Automatic polyp detection and measurement with computed tomographic colonography: A phantom study

PURPOSE: The purpose of this study is to assess the performance of computer-aided detection (CAD) software in detecting and measuring polyps for CT Colonography, based on an in vitro phantom study. MATERIAL AND METHODS: A colon phantom was constructed with a PVC pipe of 3.8 cm diameter. Nine simulat...

Descripción completa

Detalles Bibliográficos
Autores principales: Virmani, S, Lev-Toaff, AS, Ciancibello, LM
Formato: Texto
Lenguaje:English
Publicado: Department of Biomedical Imaging, Faculty of Medicine, University of Malaya, Malaysia 2009
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3097787/
https://www.ncbi.nlm.nih.gov/pubmed/21611052
http://dx.doi.org/10.2349/biij.5.3.e15
_version_ 1782203891939540992
author Virmani, S
Lev-Toaff, AS
Ciancibello, LM
author_facet Virmani, S
Lev-Toaff, AS
Ciancibello, LM
author_sort Virmani, S
collection PubMed
description PURPOSE: The purpose of this study is to assess the performance of computer-aided detection (CAD) software in detecting and measuring polyps for CT Colonography, based on an in vitro phantom study. MATERIAL AND METHODS: A colon phantom was constructed with a PVC pipe of 3.8 cm diameter. Nine simulated polyps of various sizes (3.2mm-25.4mm) were affixed inside the phantom that was placed in a water bath. The phantom was scanned on a 64-slice CT scanner with tube voltage of 120 kV and current of 205 mAs. Two separate scans were performed, with different slice thickness and reconstruction interval. The first scan (thin) had a slice thickness of 1mm and reconstruction interval 0.5mm. The second scan (thick) had a slice thickness of 2mm and reconstruction interval of 1mm. Images from both scans were processed using CT Colonography software that automatically segments the colon phantom and applies CAD that automatically highlights and provides the size (maximum and minimum diameters, volume) of each polyp. Two readers independently measured each polyp (two orthogonal diameters) using both 2D and 3D views. Readers’ manual measurements (diameters) and automatic measurements from CAD (diameters and volume) were compared to actual polyp sizes as measured by mechanical calipers. RESULTS: All polyps except the smallest (3.2mm) were detected by CAD. CAD achieved 100% sensitivity in detecting polyps ≥6mm. Mean errors in CAD automated volume measurements for thin and thick slice scans were 8.7% and 6.8%, respectively. Almost all CAD and manual readers’ 3D measurements overestimated the size of polyps to variable extent. Both over- and underestimation of polyp sizes were observed in the readers’ manual 2D measurements. Overall, Reader 1 (expert) had smaller mean error than Reader 2 (non-expert). CONCLUSION: CAD provided accurate size measurements for all polyps, and results were comparable to the two readers' manual measurements
format Text
id pubmed-3097787
institution National Center for Biotechnology Information
language English
publishDate 2009
publisher Department of Biomedical Imaging, Faculty of Medicine, University of Malaya, Malaysia
record_format MEDLINE/PubMed
spelling pubmed-30977872011-05-24 Automatic polyp detection and measurement with computed tomographic colonography: A phantom study Virmani, S Lev-Toaff, AS Ciancibello, LM Biomed Imaging Interv J Original Article PURPOSE: The purpose of this study is to assess the performance of computer-aided detection (CAD) software in detecting and measuring polyps for CT Colonography, based on an in vitro phantom study. MATERIAL AND METHODS: A colon phantom was constructed with a PVC pipe of 3.8 cm diameter. Nine simulated polyps of various sizes (3.2mm-25.4mm) were affixed inside the phantom that was placed in a water bath. The phantom was scanned on a 64-slice CT scanner with tube voltage of 120 kV and current of 205 mAs. Two separate scans were performed, with different slice thickness and reconstruction interval. The first scan (thin) had a slice thickness of 1mm and reconstruction interval 0.5mm. The second scan (thick) had a slice thickness of 2mm and reconstruction interval of 1mm. Images from both scans were processed using CT Colonography software that automatically segments the colon phantom and applies CAD that automatically highlights and provides the size (maximum and minimum diameters, volume) of each polyp. Two readers independently measured each polyp (two orthogonal diameters) using both 2D and 3D views. Readers’ manual measurements (diameters) and automatic measurements from CAD (diameters and volume) were compared to actual polyp sizes as measured by mechanical calipers. RESULTS: All polyps except the smallest (3.2mm) were detected by CAD. CAD achieved 100% sensitivity in detecting polyps ≥6mm. Mean errors in CAD automated volume measurements for thin and thick slice scans were 8.7% and 6.8%, respectively. Almost all CAD and manual readers’ 3D measurements overestimated the size of polyps to variable extent. Both over- and underestimation of polyp sizes were observed in the readers’ manual 2D measurements. Overall, Reader 1 (expert) had smaller mean error than Reader 2 (non-expert). CONCLUSION: CAD provided accurate size measurements for all polyps, and results were comparable to the two readers' manual measurements Department of Biomedical Imaging, Faculty of Medicine, University of Malaya, Malaysia 2009-07-01 /pmc/articles/PMC3097787/ /pubmed/21611052 http://dx.doi.org/10.2349/biij.5.3.e15 Text en © 2009 Biomedical Imaging and Intervention Journal http://creativecommons.org/licenses/by/2.5/ This is an open-access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
spellingShingle Original Article
Virmani, S
Lev-Toaff, AS
Ciancibello, LM
Automatic polyp detection and measurement with computed tomographic colonography: A phantom study
title Automatic polyp detection and measurement with computed tomographic colonography: A phantom study
title_full Automatic polyp detection and measurement with computed tomographic colonography: A phantom study
title_fullStr Automatic polyp detection and measurement with computed tomographic colonography: A phantom study
title_full_unstemmed Automatic polyp detection and measurement with computed tomographic colonography: A phantom study
title_short Automatic polyp detection and measurement with computed tomographic colonography: A phantom study
title_sort automatic polyp detection and measurement with computed tomographic colonography: a phantom study
topic Original Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3097787/
https://www.ncbi.nlm.nih.gov/pubmed/21611052
http://dx.doi.org/10.2349/biij.5.3.e15
work_keys_str_mv AT virmanis automaticpolypdetectionandmeasurementwithcomputedtomographiccolonographyaphantomstudy
AT levtoaffas automaticpolypdetectionandmeasurementwithcomputedtomographiccolonographyaphantomstudy
AT ciancibellolm automaticpolypdetectionandmeasurementwithcomputedtomographiccolonographyaphantomstudy