Continuous Non-Invasive Monitoring of Tidal Volumes by Measurement of Tidal Impedance in Neonatal Piglets
BACKGROUND: Electrical Impedance measurements can be used to estimate the content of intra-thoracic air and thereby give information on pulmonary ventilation. Conventional Impedance measurements mainly indicate relative changes, but no information concerning air-volume is given. The study was perfor...
Autores principales: | , , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Public Library of Science
2011
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3110249/ https://www.ncbi.nlm.nih.gov/pubmed/21687746 http://dx.doi.org/10.1371/journal.pone.0021003 |
_version_ | 1782205503761285120 |
---|---|
author | Kurth, Florian Zinnow, Fabienne Prakapenia, Alexandra Dietl, Sabrina Winkler, Stefan Ifflaender, Sascha Rüdiger, Mario Burkhardt, Wolfram |
author_facet | Kurth, Florian Zinnow, Fabienne Prakapenia, Alexandra Dietl, Sabrina Winkler, Stefan Ifflaender, Sascha Rüdiger, Mario Burkhardt, Wolfram |
author_sort | Kurth, Florian |
collection | PubMed |
description | BACKGROUND: Electrical Impedance measurements can be used to estimate the content of intra-thoracic air and thereby give information on pulmonary ventilation. Conventional Impedance measurements mainly indicate relative changes, but no information concerning air-volume is given. The study was performed to test whether a 3-point-calibration with known tidal volumes (VT) during conventional mechanical ventilation (CMV) allows subsequent calculation of VT from total Tidal-Impedance (tTI) measurements using Quadrant Impedance Measurement (QIM). In addition the distribution of TI in different regions of the thorax was examined. METHODOLOGY AND PRINCIPAL FINDINGS: QIM was performed in five neonatal piglets during volume-controlled CMV. tTI values at three different VT (4, 6, 8 ml/kg) were used to establish individual calibration curves. Subsequently, each animal was ventilated with different patterns of varying VT (2–10 ml/kg) at different PEEP levels (0, 3, 6, 9, 12 cmH(2)O). VT variation was repeated after surfactant depletion by bronchoalveolar lavage. VT was calculated from tTI values (VT(calc)) and compared to the VT delivered by the ventilator (VT(PNT)). Bland-Altman analysis revealed good agreement between VT(calc) and VT(PNT) before (bias −0.08 ml; limits of agreement −1.18 to 1.02 ml at PEEP = 3 cmH(2)O) and after surfactant depletion (bias −0.17 ml; limits of agreement −1.57 to 1.22 ml at PEEP = 3 cmH(2)O). At higher PEEP levels VT(calc) was lower than VT(PNT), when only one fixed calibration curve (at PEEP 3 cmH(2)O) was used. With a new calibration curve at each PEEP level the method showed similar accuracy at each PEEP level. TI showed a homogeneous distribution over the four assessed quadrants with a shift toward caudal regions of the thorax with increasing VT. CONCLUSION: Tidal Impedance values could be used for precise and accurate calculation of VT during CMV in this animal study, when calibrated at each PEEP level. |
format | Online Article Text |
id | pubmed-3110249 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2011 |
publisher | Public Library of Science |
record_format | MEDLINE/PubMed |
spelling | pubmed-31102492011-06-16 Continuous Non-Invasive Monitoring of Tidal Volumes by Measurement of Tidal Impedance in Neonatal Piglets Kurth, Florian Zinnow, Fabienne Prakapenia, Alexandra Dietl, Sabrina Winkler, Stefan Ifflaender, Sascha Rüdiger, Mario Burkhardt, Wolfram PLoS One Research Article BACKGROUND: Electrical Impedance measurements can be used to estimate the content of intra-thoracic air and thereby give information on pulmonary ventilation. Conventional Impedance measurements mainly indicate relative changes, but no information concerning air-volume is given. The study was performed to test whether a 3-point-calibration with known tidal volumes (VT) during conventional mechanical