Cargando…
Dosimetric effect of multileaf collimator leaf width in intensity-modulated radiotherapy delivery techniques for small- and large-volume targets
The purpose of this study was to evaluate the dosimetric effect of the leaf width of a multileaf collimator (MLC) in intensity-modulated radiotherapy (IMRT) delivery techniques for small- and large-volume targets. We retrospectively selected previously treated 5 intracranial and 5 head-neck patients...
Autores principales: | , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Medknow Publications
2011
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3119955/ https://www.ncbi.nlm.nih.gov/pubmed/21731222 http://dx.doi.org/10.4103/0971-6203.79690 |
_version_ | 1782206609372479488 |
---|---|
author | Yoganathan, S. A. Mani, Karthick Raj Das, K. J. Maria Agarwal, Arpita Kumar, Shaleen |
author_facet | Yoganathan, S. A. Mani, Karthick Raj Das, K. J. Maria Agarwal, Arpita Kumar, Shaleen |
author_sort | Yoganathan, S. A. |
collection | PubMed |
description | The purpose of this study was to evaluate the dosimetric effect of the leaf width of a multileaf collimator (MLC) in intensity-modulated radiotherapy (IMRT) delivery techniques for small- and large-volume targets. We retrospectively selected previously treated 5 intracranial and 5 head-neck patients for this study to represent small- (range, 18.37-72.75 cc; mean, 42.99 cc) and large-volume (range, 312.31-472.84 cc; mean, 361.14 cc) targets. A 6-MV photon beam data was configured for Brianlab m3 (3 mm), Varian Millennium 120 (5 mm) and Millennium 80 (10 mm) MLCs in the Eclipse treatment-planning system. Sliding window and step-shoot IMRT plans were generated for intracranial patients using all the above-mentioned MLCs; but due to the field size limitation of Brainlab MLC, we used only 5-mm and 10-mm MLCs in the head-and-neck patients. Target conformity, dose to the critical organs and dose to normal tissues were recorded and evaluated. Although the 3-mm MLC resulted in better target conformity (mean difference of 7.7% over 5-mm MLC and 12.7% over 10-mm MLC) over other MLCs for small-volume targets, it increased the total monitor units of the plans. No appreciable differences in terms of target conformity, organ at risk and normal-tissue sparing were observed between the 5-mm and 10-mm MLCs for large-volume targets. The effect of MLC leaf width was not quantifiably different in sliding window and step and shoot techniques. In addition, we observed that there was no additional benefit to the sliding-window (SW) technique when compared to the step-shoot (SS) technique as a result of reduction of MLC leaf width. |
format | Online Article Text |
id | pubmed-3119955 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2011 |
publisher | Medknow Publications |
record_format | MEDLINE/PubMed |
spelling | pubmed-31199552011-06-30 Dosimetric effect of multileaf collimator leaf width in intensity-modulated radiotherapy delivery techniques for small- and large-volume targets Yoganathan, S. A. Mani, Karthick Raj Das, K. J. Maria Agarwal, Arpita Kumar, Shaleen J Med Phys Original Article The purpose of this study was to evaluate the dosimetric effect of the leaf width of a multileaf collimator (MLC) in intensity-modulated radiotherapy (IMRT) delivery techniques for small- and large-volume targets. We retrospectively selected previously treated 5 intracranial and 5 head-neck patients for this study to represent small- (range, 18.37-72.75 cc; mean, 42.99 cc) and large-volume (range, 312.31-472.84 cc; mean, 361.14 cc) targets. A 6-MV photon beam data was configured for Brianlab m3 (3 mm), Varian Millennium 120 (5 mm) and Millennium 80 (10 mm) MLCs in the Eclipse treatment-planning system. Sliding window and step-shoot IMRT plans were generated for intracranial patients using all the above-mentioned MLCs; but due to the field size limitation of Brainlab MLC, we used only 5-mm and 10-mm MLCs in the head-and-neck patients. Target conformity, dose to the critical organs and dose to normal tissues were recorded and evaluated. Although the 3-mm MLC resulted in better target conformity (mean difference of 7.7% over 5-mm MLC and 12.7% over 10-mm MLC) over other MLCs for small-volume targets, it increased the total monitor units of the plans. No appreciable differences in terms of target conformity, organ at risk and normal-tissue sparing were observed between the 5-mm and 10-mm MLCs for large-volume targets. The effect of MLC leaf width was not quantifiably different in sliding window and step and shoot techniques. In addition, we observed that there was no additional benefit to the sliding-window (SW) technique when compared to the step-shoot (SS) technique as a result of reduction of MLC leaf width. Medknow Publications 2011 /pmc/articles/PMC3119955/ /pubmed/21731222 http://dx.doi.org/10.4103/0971-6203.79690 Text en © Journal of Medical Physics http://creativecommons.org/licenses/by/2.0/ This is an open-access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited. |
spellingShingle | Original Article Yoganathan, S. A. Mani, Karthick Raj Das, K. J. Maria Agarwal, Arpita Kumar, Shaleen Dosimetric effect of multileaf collimator leaf width in intensity-modulated radiotherapy delivery techniques for small- and large-volume targets |
title | Dosimetric effect of multileaf collimator leaf width in intensity-modulated radiotherapy delivery techniques for small- and large-volume targets |
title_full | Dosimetric effect of multileaf collimator leaf width in intensity-modulated radiotherapy delivery techniques for small- and large-volume targets |
title_fullStr | Dosimetric effect of multileaf collimator leaf width in intensity-modulated radiotherapy delivery techniques for small- and large-volume targets |
title_full_unstemmed | Dosimetric effect of multileaf collimator leaf width in intensity-modulated radiotherapy delivery techniques for small- and large-volume targets |
title_short | Dosimetric effect of multileaf collimator leaf width in intensity-modulated radiotherapy delivery techniques for small- and large-volume targets |
title_sort | dosimetric effect of multileaf collimator leaf width in intensity-modulated radiotherapy delivery techniques for small- and large-volume targets |
topic | Original Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3119955/ https://www.ncbi.nlm.nih.gov/pubmed/21731222 http://dx.doi.org/10.4103/0971-6203.79690 |
work_keys_str_mv | AT yoganathansa dosimetriceffectofmultileafcollimatorleafwidthinintensitymodulatedradiotherapydeliverytechniquesforsmallandlargevolumetargets AT manikarthickraj dosimetriceffectofmultileafcollimatorleafwidthinintensitymodulatedradiotherapydeliverytechniquesforsmallandlargevolumetargets AT daskjmaria dosimetriceffectofmultileafcollimatorleafwidthinintensitymodulatedradiotherapydeliverytechniquesforsmallandlargevolumetargets AT agarwalarpita dosimetriceffectofmultileafcollimatorleafwidthinintensitymodulatedradiotherapydeliverytechniquesforsmallandlargevolumetargets AT kumarshaleen dosimetriceffectofmultileafcollimatorleafwidthinintensitymodulatedradiotherapydeliverytechniquesforsmallandlargevolumetargets |