Assessment of EchoMRI-AH™ vs. Dual-energy X-ray Absorptiometry (DXA) to Measure Human Body Composition
BACKGROUND: The sensitivity to detect small changes in body composition (fat mass and fat-free mass) largely depends on the instrument's precision. We compared EchoMRI-AH™ and DXA (Hologic QDR-4500A) for estimating fat mass in 301 volunteers. METHODS: Body composition was evaluated in 136 males...
Autores principales: | , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
2011
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3145037/ https://www.ncbi.nlm.nih.gov/pubmed/21224826 http://dx.doi.org/10.1038/ijo.2010.268 |
_version_ | 1782209065669099520 |
---|---|
author | Galgani, Jose E. Smith, Steven R. Ravussin, Eric |
author_facet | Galgani, Jose E. Smith, Steven R. Ravussin, Eric |
author_sort | Galgani, Jose E. |
collection | PubMed |
description | BACKGROUND: The sensitivity to detect small changes in body composition (fat mass and fat-free mass) largely depends on the instrument's precision. We compared EchoMRI-AH™ and DXA (Hologic QDR-4500A) for estimating fat mass in 301 volunteers. METHODS: Body composition was evaluated in 136 males and 165 females with a large range of body mass index (19–49 kg/m(2)) and age (19–91 y old) using DXA and EchoMRI-AH™. In a subsample of 13 lean (BMI=19–25 kg/m(2)) and 21 overweight/obese (BMI>25 kg/m(2)) individuals, within-subject precision was evaluated from repeated measurements taken within one hour (n=3) and one-week apart (mean of three measurements taken on each day). RESULTS: Using Bland-Altman analysis, we compared the mean of the fat mass measurements vs. the difference in fat mass measured by both instruments. We found that EchoMRI-AH™ quantified larger amount of fat vs. DXA in non-obese (BMI<30 kg/m(2); [1.1, CI(95):-3.7 – 6.0 kg]) and obese (BMI≥30 kg/m(2); [4.2, CI(95):-1.4 – 9.8 kg]) participants. Within-subject precision (coefficient of variation %) in fat mass measured within one hour was remarkably better when measured by EchoMRI-AH™ than DXA (<0.5% vs. <1.5%, respectively; p<0.001). However, one-week apart within-subject variability showed similar values for both instruments (<2.2%; p=0.15). CONCLUSIONS: EchoMRI-AH™ yielded greater fat mass values when compared with DXA (Hologic QDR-4500A), particularly in fatter subjects. EchoMRI-AH™ and DXA showed similar one-week apart precision when fat mass was measured both in lean and overweight/obese individuals. |
format | Online Article Text |
id | pubmed-3145037 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2011 |
record_format | MEDLINE/PubMed |
spelling | pubmed-31450372012-03-01 Assessment of EchoMRI-AH™ vs. Dual-energy X-ray Absorptiometry (DXA) to Measure Human Body Composition Galgani, Jose E. Smith, Steven R. Ravussin, Eric Int J Obes (Lond) Article BACKGROUND: The sensitivity to detect small changes in body composition (fat mass and fat-free mass) largely depends on the instrument's precision. We compared EchoMRI-AH™ and DXA (Hologic QDR-4500A) for estimating fat mass in 301 volunteers. METHODS: Body composition was evaluated in 136 males and 165 females with a large range of body mass index (19–49 kg/m(2)) and age (19–91 y old) using DXA and EchoMRI-AH™. In a subsample of 13 lean (BMI=19–25 kg/m(2)) and 21 overweight/obese (BMI>25 kg/m(2)) individuals, within-subject precision was evaluated from repeated measurements taken within one hour (n=3) and one-week apart (mean of three measurements taken on each day). RESULTS: Using Bland-Altman analysis, we compared the mean of the fat mass measurements vs. the difference in fat mass measured by both instruments. We found that EchoMRI-AH™ quantified larger amount of fat vs. DXA in non-obese (BMI<30 kg/m(2); [1.1, CI(95):-3.7 – 6.0 kg]) and obese (BMI≥30 kg/m(2); [4.2, CI(95):-1.4 – 9.8 kg]) participants. Within-subject precision (coefficient of variation %) in fat mass measured within one hour was remarkably better when measured by EchoMRI-AH™ than DXA (<0.5% vs. <1.5%, respectively; p<0.001). However, one-week apart within-subject variability showed similar values for both instruments (<2.2%; p=0.15). CONCLUSIONS: EchoMRI-AH™ yielded greater fat mass values when compared with DXA (Hologic QDR-4500A), particularly in fatter subjects. EchoMRI-AH™ and DXA showed similar one-week apart precision when fat mass was measured both in lean and overweight/obese individuals. 2011-01-11 2011-09 /pmc/articles/PMC3145037/ /pubmed/21224826 http://dx.doi.org/10.1038/ijo.2010.268 Text en Users may view, print, copy, download and text and data- mine the content in such documents, for the purposes of academic research, subject always to the full Conditions of use: http://www.nature.com/authors/editorial_policies/license.html#terms |
spellingShingle | Article Galgani, Jose E. Smith, Steven R. Ravussin, Eric Assessment of EchoMRI-AH™ vs. Dual-energy X-ray Absorptiometry (DXA) to Measure Human Body Composition |
title | Assessment of EchoMRI-AH™ vs. Dual-energy X-ray Absorptiometry (DXA) to Measure Human Body Composition |
title_full | Assessment of EchoMRI-AH™ vs. Dual-energy X-ray Absorptiometry (DXA) to Measure Human Body Composition |
title_fullStr | Assessment of EchoMRI-AH™ vs. Dual-energy X-ray Absorptiometry (DXA) to Measure Human Body Composition |
title_full_unstemmed | Assessment of EchoMRI-AH™ vs. Dual-energy X-ray Absorptiometry (DXA) to Measure Human Body Composition |
title_short | Assessment of EchoMRI-AH™ vs. Dual-energy X-ray Absorptiometry (DXA) to Measure Human Body Composition |
title_sort | assessment of echomri-ah™ vs. dual-energy x-ray absorptiometry (dxa) to measure human body composition |
topic | Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3145037/ https://www.ncbi.nlm.nih.gov/pubmed/21224826 http://dx.doi.org/10.1038/ijo.2010.268 |
work_keys_str_mv | AT galganijosee assessmentofechomriahvsdualenergyxrayabsorptiometrydxatomeasurehumanbodycomposition AT smithstevenr assessmentofechomriahvsdualenergyxrayabsorptiometrydxatomeasurehumanbodycomposition AT ravussineric assessmentofechomriahvsdualenergyxrayabsorptiometrydxatomeasurehumanbodycomposition |