True logarithmic amplification of frequency clock in SS-OCT for calibration

With swept source optical coherence tomography (SS-OCT), imprecise signal calibration prevents optimal imaging of biological tissues such as coronary artery. This work demonstrates an approach using a true logarithmic amplifier to precondition the clock signal, with the effort to minimize the noises...

Descripción completa

Detalles Bibliográficos
Autores principales: Liu, Bin, Azimi, Ehsan, Brezinski, Mark E.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Optical Society of America 2011
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3114241/
https://www.ncbi.nlm.nih.gov/pubmed/21698036
http://dx.doi.org/10.1364/BOE.2.001769
Descripción
Sumario:With swept source optical coherence tomography (SS-OCT), imprecise signal calibration prevents optimal imaging of biological tissues such as coronary artery. This work demonstrates an approach using a true logarithmic amplifier to precondition the clock signal, with the effort to minimize the noises and phase errors for optimal calibration. This method was validated and tested with a high-speed SS-OCT. The experimental results manifest its superior ability on optimization of the calibration and improvement of the imaging performance. Particularly, this hardware-based approach is suitable for real-time calibration in a high-speed system where computation time is constrained.