Comparison of Depth of Cure, Hardness and Heat Generation of LED and High Intensity QTH Light Sources

OBJECTIVES: To compare curing performance of a second generation LED curing light with a high power tungsten quartz halogen (QTH). METHODS: A hybrid composite resin (Filtek Z 250, 3M, USA) was used as test material and cured using a second generation LED light (Translux Power Blue™, Heraus Kulzer ,G...

Descripción completa

Detalles Bibliográficos
Autores principales: Mousavinasab, Sayed Mostafa, Meyers, Ian
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Dental Investigations Society 2011
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3137443/
https://www.ncbi.nlm.nih.gov/pubmed/21769271
_version_ 1782208290742075392
author Mousavinasab, Sayed Mostafa
Meyers, Ian
author_facet Mousavinasab, Sayed Mostafa
Meyers, Ian
author_sort Mousavinasab, Sayed Mostafa
collection PubMed
description OBJECTIVES: To compare curing performance of a second generation LED curing light with a high power tungsten quartz halogen (QTH). METHODS: A hybrid composite resin (Filtek Z 250, 3M, USA) was used as test material and cured using a second generation LED light (Translux Power Blue™, Heraus Kulzer ,Germany) or a very high power QTH light unit (EMS, Switzerland). A two split aluminum mold was used to prepare ten samples with LED light source cured for forty seconds and ten samples prepared using high power QTH light unit, cured for four or six seconds recommended exposure time. Hardness, depth of cure (DOC) and thermal rise during exposure time by these light sources were measured. The data submitted to analysis of variance (ANOVA), Tukey’s and student’s t tests at 5% significance level. RESULTS: Significant differences were found in hardness, DOC of samples cured by above mentioned light sources and also in thermal rises during exposure time. The curing performance of the tested QTH was not as well as the LED light. TPB light source produced the maximum hardness (81.25, 73.29, 65.49,55.83 and 24.53 for 0 mm, 1 mm, 2 mm, 3 mm and 4 mm intervals) and DOC (2.64 mm) values with forty seconds irradiation time and the high power (QTH) the least hardness (73.27, 61.51 and 31.59 for 0 mm, 1 mm and 2 mm, respectively) and DOC (2 mm) values with four seconds irradiation time. Thermal rises during 4 s and 6 s curing time using high power QTH and tested LED were 1.88°C, 3°C and 1.87°C, respectively. CONCLUSIONS: The used high power LED light produced greater hardness and depth of cure during forty seconds exposure time compared to high power QTH light with four or six seconds curing time. Thermal rise during 6 s curing time with QTH was greater compared to thermal changes occurred during 40 s curing time with tested LED light source. There was no difference seen in thermal changes caused by LED light with 40 s and QTH light with 4 s exposure time.
format Online
Article
Text
id pubmed-3137443
institution National Center for Biotechnology Information
language English
publishDate 2011
publisher Dental Investigations Society
record_format MEDLINE/PubMed
spelling pubmed-31374432011-07-18 Comparison of Depth of Cure, Hardness and Heat Generation of LED and High Intensity QTH Light Sources Mousavinasab, Sayed Mostafa Meyers, Ian Eur J Dent Original Articles OBJECTIVES: To compare curing performance of a second generation LED curing light with a high power tungsten quartz halogen (QTH). METHODS: A hybrid composite resin (Filtek Z 250, 3M, USA) was used as test material and cured using a second generation LED light (Translux Power Blue™, Heraus Kulzer ,Germany) or a very high power QTH light unit (EMS, Switzerland). A two split aluminum mold was used to prepare ten samples with LED light source cured for forty seconds and ten samples prepared using high power QTH light unit, cured for four or six seconds recommended exposure time. Hardness, depth of cure (DOC) and thermal rise during exposure time by these light sources were measured. The data submitted to analysis of variance (ANOVA), Tukey’s and student’s t tests at 5% significance level. RESULTS: Significant differences were found in hardness, DOC of samples cured by above mentioned light sources and also in thermal rises during exposure time. The curing performance of the tested QTH was not as well as the LED light. TPB light source produced the maximum hardness (81.25, 73.29, 65.49,55.83 and 24.53 for 0 mm, 1 mm, 2 mm, 3 mm and 4 mm intervals) and DOC (2.64 mm) values with forty seconds irradiation time and the high power (QTH) the least hardness (73.27, 61.51 and 31.59 for 0 mm, 1 mm and 2 mm, respectively) and DOC (2 mm) values with four seconds irradiation time. Thermal rises during 4 s and 6 s curing time using high power QTH and tested LED were 1.88°C, 3°C and 1.87°C, respectively. CONCLUSIONS: The used high power LED light produced greater hardness and depth of cure during forty seconds exposure time compared to high power QTH light with four or six seconds curing time. Thermal rise during 6 s curing time with QTH was greater compared to thermal changes occurred during 40 s curing time with tested LED light source. There was no difference seen in thermal changes caused by LED light with 40 s and QTH light with 4 s exposure time. Dental Investigations Society 2011-07 /pmc/articles/PMC3137443/ /pubmed/21769271 Text en Copyright 2011 European Journal of Dentistry. All rights reserved.
spellingShingle Original Articles
Mousavinasab, Sayed Mostafa
Meyers, Ian
Comparison of Depth of Cure, Hardness and Heat Generation of LED and High Intensity QTH Light Sources
title Comparison of Depth of Cure, Hardness and Heat Generation of LED and High Intensity QTH Light Sources
title_full Comparison of Depth of Cure, Hardness and Heat Generation of LED and High Intensity QTH Light Sources
title_fullStr Comparison of Depth of Cure, Hardness and Heat Generation of LED and High Intensity QTH Light Sources
title_full_unstemmed Comparison of Depth of Cure, Hardness and Heat Generation of LED and High Intensity QTH Light Sources
title_short Comparison of Depth of Cure, Hardness and Heat Generation of LED and High Intensity QTH Light Sources
title_sort comparison of depth of cure, hardness and heat generation of led and high intensity qth light sources
topic Original Articles
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3137443/
https://www.ncbi.nlm.nih.gov/pubmed/21769271
work_keys_str_mv AT mousavinasabsayedmostafa comparisonofdepthofcurehardnessandheatgenerationofledandhighintensityqthlightsources
AT meyersian comparisonofdepthofcurehardnessandheatgenerationofledandhighintensityqthlightsources