Relative dosimetrical verification in high dose rate brachytherapy using two-dimensional detector array IMatriXX
For high dose rate (HDR) brachytherapy, independent treatment verification is needed to ensure that the treatment is performed as per prescription. This study demonstrates dosimetric quality assurance of the HDR brachytherapy using a commercially available two-dimensional ion chamber array called IM...
Autores principales: | , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Medknow Publications Pvt Ltd
2011
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3159223/ https://www.ncbi.nlm.nih.gov/pubmed/21897562 http://dx.doi.org/10.4103/0971-6203.83491 |
_version_ | 1782210428075507712 |
---|---|
author | Manikandan, A. Biplab, Sarkar David, Perianayagam A. Holla, R. Vivek, T. R. Sujatha, N. |
author_facet | Manikandan, A. Biplab, Sarkar David, Perianayagam A. Holla, R. Vivek, T. R. Sujatha, N. |
author_sort | Manikandan, A. |
collection | PubMed |
description | For high dose rate (HDR) brachytherapy, independent treatment verification is needed to ensure that the treatment is performed as per prescription. This study demonstrates dosimetric quality assurance of the HDR brachytherapy using a commercially available two-dimensional ion chamber array called IMatriXX, which has a detector separation of 0.7619 cm. The reference isodose length, step size, and source dwell positional accuracy were verified. A total of 24 dwell positions, which were verified for positional accuracy gave a total error (systematic and random) of –0.45 mm, with a standard deviation of 1.01 mm and maximum error of 1.8 mm. Using a step size of 5 mm, reference isodose length (the length of 100% isodose line) was verified for single and multiple catheters of same and different source loadings. An error ≤1 mm was measured in 57% of tests analyzed. Step size verification for 2, 3, 4, and 5 cm was performed and 70% of the step size errors were below 1 mm, with maximum of 1.2 mm. The step size ≤1 cm could not be verified by the IMatriXX as it could not resolve the peaks in dose profile. |
format | Online Article Text |
id | pubmed-3159223 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2011 |
publisher | Medknow Publications Pvt Ltd |
record_format | MEDLINE/PubMed |
spelling | pubmed-31592232011-09-06 Relative dosimetrical verification in high dose rate brachytherapy using two-dimensional detector array IMatriXX Manikandan, A. Biplab, Sarkar David, Perianayagam A. Holla, R. Vivek, T. R. Sujatha, N. J Med Phys Technical Note For high dose rate (HDR) brachytherapy, independent treatment verification is needed to ensure that the treatment is performed as per prescription. This study demonstrates dosimetric quality assurance of the HDR brachytherapy using a commercially available two-dimensional ion chamber array called IMatriXX, which has a detector separation of 0.7619 cm. The reference isodose length, step size, and source dwell positional accuracy were verified. A total of 24 dwell positions, which were verified for positional accuracy gave a total error (systematic and random) of –0.45 mm, with a standard deviation of 1.01 mm and maximum error of 1.8 mm. Using a step size of 5 mm, reference isodose length (the length of 100% isodose line) was verified for single and multiple catheters of same and different source loadings. An error ≤1 mm was measured in 57% of tests analyzed. Step size verification for 2, 3, 4, and 5 cm was performed and 70% of the step size errors were below 1 mm, with maximum of 1.2 mm. The step size ≤1 cm could not be verified by the IMatriXX as it could not resolve the peaks in dose profile. Medknow Publications Pvt Ltd 2011 /pmc/articles/PMC3159223/ /pubmed/21897562 http://dx.doi.org/10.4103/0971-6203.83491 Text en © Journal of Medical Physics http://creativecommons.org/licenses/by-nc-sa/3.0 This is an open-access article distributed under the terms of the Creative Commons Attribution-Noncommercial-Share Alike 3.0 Unported, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited. |
spellingShingle | Technical Note Manikandan, A. Biplab, Sarkar David, Perianayagam A. Holla, R. Vivek, T. R. Sujatha, N. Relative dosimetrical verification in high dose rate brachytherapy using two-dimensional detector array IMatriXX |
title | Relative dosimetrical verification in high dose rate brachytherapy using two-dimensional detector array IMatriXX |
title_full | Relative dosimetrical verification in high dose rate brachytherapy using two-dimensional detector array IMatriXX |
title_fullStr | Relative dosimetrical verification in high dose rate brachytherapy using two-dimensional detector array IMatriXX |
title_full_unstemmed | Relative dosimetrical verification in high dose rate brachytherapy using two-dimensional detector array IMatriXX |
title_short | Relative dosimetrical verification in high dose rate brachytherapy using two-dimensional detector array IMatriXX |
title_sort | relative dosimetrical verification in high dose rate brachytherapy using two-dimensional detector array imatrixx |
topic | Technical Note |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3159223/ https://www.ncbi.nlm.nih.gov/pubmed/21897562 http://dx.doi.org/10.4103/0971-6203.83491 |
work_keys_str_mv | AT manikandana relativedosimetricalverificationinhighdoseratebrachytherapyusingtwodimensionaldetectorarrayimatrixx AT biplabsarkar relativedosimetricalverificationinhighdoseratebrachytherapyusingtwodimensionaldetectorarrayimatrixx AT davidperianayagama relativedosimetricalverificationinhighdoseratebrachytherapyusingtwodimensionaldetectorarrayimatrixx AT hollar relativedosimetricalverificationinhighdoseratebrachytherapyusingtwodimensionaldetectorarrayimatrixx AT vivektr relativedosimetricalverificationinhighdoseratebrachytherapyusingtwodimensionaldetectorarrayimatrixx AT sujathan relativedosimetricalverificationinhighdoseratebrachytherapyusingtwodimensionaldetectorarrayimatrixx |