Measurements of patient’s setup variation in intensity-modulated radiation therapy of head and neck cancer using electronic portal imaging device

PURPOSE: To measure the interfraction setup variation of patient undergoing intensity-modulated radiation therapy (IMRT) of head and neck cancer. The data was used to define adequate treatment CTV-to-PTV margin. MATERIALS AND METHODS: During March to September 2006, data was collected from 9 head an...

Descripción completa

Detalles Bibliográficos
Autores principales: Naiyanet, N, Oonsiri, S, Lertbutsayanukul, C, Suriyapee, S
Formato: Texto
Lenguaje:English
Publicado: Department of Biomedical Imaging, Faculty of Medicine, University of Malaya, Malaysia 2007
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3097651/
https://www.ncbi.nlm.nih.gov/pubmed/21614263
http://dx.doi.org/10.2349/biij.3.1.e30
_version_ 1782203859690586112
author Naiyanet, N
Oonsiri, S
Lertbutsayanukul, C
Suriyapee, S
author_facet Naiyanet, N
Oonsiri, S
Lertbutsayanukul, C
Suriyapee, S
author_sort Naiyanet, N
collection PubMed
description PURPOSE: To measure the interfraction setup variation of patient undergoing intensity-modulated radiation therapy (IMRT) of head and neck cancer. The data was used to define adequate treatment CTV-to-PTV margin. MATERIALS AND METHODS: During March to September 2006, data was collected from 9 head and neck cancer patients treated with dynamic IMRT using 6 MV X-ray beam from Varian Clinac 23EX. Weekly portal images of setup fields which were anterior-posterior and lateral portal images were acquired for each patient with an amorphous silicon EPID, Varian aS500. These images were matched with the reference image from Varian Acuity simulator using the Varis vision software (Version 7.3.10). Six anatomical landmarks were selected for comparison. The displacement of portal image from the reference image was recorded in X (Left-Right, L-R), Y (Superior-Inferior, S-I) direction for anterior field and Z (Anterior-Posterior, A-P), Y (S-I) direction for lateral field. The systematic and random error for individual and population were calculated. Then the population-based margins were obtained. RESULTS: A total of 135 images (27 simulation images and 108 portal images) and 405 match points was evaluated. The systematic error ranged from 0 to 7.5 mm and the random error ranged from 0.3 to 4.8 mm for all directions. The population-based margin ranged from 2.3 to 4.5 mm (L-R), 3.5 to 4.9 mm (S-I) for anterior field and 3.4 to 4.7 mm (A-P), 2.6 to 3.7 mm (S-I) for the lateral field. These margins were comparable to the margin that was prescribed at the King Chulalongkorn Memorial Hospital (5-10 mm) for head and neck cancer. CONCLUSION: The population-based margin is less than 5 mm, thus the margin provides sufficient coverage for all of the patients.
format Text
id pubmed-3097651
institution National Center for Biotechnology Information
language English
publishDate 2007
publisher Department of Biomedical Imaging, Faculty of Medicine, University of Malaya, Malaysia
record_format MEDLINE/PubMed
spelling pubmed-30976512011-05-24 Measurements of patient’s setup variation in intensity-modulated radiation therapy of head and neck cancer using electronic portal imaging device Naiyanet, N Oonsiri, S Lertbutsayanukul, C Suriyapee, S Biomed Imaging Interv J Original Article PURPOSE: To measure the interfraction setup variation of patient undergoing intensity-modulated radiation therapy (IMRT) of head and neck cancer. The data was used to define adequate treatment CTV-to-PTV margin. MATERIALS AND METHODS: During March to September 2006, data was collected from 9 head and neck cancer patients treated with dynamic IMRT using 6 MV X-ray beam from Varian Clinac 23EX. Weekly portal images of setup fields which were anterior-posterior and lateral portal images were acquired for each patient with an amorphous silicon EPID, Varian aS500. These images were matched with the reference image from Varian Acuity simulator using the Varis vision software (Version 7.3.10). Six anatomical landmarks were selected for comparison. The displacement of portal image from the reference image was recorded in X (Left-Right, L-R), Y (Superior-Inferior, S-I) direction for anterior field and Z (Anterior-Posterior, A-P), Y (S-I) direction for lateral field. The systematic and random error for individual and population were calculated. Then the population-based margins were obtained. RESULTS: A total of 135 images (27 simulation images and 108 portal images) and 405 match points was evaluated. The systematic error ranged from 0 to 7.5 mm and the random error ranged from 0.3 to 4.8 mm for all directions. The population-based margin ranged from 2.3 to 4.5 mm (L-R), 3.5 to 4.9 mm (S-I) for anterior field and 3.4 to 4.7 mm (A-P), 2.6 to 3.7 mm (S-I) for the lateral field. These margins were comparable to the margin that was prescribed at the King Chulalongkorn Memorial Hospital (5-10 mm) for head and neck cancer. CONCLUSION: The population-based margin is less than 5 mm, thus the margin provides sufficient coverage for all of the patients. Department of Biomedical Imaging, Faculty of Medicine, University of Malaya, Malaysia 2007-01-01 /pmc/articles/PMC3097651/ /pubmed/21614263 http://dx.doi.org/10.2349/biij.3.1.e30 Text en © 2007 Biomedical Imaging and Intervention Journal http://creativecommons.org/licenses/by/2.5/ This is an open-access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
spellingShingle Original Article
Naiyanet, N
Oonsiri, S
Lertbutsayanukul, C
Suriyapee, S
Measurements of patient’s setup variation in intensity-modulated radiation therapy of head and neck cancer using electronic portal imaging device
title Measurements of patient’s setup variation in intensity-modulated radiation therapy of head and neck cancer using electronic portal imaging device
title_full Measurements of patient’s setup variation in intensity-modulated radiation therapy of head and neck cancer using electronic portal imaging device
title_fullStr Measurements of patient’s setup variation in intensity-modulated radiation therapy of head and neck cancer using electronic portal imaging device
title_full_unstemmed Measurements of patient’s setup variation in intensity-modulated radiation therapy of head and neck cancer using electronic portal imaging device
title_short Measurements of patient’s setup variation in intensity-modulated radiation therapy of head and neck cancer using electronic portal imaging device
title_sort measurements of patient’s setup variation in intensity-modulated radiation therapy of head and neck cancer using electronic portal imaging device
topic Original Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3097651/
https://www.ncbi.nlm.nih.gov/pubmed/21614263
http://dx.doi.org/10.2349/biij.3.1.e30
work_keys_str_mv AT naiyanetn measurementsofpatientssetupvariationinintensitymodulatedradiationtherapyofheadandneckcancerusingelectronicportalimagingdevice
AT oonsiris measurementsofpatientssetupvariationinintensitymodulatedradiationtherapyofheadandneckcancerusingelectronicportalimagingdevice
AT lertbutsayanukulc measurementsofpatientssetupvariationinintensitymodulatedradiationtherapyofheadandneckcancerusingelectronicportalimagingdevice
AT suriyapees measurementsofpatientssetupvariationinintensitymodulatedradiationtherapyofheadandneckcancerusingelectronicportalimagingdevice