011-04-04 Patient-Specific Unicompartmental Knee Resurfacing Arthroplasty: Use of a Novel Interference Lock to Reduce Tibial Insert Micromotion and Backside Wear Steklov, Nick Chao, Nam Srivastav, Sudesh Open Biomed Eng J Article Micromotion has long been associated with wear of polyethylene tibial inserts, potentially causing failure of unicompartmental knee replacement systems. One cause of micromotion is the locking mechanism between the undersurface of the polyethylene and the tibial tray. The objective of this study was to investigate the use of new novel lock designs for reducing the micromotion associated with a patient specific tibial implant. Micromotion occurring between the tibial insert and tibial tray was measured using the DVRT method for two new lock designs and compared to the micromotion measured for the prior generation lock design. In total, 18 samples were tested, six in each of the three designs: prior generation CLEARANCE lock design, new LINE-TO-LINE design, new INTERFERENCE design. At the lowest loading of 100N, the motion index of the new interference lock design (4.9 ±2.6 microns) was less than the motion index of both the new line-to-line lock design (11.9±4.5 microns) and the prior generation clearance lock design (154.9±84.4 microns). Likewise, at the higher loading of 400N, the motion index of the new interference lock design (58.7 ±7.2 microns) was again less than the motion index of the two alternative lock designs (new line-to-line: 135.5±57.6 microns; prior generation clearance: 371.8±151.4 microns). When compared to previously published micromotion data for TKR systems, the results demonstrate that a novel interference design locking mechanism as part of a patient-specific unicompartmental knee arthroplasty system can significantly reduce micromotion that is believed to be one of the causes of osteolysis. Bentham Open 2010-08-17 /pmc/articles/PMC3069356/ /pubmed/21464910 http://dx.doi.org/10.2174/1874120701004010156 Text en © Steklov et al.; Licensee Bentham Open. http://creativecommons.org/licenses/by-nc/3.0/ This is an open access article licensed under the terms of the Creative Commons Attribution Non-Commercial License (http://creativecommons.org/licenses/by-nc/3.0/) which permits unrestricted, non-commercial use, distribution and reproduction in any medium, provided the work is properly cited.
spellingShingle Article
Steklov, Nick
Chao, Nam
Srivastav, Sudesh
Patient-Specific Unicompartmental Knee Resurfacing Arthroplasty: Use of a Novel Interference Lock to Reduce Tibial Insert Micromotion and Backside Wear
title Patient-Specific Unicompartmental Knee Resurfacing Arthroplasty: Use of a Novel Interference Lock to Reduce Tibial Insert Micromotion and Backside Wear
title_full Patient-Specific Unicompartmental Knee Resurfacing Arthroplasty: Use of a Novel Interference Lock to Reduce Tibial Insert Micromotion and Backside Wear
title_fullStr Patient-Specific Unicompartmental Knee Resurfacing Arthroplasty: Use of a Novel Interference Lock to Reduce Tibial Insert Micromotion and Backside Wear
title_full_unstemmed Patient-Specific Unicompartmental Knee Resurfacing Arthroplasty: Use of a Novel Interference Lock to Reduce Tibial Insert Micromotion and Backside Wear
title_short Patient-Specific Unicompartmental Knee Resurfacing Arthroplasty: Use of a Novel Interference Lock to Reduce Tibial Insert Micromotion and Backside Wear
title_sort patient-specific unicompartmental knee resurfacing arthroplasty: use of a novel interference lock to reduce tibial insert micromotion and backside wear
topic Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3069356/
https://www.ncbi.nlm.nih.gov/pubmed/21464910
http://dx.doi.org/10.2174/1874120701004010156
work_keys_str_mv AT steklovnick patientspecificunicompartmentalkneeresurfacingarthroplastyuseofanovelinterferencelocktoreducetibialinsertmicromotionandbacksidewear
AT chaonam patientspecificunicompartmentalkneeresurfacingarthroplastyuseofanovelinterferencelocktoreducetibialinsertmicromotionandbacksidewear
AT srivastavsudesh patientspecificunicompartmentalkneeresurfacingarthroplastyuseofanovelinterferencelocktoreducetibialinsertmicromotionandbacksidewear