Electrolytes assessed by point-of-care testing – Are the values comparable with results obtained from the central laboratory?

BACKGROUND AND AIMS: When dealing with very sick patients, the speed and accuracy of tests to detect metabolic derangements is very important. We evaluated if there was agreement between whole blood electrolytes measured by a point-of-care device and serum electrolytes measured using indirect ion-se...

Descripción completa

Detalles Bibliográficos
Autores principales: Chacko, Binila, Peter, John V, Patole, Shalom, Fleming, Jude J, Selvakumar, Ratnasamy
Formato: Texto
Lenguaje:English
Publicado: Medknow Publications 2011
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3097538/
https://www.ncbi.nlm.nih.gov/pubmed/21633542
http://dx.doi.org/10.4103/0972-5229.78219
_version_ 1782203833052561408
author Chacko, Binila
Peter, John V
Patole, Shalom
Fleming, Jude J
Selvakumar, Ratnasamy
author_facet Chacko, Binila
Peter, John V
Patole, Shalom
Fleming, Jude J
Selvakumar, Ratnasamy
author_sort Chacko, Binila
collection PubMed
description BACKGROUND AND AIMS: When dealing with very sick patients, the speed and accuracy of tests to detect metabolic derangements is very important. We evaluated if there was agreement between whole blood electrolytes measured by a point-of-care device and serum electrolytes measured using indirect ion-selective electrodes. MATERIALS AND METHODS: In this prospective study, electrolytes were analyzed in 44 paired samples drawn from critically ill patients. Whole blood electrolytes were analyzed using a point-of-care blood gas analyzer and serum electrolytes were analyzed in the central laboratory on samples transported through a rapid transit pneumatic system. Agreement was summarized by the mean difference with 95% limits of agreement (LOA) and Lin’s concordance correlation (p (c)). RESULTS: There was a significant difference in the mean (±standard deviation) sodium value between whole blood and serum samples (135.8 ± 5.7 mmol/L vs. 139.9 ± 5.4 mmol/L, P < 0.001), with the agreement being modest (p(c) = 0.71; mean difference −4.0; 95% LOA −8.78 to 0.65). Although the agreement between whole blood and serum potassium was good (p(c) = 0.96), and the average difference small (−0.3; 95% LOA −0.72 to 0.13), individual differences were clinically significant, particularly at lower potassium values. For potassium values <3.0 mmol/L, the concordance was low (p(c) = 0.53) and the LOA was wide (1.0 to −0.13). The concordance for potassium was good (p(c) = 0.96) for values ≥3.0 (mean difference −0.2; 95% LOA −0.48 to 0.06). CONCLUSIONS: Clinicians should be aware of the difference between whole blood and serum electrolytes, particularly when urgent samples are tested at point of care and routine follow-up electrolytes are sent to the central laboratory. A correction factor needs to be determined at each center.
format Text
id pubmed-3097538
institution National Center for Biotechnology Information
language English
publishDate 2011
publisher Medknow Publications
record_format MEDLINE/PubMed
spelling pubmed-30975382011-06-01 Electrolytes assessed by point-of-care testing – Are the values comparable with results obtained from the central laboratory? Chacko, Binila Peter, John V Patole, Shalom Fleming, Jude J Selvakumar, Ratnasamy Indian J Crit Care Med Research Article BACKGROUND AND AIMS: When dealing with very sick patients, the speed and accuracy of tests to detect metabolic derangements is very important. We evaluated if there was agreement between whole blood electrolytes measured by a point-of-care device and serum electrolytes measured using indirect ion-selective electrodes. MATERIALS AND METHODS: In this prospective study, electrolytes were analyzed in 44 paired samples drawn from critically ill patients. Whole blood electrolytes were analyzed using a point-of-care blood gas analyzer and serum electrolytes were analyzed in the central laboratory on samples transported through a rapid transit pneumatic system. Agreement was summarized by the mean difference with 95% limits of agreement (LOA) and Lin’s concordance correlation (p (c)). RESULTS: There was a significant difference in the mean (±standard deviation) sodium value between whole blood and serum samples (135.8 ± 5.7 mmol/L vs. 139.9 ± 5.4 mmol/L, P < 0.001), with the agreement being modest (p(c) = 0.71; mean difference −4.0; 95% LOA −8.78 to 0.65). Although the agreement between whole blood and serum potassium was good (p(c) = 0.96), and the average difference small (−0.3; 95% LOA −0.72 to 0.13), individual differences were clinically significant, particularly at lower potassium values. For potassium values <3.0 mmol/L, the concordance was low (p(c) = 0.53) and the LOA was wide (1.0 to −0.13). The concordance for potassium was good (p(c) = 0.96) for values ≥3.0 (mean difference −0.2; 95% LOA −0.48 to 0.06). CONCLUSIONS: Clinicians should be aware of the difference between whole blood and serum electrolytes, particularly when urgent samples are tested at point of care and routine follow-up electrolytes are sent to the central laboratory. A correction factor needs to be determined at each center. Medknow Publications 2011 /pmc/articles/PMC3097538/ /pubmed/21633542 http://dx.doi.org/10.4103/0972-5229.78219 Text en © Indian Journal of Critical Care Medicine http://creativecommons.org/licenses/by/2.0/ This is an open-access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
spellingShingle Research Article
Chacko, Binila
Peter, John V
Patole, Shalom
Fleming, Jude J
Selvakumar, Ratnasamy
Electrolytes assessed by point-of-care testing – Are the values comparable with results obtained from the central laboratory?
title Electrolytes assessed by point-of-care testing – Are the values comparable with results obtained from the central laboratory?
title_full Electrolytes assessed by point-of-care testing – Are the values comparable with results obtained from the central laboratory?
title_fullStr Electrolytes assessed by point-of-care testing – Are the values comparable with results obtained from the central laboratory?
title_full_unstemmed Electrolytes assessed by point-of-care testing – Are the values comparable with results obtained from the central laboratory?
title_short Electrolytes assessed by point-of-care testing – Are the values comparable with results obtained from the central laboratory?
title_sort electrolytes assessed by point-of-care testing – are the values comparable with results obtained from the central laboratory?
topic Research Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3097538/
https://www.ncbi.nlm.nih.gov/pubmed/21633542
http://dx.doi.org/10.4103/0972-5229.78219
work_keys_str_mv AT chackobinila electrolytesassessedbypointofcaretestingarethevaluescomparablewithresultsobtainedfromthecentrallaboratory
AT peterjohnv electrolytesassessedbypointofcaretestingarethevaluescomparablewithresultsobtainedfromthecentrallaboratory
AT patoleshalom electrolytesassessedbypointofcaretestingarethevaluescomparablewithresultsobtainedfromthecentrallaboratory
AT flemingjudej electrolytesassessedbypointofcaretestingarethevaluescomparablewithresultsobtainedfromthecentrallaboratory
AT selvakumarratnasamy electrolytesassessedbypointofcaretestingarethevaluescomparablewithresultsobtainedfromthecentrallaboratory