Expansion of the Multi-Link Frontier™ Coronary Bifurcation Stent: Micro-Computed Tomographic Assessment in Human Autopsy and Porcine Heart Samples
BACKGROUND: Treatment of coronary bifurcation lesions remains challenging, beyond the introduction of drug eluting stents. Dedicated stent systems are available to improve the technical approach to the treatment of these lesions. However dedicated stent systems have so far not reduced the incidence...
Autores principales: | , , , , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Public Library of Science
2011
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3140974/ https://www.ncbi.nlm.nih.gov/pubmed/21814552 http://dx.doi.org/10.1371/journal.pone.0021778 |
_version_ | 1782208603391787008 |
---|---|
author | Kralev, Stefan Haag, Benjamin Spannenberger, Jens Lang, Siegfried Brockmann, Marc A. Bartling, Soenke Marx, Alexander Haase, Karl-Konstantin Borggrefe, Martin Süselbeck, Tim |
author_facet | Kralev, Stefan Haag, Benjamin Spannenberger, Jens Lang, Siegfried Brockmann, Marc A. Bartling, Soenke Marx, Alexander Haase, Karl-Konstantin Borggrefe, Martin Süselbeck, Tim |
author_sort | Kralev, Stefan |
collection | PubMed |
description | BACKGROUND: Treatment of coronary bifurcation lesions remains challenging, beyond the introduction of drug eluting stents. Dedicated stent systems are available to improve the technical approach to the treatment of these lesions. However dedicated stent systems have so far not reduced the incidence of stent restenosis. The aim of this study was to assess the expansion of the Multi-Link (ML) Frontier™ stent in human and porcine coronary arteries to provide the cardiologist with useful in-vitro information for stent implantation and selection. METHODOLOGY/PRINCIPAL FINDINGS: Nine ML Frontier™ stents were implanted in seven human autopsy heart samples with known coronary artery disease and five ML Frontier™ stents were implanted in five porcine hearts. Proximal, distal and side branch diameters (PD, DD, SBD, respectively), corresponding opening areas (PA, DA, SBA) and the mean stent length (L) were assessed by micro-computed tomography (micro-CT). PD and PA were significantly smaller in human autopsy heart samples than in porcine heart samples (3.54±0.47 mm vs. 4.04±0.22 mm, p = 0.048; 10.00±2.42 mm(2) vs. 12.84±1.38 mm(2), p = 0.034, respectively) and than those given by the manufacturer (3.54±0.47 mm vs. 4.03 mm, p = 0.014). L was smaller in human autopsy heart samples than in porcine heart samples, although data did not reach significance (16.66±1.30 mm vs. 17.30±0.51 mm, p = 0.32), and significantly smaller than that given by the manufacturer (16.66±1.30 mm vs. 18 mm, p = 0.015). CONCLUSIONS/SIGNIFICANCE: Micro-CT is a feasible tool for exact surveying of dedicated stent systems and could make a contribution to the development of these devices. The proximal diameter and proximal area of the stent system were considerably smaller in human autopsy heart samples than in porcine heart samples and than those given by the manufacturer. Special consideration should be given to the stent deployment procedure (and to the follow-up) of dedicated stent systems, considering final intravascular ultrasound or optical coherence tomography to visualize (and if necessary optimize) stent expansion. |
format | Online Article Text |
id | pubmed-3140974 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2011 |
publisher | Public Library of Science |
record_format | MEDLINE/PubMed |
spelling | pubmed-31409742011-08-03 Expansion of the Multi-Link Frontier™ Coronary Bifurcation Stent: Micro-Computed Tomographic Assessment in Human Autopsy and Porcine Heart Samples Kralev, Stefan Haag, Benjamin Spannenberger, Jens Lang, Siegfried Brockmann, Marc A. Bartling, Soenke Marx, Alexander Haase, Karl-Konstantin Borggrefe, Martin Süselbeck, Tim PLoS One Research Article BACKGROUND: Treatment of coronary bifurcation lesions remains challenging, beyond the introduction of drug eluting stents. Dedicated stent systems are available to improve the technical approach to the treatment of these lesions. However dedicated stent systems have so far not reduced the incidence of stent restenosis. The aim of this study was to assess the expansion of the Multi-Link (ML) Frontier™ stent in human and porcine coronary arteries to provide the cardiologist with useful in-vitro information for stent implantation and selection. METHODOLOGY/PRINCIPAL FINDINGS: Nine ML Frontier™ stents were implanted in seven human autopsy heart samples with known coronary artery disease and five ML Frontier™ stents were implanted in five