A comparison of the accuracy of ulnar versus median nerve stimulation for neuromuscular monitoring

BACKGROUND: Inexperienced anesthesiologists are frequently unclear as to whether to stimulate the ulnar or median nerve to monitor the adductor pollicis. The primary purpose of this study was to determine whether monitoring the adductor pollicis by positioning the stimulating electrodes over the med...

Descripción completa

Detalles Bibliográficos
Autores principales: Lee, Hee Jong, Kim, Kyo Sang, Shim, Jae Chul, Yoon, Sung Wook
Formato: Online Artículo Texto
Lenguaje:English
Publicado: The Korean Society of Anesthesiologists 2011
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3110291/
https://www.ncbi.nlm.nih.gov/pubmed/21716963
http://dx.doi.org/10.4097/kjae.2011.60.5.334
Descripción
Sumario:BACKGROUND: Inexperienced anesthesiologists are frequently unclear as to whether to stimulate the ulnar or median nerve to monitor the adductor pollicis. The primary purpose of this study was to determine whether monitoring the adductor pollicis by positioning the stimulating electrodes over the median nerve is an acceptable alternative to applying electrodes over the ulnar nerve. METHODS: In 20 patients anesthetized with propofol and remifentanil, one pair of stimulating electrodes was positioned over the ulnar nerve. A second pair was placed over the median nerve on the other hand. The acceleromyographic response was monitored on both hands. Rocuronium 0.6 mg/kg was administered. Single twitch (ST) and train-of-four (TOF) stimulations were applied alternatively to both sites. RESULTS: None of the patients showed a twitch response at either site after injection of rocuronium. There were no differences in the mean supramaximal threshold, mean initial TOF ratio, or mean initial ST ratio between the two sites. Bland-Altman analysis revealed a bias (limit of agreement) in the TOF and ST ratios over the median nerve of 7% (± 31%) and 26% (± 73%), respectively, as compared with the ulnar nerve. The median nerve TOF ratio was overestimated by 16.2%, as compared with that of the ulnar nerve value, and the median nerve ST ratio was overestimated by 72.9%, as compared to that of the ulnar nerve. CONCLUSIONS: The ulnar and median nerves cannot be used interchangeably for accurate neuromuscular monitoring.