Hemoglobin estimation by the HemoCue(® )portable hemoglobin photometer in a resource poor setting

BACKGROUND: In resource poor settings where automated hematology analyzers are not available, the Cyanmethemoglobin method is often used. This method though cheaper, takes more time. In blood donations, the semi-quantitative gravimetric copper sulfate method which is very easy and inexpensive may be...

Descripción completa

Detalles Bibliográficos
Autores principales: Nkrumah, Bernard, Nguah, Samuel Blay, Sarpong, Nimako, Dekker, Denise, Idriss, Ali, May, Juergen, Adu-Sarkodie, Yaw
Formato: Texto
Lenguaje:English
Publicado: BioMed Central 2011
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3095531/
https://www.ncbi.nlm.nih.gov/pubmed/21510885
http://dx.doi.org/10.1186/1472-6890-11-5
_version_ 1782203657808248832
author Nkrumah, Bernard
Nguah, Samuel Blay
Sarpong, Nimako
Dekker, Denise
Idriss, Ali
May, Juergen
Adu-Sarkodie, Yaw
author_facet Nkrumah, Bernard
Nguah, Samuel Blay
Sarpong, Nimako
Dekker, Denise
Idriss, Ali
May, Juergen
Adu-Sarkodie, Yaw
author_sort Nkrumah, Bernard
collection PubMed
description BACKGROUND: In resource poor settings where automated hematology analyzers are not available, the Cyanmethemoglobin method is often used. This method though cheaper, takes more time. In blood donations, the semi-quantitative gravimetric copper sulfate method which is very easy and inexpensive may be used but does not provide an acceptable degree of accuracy. The HemoCue(® )hemoglobin photometer has been used for these purposes. This study was conducted to generate data to support or refute its use as a point-of-care device for hemoglobin estimation in mobile blood donations and critical care areas in health facilities. METHOD: EDTA blood was collected from study participants drawn from five groups: pre-school children, school children, pregnant women, non-pregnant women and men. Blood collected was immediately processed to estimate the hemoglobin concentration using three different methods (HemoCue(®), Sysmex KX21N and Cyanmethemoglobin). Agreement between the test methods was assessed by the method of Bland and Altman. The Intraclass correlation coefficient (ICC) was used to determine the within subject variability of measured hemoglobin. RESULTS: Of 398 subjects, 42% were males with the overall mean age being 19.4 years. The overall mean hemoglobin as estimated by each method was 10.4 g/dl for HemoCue, 10.3 g/dl for Sysmex KX21N and 10.3 g/dl for Cyanmethemoglobin. Pairwise analysis revealed that the hemoglobin determined by the HemoCue method was higher than that measured by the KX21N and Cyanmethemoglobin. Comparing the hemoglobin determined by the HemoCue to Cyanmethemoglobin, the concordance correlation coefficient was 0.995 (95% CI: 0.994-0.996, p < 0.001). The Bland and Altman's limit of agreement was -0.389 - 0.644 g/dl with the mean difference being 0.127 (95% CI: 0.102-0.153) and a non-significant difference in variability between the two measurements (p = 0.843). After adjusting to assess the effect of other possible confounders such as sex, age and category of person, there was no significant difference in the hemoglobin determined by the HemoCue compared to Cyanmethemoglobin (coef = -0.127, 95% CI: -0.379 - 0.634). CONCLUSION: Hemoglobin determined by the HemoCue method is comparable to that determined by the other methods. The HemoCue photometer is therefore recommended for use as on-the-spot device for determining hemoglobin in resource poor setting.
format Text
id pubmed-3095531
institution National Center for Biotechnology Information
language English
publishDate 2011
publisher BioMed Central
record_format MEDLINE/PubMed
