Relative dosimetrical verification in high dose rate brachytherapy using two-dimensional detector array IMatriXX

For high dose rate (HDR) brachytherapy, independent treatment verification is needed to ensure that the treatment is performed as per prescription. This study demonstrates dosimetric quality assurance of the HDR brachytherapy using a commercially available two-dimensional ion chamber array called IM...

Descripción completa

Detalles Bibliográficos
Autores principales: Manikandan, A., Biplab, Sarkar, David, Perianayagam A., Holla, R., Vivek, T. R., Sujatha, N.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Medknow Publications Pvt Ltd 2011
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3159223/
https://www.ncbi.nlm.nih.gov/pubmed/21897562
http://dx.doi.org/10.4103/0971-6203.83491
_version_ 1782210428075507712
author Manikandan, A.
Biplab, Sarkar
David, Perianayagam A.
Holla, R.
Vivek, T. R.
Sujatha, N.
author_facet Manikandan, A.
Biplab, Sarkar
David, Perianayagam A.
Holla, R.
Vivek, T. R.
Sujatha, N.
author_sort Manikandan, A.
collection PubMed
description For high dose rate (HDR) brachytherapy, independent treatment verification is needed to ensure that the treatment is performed as per prescription. This study demonstrates dosimetric quality assurance of the HDR brachytherapy using a commercially available two-dimensional ion chamber array called IMatriXX, which has a detector separation of 0.7619 cm. The reference isodose length, step size, and source dwell positional accuracy were verified. A total of 24 dwell positions, which were verified for positional accuracy gave a total error (systematic and random) of –0.45 mm, with a standard deviation of 1.01 mm and maximum error of 1.8 mm. Using a step size of 5 mm, reference isodose length (the length of 100% isodose line) was verified for single and multiple catheters of same and different source loadings. An error ≤1 mm was measured in 57% of tests analyzed. Step size verification for 2, 3, 4, and 5 cm was performed and 70% of the step size errors were below 1 mm, with maximum of 1.2 mm. The step size ≤1 cm could not be verified by the IMatriXX as it could not resolve the peaks in dose profile.
format Online
Article
Text
id pubmed-3159223
institution National Center for Biotechnology Information
language English
publishDate 2011
publisher Medknow Publications Pvt Ltd
record_format MEDLINE/PubMed
spelling pubmed-31592232011-09-06 Relative dosimetrical verification in high dose rate brachytherapy using two-dimensional detector array IMatriXX Manikandan, A. Biplab, Sarkar David, Perianayagam A. Holla, R. Vivek, T. R. Sujatha, N. J Med Phys Technical Note For high dose rate (HDR) brachytherapy, independent treatment verification is needed to ensure that the treatment is performed as per prescription. This study demonstrates dosimetric quality assurance of the HDR brachytherapy using a commercially available two-dimensional ion chamber array called IMatriXX, which has a detector separation of 0.7619 cm. The reference isodose length, step size, and source dwell positional accuracy were verified. A total of 24 dwell positions, which were verified for positional accuracy gave a total error (systematic and random) of –0.45 mm, with a standard deviation of 1.01 mm and maximum error of 1.8 mm. Using a step size of 5 mm, reference isodose length (the length of 100% isodose line) was verified for single and multiple catheters of same and different source loadings. An error ≤1 mm was measured in 57% of tests analyzed. Step size verification for 2, 3, 4, and 5 cm was performed and 70% of the step size errors were below 1 mm, with maximum of 1.2 mm. The step size ≤1 cm could not be verified by the IMatriXX as it could not resolve the peaks in dose profile. Medknow Publications Pvt Ltd 2011 /pmc/articles/PMC3159223/ /pubmed/21897562 http://dx.doi.org/10.4103/0971-6203.83491 Text en © Journal of Medical Physics http://creativecommons.org/licenses/by-nc-sa/3.0 This is an open-access article distributed under the terms of the Creative Commons Attribution-Noncommercial-Share Alike 3.0 Unported, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
spellingShingle Technical Note
Manikandan, A.
Biplab, Sarkar
David, Perianayagam A.
Holla, R.
Vivek, T. R.
Sujatha, N.
Relative dosimetrical verification in high dose rate brachytherapy using two-dimensional detector array IMatriXX
title Relative dosimetrical verification in high dose rate brachytherapy using two-dimensional detector array IMatriXX
title_full Relative dosimetrical verification in high dose rate brachytherapy using two-dimensional detector array IMatriXX
title_fullStr Relative dosimetrical verification in high dose rate brachytherapy using two-dimensional detector array IMatriXX
title_full_unstemmed Relative dosimetrical verification in high dose rate brachytherapy using two-dimensional detector array IMatriXX
title_short Relative dosimetrical verification in high dose rate brachytherapy using two-dimensional detector array IMatriXX
title_sort relative dosimetrical verification in high dose rate brachytherapy using two-dimensional detector array imatrixx
topic Technical Note
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3159223/
https://www.ncbi.nlm.nih.gov/pubmed/21897562
http://dx.doi.org/10.4103/0971-6203.83491
work_keys_str_mv AT manikandana relativedosimetricalverificationinhighdoseratebrachytherapyusingtwodimensionaldetectorarrayimatrixx
AT biplabsarkar relativedosimetricalverificationinhighdoseratebrachytherapyusingtwodimensionaldetectorarrayimatrixx
AT davidperianayagama relativedosimetricalverificationinhighdoseratebrachytherapyusingtwodimensionaldetectorarrayimatrixx
AT hollar relativedosimetricalverificationinhighdoseratebrachytherapyusingtwodimensionaldetectorarrayimatrixx
AT vivektr relativedosimetricalverificationinhighdoseratebrachytherapyusingtwodimensionaldetectorarrayimatrixx
AT sujathan relativedosimetricalverificationinhighdoseratebrachytherapyusingtwodimensionaldetectorarrayimatrixx