Experience with the Pascal(®) photocoagulator: An analysis of over 1200 laser procedures with regard to parameter refinement
AIM: To systematically refine and recommend parameter settings of spot size, power, and treatment duration using the Pascal(®) photocoagulator, a multi-spot, semi-automated, short-duration laser system. MATERIALS AND METHODS: A retrospective consecutive series with 752 Caucasian eyes and 1242 laser...
Autores principales: | , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Medknow Publications
2011
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3116566/ https://www.ncbi.nlm.nih.gov/pubmed/21350276 http://dx.doi.org/10.4103/0301-4738.77007 |
Sumario: | AIM: To systematically refine and recommend parameter settings of spot size, power, and treatment duration using the Pascal(®) photocoagulator, a multi-spot, semi-automated, short-duration laser system. MATERIALS AND METHODS: A retrospective consecutive series with 752 Caucasian eyes and 1242 laser procedures over two years were grouped into, (1) 374 macular focal / grid photocoagulation (FP), (2), 666 panretinal photocoagulation (PRP), and (3) 202 barrage photocoagulation (BP). Parameters for power, duration, spot number, and spot size were recorded for every group. RESULTS: Power parameters for all groups showed a non-gaussian distribution; FP group, median 190 mW, range 100 – 950 mW, and PRP group, median 800 mW, range 100 – 2000 mW. On subgroup comparison, for similar spot size, as treatment duration decreased, the power required increased, albeit in a much lesser proportion than that given by energy = power × time. Most frequently used patterns were single spot (89% of cases) in FP, 5 × 5 box (72%) in PRP, and 2 × 2 box (78%) in BP. Spot diameters as high as ≈ 700 μm on retina were given in the PRP group. Single session PRP was attempted in six eyes with a median spot count of 3500. CONCLUSION: Overall, due to the small duration of its pulse, the Pascal(®) photocoagulator tends to use higher powers, although much lower cumulative energies, than those used in a conventional laser. The consequent lesser heat dissipation, especially lateral, can allow one to use relatively larger spot sizes and give more closely spaced burns, without incurring significant side effects. |
---|