spelling pubmed-30955312011-05-17 Hemoglobin estimation by the HemoCue(® )portable hemoglobin photometer in a resource poor setting Nkrumah, Bernard Nguah, Samuel Blay Sarpong, Nimako Dekker, Denise Idriss, Ali May, Juergen Adu-Sarkodie, Yaw BMC Clin Pathol Research Article BACKGROUND: In resource poor settings where automated hematology analyzers are not available, the Cyanmethemoglobin method is often used. This method though cheaper, takes more time. In blood donations, the semi-quantitative gravimetric copper sulfate method which is very easy and inexpensive may be used but does not provide an acceptable degree of accuracy. The HemoCue(® )hemoglobin photometer has been used for these purposes. This study was conducted to generate data to support or refute its use as a point-of-care device for hemoglobin estimation in mobile blood donations and critical care areas in health facilities. METHOD: EDTA blood was collected from study participants drawn from five groups: pre-school children, school children, pregnant women, non-pregnant women and men. Blood collected was immediately processed to estimate the hemoglobin concentration using three different methods (HemoCue(®), Sysmex KX21N and Cyanmethemoglobin). Agreement between the test methods was assessed by the method of Bland and Altman. The Intraclass correlation coefficient (ICC) was used to determine the within subject variability of measured hemoglobin. RESULTS: Of 398 subjects, 42% were males with the overall mean age being 19.4 years. The overall mean hemoglobin as estimated by each method was 10.4 g/dl for HemoCue, 10.3 g/dl for Sysmex KX21N and 10.3 g/dl for Cyanmethemoglobin. Pairwise analysis revealed that the hemoglobin determined by the HemoCue method was higher than that measured by the KX21N and Cyanmethemoglobin. Comparing the hemoglobin determined by the HemoCue to Cyanmethemoglobin, the concordance correlation coefficient was 0.995 (95% CI: 0.994-0.996, p < 0.001). The Bland and Altman's limit of agreement was -0.389 - 0.644 g/dl with the mean difference being 0.127 (95% CI: 0.102-0.153) and a non-significant difference in variability between the two measurements (p = 0.843). After adjusting to assess the effect of other possible confounders such as sex, age and category of person, there was no significant difference in the hemoglobin determined by the HemoCue compared to Cyanmethemoglobin (coef = -0.127, 95% CI: -0.379 - 0.634). CONCLUSION: Hemoglobin determined by the HemoCue method is comparable to that determined by the other methods. The HemoCue photometer is therefore recommended for use as on-the-spot device for determining hemoglobin in resource poor setting. BioMed Central 2011-04-21 /pmc/articles/PMC3095531/ /pubmed/21510885 http://dx.doi.org/10.1186/1472-6890-11-5 Text en Copyright ©2011 Nkrumah et al; licensee BioMed Central Ltd. http://creativecommons.org/licenses/by/2.0 This is an Open Access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/2.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
spellingShingle Research Article
Nkrumah, Bernard
Nguah, Samuel Blay
Sarpong, Nimako
Dekker, Denise
Idriss, Ali
May, Juergen
Adu-Sarkodie, Yaw
Hemoglobin estimation by the HemoCue(® )portable hemoglobin photometer in a resource poor setting
title Hemoglobin estimation by the HemoCue(® )portable hemoglobin photometer in a resource poor setting
title_full Hemoglobin estimation by the HemoCue(® )portable hemoglobin photometer in a resource poor setting
title_fullStr Hemoglobin estimation by the HemoCue(® )portable hemoglobin photometer in a resource poor setting
title_full_unstemmed Hemoglobin estimation by the HemoCue(® )portable hemoglobin photometer in a resource poor setting
title_short Hemoglobin estimation by the HemoCue(® )portable hemoglobin photometer in a resource poor setting
title_sort hemoglobin estimation by the hemocue(® )portable hemoglobin photometer in a resource poor setting
topic Research Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3095531/
https://www.ncbi.nlm.nih.gov/pubmed/21510885
http://dx.doi.org/10.1186/1472-6890-11-5
work_keys_str_mv AT nkrumahbernard hemoglobinestimationbythehemocueportablehemoglobinphotometerinaresourcepoorsetting
AT nguahsamuelblay hemoglobinestimationbythehemocueportablehemoglobinphotometerinaresourcepoorsetting
AT sarpongnimako hemoglobinestimationbythehemocueportablehemoglobinphotometerinaresourcepoorsetting
AT dekkerdenise hemoglobinestimationbythehemocueportablehemoglobinphotometerinaresourcepoorsetting
AT idrissali hemoglobinestimationbythehemocueportablehemoglobinphotometerinaresourcepoorsetting
AT mayjuergen hemoglobinestimationbythehemocueportablehemoglobinphotometerinaresourcepoorsetting
AT adusarkodieyaw hemoglobinestimationbythehemocueportablehemoglobinphotometerinaresourcepoorsetting