A randomized controlled study to evaluate and compare Truview blade with Macintosh blade for laryngoscopy and intubation under general anesthesia

BACKGROUND: The Truview EVO2™ laryngoscope is a recently introduced device with a unique blade that provides a magnified laryngeal view at 42° anterior reflected view. It facilitates visualization of the glottis without alignment of oral, pharyngeal, and tracheal axes. We compared the view obtained...

Descripción completa

Detalles Bibliográficos
Autores principales: Timanaykar, Ramesh T, Anand, Lakesh K, Palta, Sanjeev
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Medknow Publications Pvt Ltd 2011
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3127299/
https://www.ncbi.nlm.nih.gov/pubmed/21772680
http://dx.doi.org/10.4103/0970-9185.81838
_version_ 1782207334755336192
author Timanaykar, Ramesh T
Anand, Lakesh K
Palta, Sanjeev
author_facet Timanaykar, Ramesh T
Anand, Lakesh K
Palta, Sanjeev
author_sort Timanaykar, Ramesh T
collection PubMed
description BACKGROUND: The Truview EVO2™ laryngoscope is a recently introduced device with a unique blade that provides a magnified laryngeal view at 42° anterior reflected view. It facilitates visualization of the glottis without alignment of oral, pharyngeal, and tracheal axes. We compared the view obtained at laryngoscopy, intubating conditions and hemodynamic parameters of Truview with Macintosh blade. MATERIALS AND METHODS: In prospective, randomized and controlled manner, 200 patients of ASA I and II of either sex (20–50 years), presenting for surgery requiring tracheal intubation, were assigned to undergo intubation using a Truview or Macintosh laryngoscope. Visualization of the vocal cord, ease of intubation, time taken for intubation, number of attempts, and hemodynamic parameters were evaluated. RESULTS: Truview provided better results for the laryngeal view using Cormack and Lehane grading, particularly in patients with higher airway Mallampati grading (P < 0.05). The time taken for intubation (33.06±5.6 vs. 23.11±57 seconds) was more with Truview than with Macintosh blade (P < 0.01). The Percentage of Glottic Opening (POGO) score was significantly higher (97.26±8) in Truview as that observed with Macintosh blade (83.70±21.5). Hemodynamic parameters increased after tracheal intubation from pre-intubation value (P < 0.05) in both the groups, but they were comparable amongst the groups. No postoperative adverse events were noted. CONCLUSION: Tracheal intubation using Truview blade provided consistently improved laryngeal view as compared to Macintosh blade without the need to align the oral, pharyngeal and tracheal axes, with equal attempts for successful intubation and similar changes in hemodynamics. However, the time taken for intubation was more with Truview.
format Online
Article
Text
id pubmed-3127299
institution National Center for Biotechnology Information
language English
publishDate 2011
publisher Medknow Publications Pvt Ltd
record_format MEDLINE/PubMed
spelling pubmed-31272992011-07-19 A randomized controlled study to evaluate and compare Truview blade with Macintosh blade for laryngoscopy and intubation under general anesthesia Timanaykar, Ramesh T Anand, Lakesh K Palta, Sanjeev J Anaesthesiol Clin Pharmacol Original Article BACKGROUND: The Truview EVO2™ laryngoscope is a recently introduced device with a unique blade that provides a magnified laryngeal view at 42° anterior reflected view. It facilitates visualization of the glottis without alignment of oral, pharyngeal, and tracheal axes. We compared the view obtained at laryngoscopy, intubating conditions and hemodynamic parameters of Truview with Macintosh blade. MATERIALS AND METHODS: In prospective, randomized and controlled manner, 200 patients of ASA I and II of either sex (20–50 years), presenting for surgery requiring tracheal intubation, were assigned to undergo intubation using a Truview or Macintosh laryngoscope. Visualization of the vocal cord, ease of intubation, time taken for intubation, number of attempts, and hemodynamic parameters were evaluated. RESULTS: Truview provided better results for the laryngeal view using Cormack and Lehane grading, particularly in patients with higher airway Mallampati grading (P < 0.05). The time taken for intubation (33.06±5.6 vs. 23.11±57 seconds) was more with Truview than with Macintosh blade (P < 0.01). The Percentage of Glottic Opening (POGO) score was significantly higher (97.26±8) in Truview as that observed with Macintosh blade (83.70±21.5). Hemodynamic parameters increased after tracheal intubation from pre-intubation value (P < 0.05) in both the groups, but they were comparable amongst the groups. No postoperative adverse events were noted. CONCLUSION: Tracheal intubation using Truview blade provided consistently improved laryngeal view as compared to Macintosh blade without the need to align the oral, pharyngeal and tracheal axes, with equal attempts for successful intubation and similar changes in hemodynamics. However, the time taken for intubation was more with Truview. Medknow Publications Pvt Ltd 2011 /pmc/articles/PMC3127299/ /pubmed/21772680 http://dx.doi.org/10.4103/0970-9185.81838 Text en © Journal of Anaesthesiology Clinical Pharmacology http://creativecommons.org/licenses/by-nc-sa/3.0 This is an open-access article distributed under the terms of the Creative Commons Attribution-Noncommercial-Share Alike 3.0 Unported, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
spellingShingle Original Article
Timanaykar, Ramesh T
Anand, Lakesh K
Palta, Sanjeev
A randomized controlled study to evaluate and compare Truview blade with Macintosh blade for laryngoscopy and intubation under general anesthesia
title A randomized controlled study to evaluate and compare Truview blade with Macintosh blade for laryngoscopy and intubation under general anesthesia
title_full A randomized controlled study to evaluate and compare Truview blade with Macintosh blade for laryngoscopy and intubation under general anesthesia
title_fullStr A randomized controlled study to evaluate and compare Truview blade with Macintosh blade for laryngoscopy and intubation under general anesthesia
title_full_unstemmed A randomized controlled study to evaluate and compare Truview blade with Macintosh blade for laryngoscopy and intubation under general anesthesia
title_short A randomized controlled study to evaluate and compare Truview blade with Macintosh blade for laryngoscopy and intubation under general anesthesia
title_sort randomized controlled study to evaluate and compare truview blade with macintosh blade for laryngoscopy and intubation under general anesthesia
topic Original Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3127299/
https://www.ncbi.nlm.nih.gov/pubmed/21772680
http://dx.doi.org/10.4103/0970-9185.81838
work_keys_str_mv AT timanaykarramesht arandomizedcontrolledstudytoevaluateandcomparetruviewbladewithmacintoshbladeforlaryngoscopyandintubationundergeneralanesthesia
AT anandlakeshk arandomizedcontrolledstudytoevaluateandcomparetruviewbladewithmacintoshbladeforlaryngoscopyandintubationundergeneralanesthesia
AT paltasanjeev arandomizedcontrolledstudytoevaluateandcomparetruviewbladewithmacintoshbladeforlaryngoscopyandintubationundergeneralanesthesia
AT timanaykarramesht randomizedcontrolledstudytoevaluateandcomparetruviewbladewithmacintoshbladeforlaryngoscopyandintubationundergeneralanesthesia
AT anandlakeshk randomizedcontrolledstudytoevaluateandcomparetruviewbladewithmacintoshbladeforlaryngoscopyandintubationundergeneralanesthesia
AT paltasanjeev randomizedcontrolledstudytoevaluateandcomparetruviewbladewithmacintoshbladeforlaryngoscopyandintubationundergeneralanesthesia