porcine hearts. Proximal, distal and side branch diameters (PD, DD, SBD, respectively), corresponding opening areas (PA, DA, SBA) and the mean stent length (L) were assessed by micro-computed tomography (micro-CT). PD and PA were significantly smaller in human autopsy heart samples than in porcine heart samples (3.54±0.47 mm vs. 4.04±0.22 mm, p = 0.048; 10.00±2.42 mm(2) vs. 12.84±1.38 mm(2), p = 0.034, respectively) and than those given by the manufacturer (3.54±0.47 mm vs. 4.03 mm, p = 0.014). L was smaller in human autopsy heart samples than in porcine heart samples, although data did not reach significance (16.66±1.30 mm vs. 17.30±0.51 mm, p = 0.32), and significantly smaller than that given by the manufacturer (16.66±1.30 mm vs. 18 mm, p = 0.015). CONCLUSIONS/SIGNIFICANCE: Micro-CT is a feasible tool for exact surveying of dedicated stent systems and could make a contribution to the development of these devices. The proximal diameter and proximal area of the stent system were considerably smaller in human autopsy heart samples than in porcine heart samples and than those given by the manufacturer. Special consideration should be given to the stent deployment procedure (and to the follow-up) of dedicated stent systems, considering final intravascular ultrasound or optical coherence tomography to visualize (and if necessary optimize) stent expansion. Public Library of Science 2011-07-21 /pmc/articles/PMC3140974/ /pubmed/21814552 http://dx.doi.org/10.1371/journal.pone.0021778 Text en Kralev et al. http://creativecommons.org/licenses/by/4.0/ This is an open-access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are properly credited. |
spellingShingle | Research Article Kralev, Stefan Haag, Benjamin Spannenberger, Jens Lang, Siegfried Brockmann, Marc A. Bartling, Soenke Marx, Alexander Haase, Karl-Konstantin Borggrefe, Martin Süselbeck, Tim Expansion of the Multi-Link Frontier™ Coronary Bifurcation Stent: Micro-Computed Tomographic Assessment in Human Autopsy and Porcine Heart Samples |
title | Expansion of the Multi-Link Frontier™ Coronary Bifurcation Stent: Micro-Computed Tomographic Assessment in Human Autopsy and Porcine Heart Samples |
title_full | Expansion of the Multi-Link Frontier™ Coronary Bifurcation Stent: Micro-Computed Tomographic Assessment in Human Autopsy and Porcine Heart Samples |
title_fullStr | Expansion of the Multi-Link Frontier™ Coronary Bifurcation Stent: Micro-Computed Tomographic Assessment in Human Autopsy and Porcine Heart Samples |
title_full_unstemmed | Expansion of the Multi-Link Frontier™ Coronary Bifurcation Stent: Micro-Computed Tomographic Assessment in Human Autopsy and Porcine Heart Samples |
title_short | Expansion of the Multi-Link Frontier™ Coronary Bifurcation Stent: Micro-Computed Tomographic Assessment in Human Autopsy and Porcine Heart Samples |
title_sort | expansion of the multi-link frontier™ coronary bifurcation stent: micro-computed tomographic assessment in human autopsy and porcine heart samples |
topic | Research Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3140974/ https://www.ncbi.nlm.nih.gov/pubmed/21814552 http://dx.doi.org/10.1371/journal.pone.0021778 |
work_keys_str_mv | AT kralevstefan expansionofthemultilinkfrontiercoronarybifurcationstentmicrocomputedtomographicassessmentinhumanautopsyandporcineheartsamples AT haagbenjamin expansionofthemultilinkfrontiercoronarybifurcationstentmicrocomputedtomographicassessmentinhumanautopsyandporcineheartsamples AT spannenbergerjens expansionofthemultilinkfrontiercoronarybifurcationstentmicrocomputedtomographicassessmentinhumanautopsyandporcineheartsamples AT langsiegfried expansionofthemultilinkfrontiercoronarybifurcationstentmicrocomputedtomographicassessmentinhumanautopsyandporcineheartsamples AT brockmannmarca expansionofthemultilinkfrontiercoronarybifurcationstentmicrocomputedtomographicassessmentinhumanautopsyandporcineheartsamples AT bartlingsoenke expansionofthemultilinkfrontiercoronarybifurcationstentmicrocomputedtomographicassessmentinhumanautopsyandporcineheartsamples AT marxalexander expansionofthemultilinkfrontiercoronarybifurcationstentmicrocomputedtomographicassessmentinhumanautopsyandporcineheartsamples AT haasekarlkonstantin expansionofthemultilinkfrontiercoronarybifurcationstentmicrocomputedtomographicassessmentinhumanautopsyandporcineheartsamples AT borggrefemartin expansionofthemultilinkfrontiercoronarybifurcationstentmicrocomputedtomographicassessmentinhumanautopsyandporcineheartsamples AT suselbecktim expansionofthemultilinkfrontiercoronarybifurcationstentmicrocomputedtomographicassessmentinhumanautopsyandporcineheartsamples |