Image Artifacts in Concurrent Transcranial Magnetic Stimulation (TMS) and fMRI Caused by Leakage Currents: Modeling and Compensation

PURPOSE: To characterize and eliminate a new type of image artifact in concurrent transcranial magnetic stimulation and functional MRI (TMS-fMRI) caused by small leakage currents originating from the high-voltage capacitors in the TMS stimulator system. MATERIALS AND METHODS: The artifacts in echo-p...

Descripción completa

Detalles Bibliográficos
Autores principales: Weiskopf, Nikolaus, Josephs, Oliver, Ruff, Christian C, Blankenburg, Felix, Featherstone, Eric, Thomas, Anthony, Bestmann, Sven, Driver, Jon, Deichmann, Ralf
Formato: Texto
Lenguaje:English
Publicado: Wiley Subscription Services, Inc., A Wiley Company 2009
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3077517/
https://www.ncbi.nlm.nih.gov/pubmed/19388099
http://dx.doi.org/10.1002/jmri.21749
Descripción
Sumario:PURPOSE: To characterize and eliminate a new type of image artifact in concurrent transcranial magnetic stimulation and functional MRI (TMS-fMRI) caused by small leakage currents originating from the high-voltage capacitors in the TMS stimulator system. MATERIALS AND METHODS: The artifacts in echo-planar images (EPI) caused by leakage currents were characterized and quantified in numerical simulations and phantom studies with different phantom-coil geometries. A relay-diode combination was devised and inserted in the TMS circuit that shorts the leakage current. Its effectiveness for artifact reduction was assessed in a phantom scan resembling a realistic TMS-fMRI experiment. RESULTS: The leakage-current-induced signal changes exhibited a multipolar spatial pattern and the maxima exceeded 1% at realistic coil-cortex distances. The relay-diode combination effectively reduced the artifact to a negligible level. CONCLUSION: The leakage-current artifacts potentially obscure effects of interest or lead to false-positives. Since the artifact depends on the experimental setup and design (eg, amplitude of the leakage current, coil orientation, paradigm, EPI parameters), we recommend its assessment for each experiment. The relay-diode combination can eliminate the artifacts if necessary. J. Magn. Reson. Imaging 2009;29:1211–1217. © 2009 Wiley-Liss, Inc.