ventilation (CMV) allows subsequent calculation of VT from total Tidal-Impedance (tTI) measurements using Quadrant Impedance Measurement (QIM). In addition the distribution of TI in different regions of the thorax was examined. METHODOLOGY AND PRINCIPAL FINDINGS: QIM was performed in five neonatal piglets during volume-controlled CMV. tTI values at three different VT (4, 6, 8 ml/kg) were used to establish individual calibration curves. Subsequently, each animal was ventilated with different patterns of varying VT (2–10 ml/kg) at different PEEP levels (0, 3, 6, 9, 12 cmH(2)O). VT variation was repeated after surfactant depletion by bronchoalveolar lavage. VT was calculated from tTI values (VT(calc)) and compared to the VT delivered by the ventilator (VT(PNT)). Bland-Altman analysis revealed good agreement between VT(calc) and VT(PNT) before (bias −0.08 ml; limits of agreement −1.18 to 1.02 ml at PEEP = 3 cmH(2)O) and after surfactant depletion (bias −0.17 ml; limits of agreement −1.57 to 1.22 ml at PEEP = 3 cmH(2)O). At higher PEEP levels VT(calc) was lower than VT(PNT), when only one fixed calibration curve (at PEEP 3 cmH(2)O) was used. With a new calibration curve at each PEEP level the method showed similar accuracy at each PEEP level. TI showed a homogeneous distribution over the four assessed quadrants with a shift toward caudal regions of the thorax with increasing VT. CONCLUSION: Tidal Impedance values could be used for precise and accurate calculation of VT during CMV in this animal study, when calibrated at each PEEP level. Public Library of Science 2011-06-07 /pmc/articles/PMC3110249/ /pubmed/21687746 http://dx.doi.org/10.1371/journal.pone.0021003 Text en Kurth et al. http://creativecommons.org/licenses/by/4.0/ This is an open-access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are properly credited. |
spellingShingle | Research Article Kurth, Florian Zinnow, Fabienne Prakapenia, Alexandra Dietl, Sabrina Winkler, Stefan Ifflaender, Sascha Rüdiger, Mario Burkhardt, Wolfram Continuous Non-Invasive Monitoring of Tidal Volumes by Measurement of Tidal Impedance in Neonatal Piglets |
title | Continuous Non-Invasive Monitoring of Tidal Volumes by Measurement of Tidal Impedance in Neonatal Piglets |
title_full | Continuous Non-Invasive Monitoring of Tidal Volumes by Measurement of Tidal Impedance in Neonatal Piglets |
title_fullStr | Continuous Non-Invasive Monitoring of Tidal Volumes by Measurement of Tidal Impedance in Neonatal Piglets |
title_full_unstemmed | Continuous Non-Invasive Monitoring of Tidal Volumes by Measurement of Tidal Impedance in Neonatal Piglets |
title_short | Continuous Non-Invasive Monitoring of Tidal Volumes by Measurement of Tidal Impedance in Neonatal Piglets |
title_sort | continuous non-invasive monitoring of tidal volumes by measurement of tidal impedance in neonatal piglets |
topic | Research Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3110249/ https://www.ncbi.nlm.nih.gov/pubmed/21687746 http://dx.doi.org/10.1371/journal.pone.0021003 |
work_keys_str_mv | AT kurthflorian continuousnoninvasivemonitoringoftidalvolumesbymeasurementoftidalimpedanceinneonatalpiglets AT zinnowfabienne continuousnoninvasivemonitoringoftidalvolumesbymeasurementoftidalimpedanceinneonatalpiglets AT prakapeniaalexandra continuousnoninvasivemonitoringoftidalvolumesbymeasurementoftidalimpedanceinneonatalpiglets AT dietlsabrina continuousnoninvasivemonitoringoftidalvolumesbymeasurementoftidalimpedanceinneonatalpiglets AT winklerstefan continuousnoninvasivemonitoringoftidalvolumesbymeasurementoftidalimpedanceinneonatalpiglets AT ifflaendersascha continuousnoninvasivemonitoringoftidalvolumesbymeasurementoftidalimpedanceinneonatalpiglets AT rudigermario continuousnoninvasivemonitoringoftidalvolumesbymeasurementoftidalimpedanceinneonatalpiglets AT burkhardtwolfram continuousnoninvasivemonitoringoftidalvolumesbymeasurementoftidalimpedanceinneonatalpiglets |