Dosimetric analysis of beam-matching procedure of two similar linear accelerators
Linear accelerators from the same vendor in a radiation therapy center are usually beam-matched following Vendor's acceptance criteria. This protocol is limited to check the difference at particular points on the ionization curve for depth dose or beam profiles. This article describes different...
Autores principales: | , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Medknow Publications Pvt Ltd
2011
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3159224/ https://www.ncbi.nlm.nih.gov/pubmed/21897563 http://dx.doi.org/10.4103/0971-6203.83497 |
_version_ | 1782210428289417216 |
---|---|
author | Bhangle, Janhavi R. Narayanan, V. K. Sathiya Kumar, Namitha K. Vaitheeswaran, R. |
author_facet | Bhangle, Janhavi R. Narayanan, V. K. Sathiya Kumar, Namitha K. Vaitheeswaran, R. |
author_sort | Bhangle, Janhavi R. |
collection | PubMed |
description | Linear accelerators from the same vendor in a radiation therapy center are usually beam-matched following Vendor's acceptance criteria. This protocol is limited to check the difference at particular points on the ionization curve for depth dose or beam profiles. This article describes different tests done after commissioning to evaluate the level of agreement between matched beams of two ONCOR Impression plus linear accelerators from Siemens. Total scatter factors, collimator scatter factors, wedge transmission factors were measured in water for 6-MV photon. All these factors for ONCOR2 were within ±1% of those values for ONCOR1. Along with these point dose measurements we have essentially used γ-index to compare the planar dose distribution from two beam-matched accelerators. For this study a set of ready packed EDR2 films was exposed on both accelerators. The set consisted of films for percentage depth dose, beam profiles, a pyramid shape, multileaf collimator's positional and dose delivery accuracy, and a film to compare head scatter at tray level. To include treatment planning system calculations, a film kept in axial plane was exposed to 3DCRT and IMRT plans with actual gantry angles and monitor units. These films were analyzed for γ in OmniPro IMRT software using different combinations of Δdose and - Δdistances. All these films have shown good agreement for - Δdistance of 3 mm and Δdose of 3 %. |
format | Online Article Text |
id | pubmed-3159224 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2011 |
publisher | Medknow Publications Pvt Ltd |
record_format | MEDLINE/PubMed |
spelling | pubmed-31592242011-09-06 Dosimetric analysis of beam-matching procedure of two similar linear accelerators Bhangle, Janhavi R. Narayanan, V. K. Sathiya Kumar, Namitha K. Vaitheeswaran, R. J Med Phys Technical Note Linear accelerators from the same vendor in a radiation therapy center are usually beam-matched following Vendor's acceptance criteria. This protocol is limited to check the difference at particular points on the ionization curve for depth dose or beam profiles. This article describes different tests done after commissioning to evaluate the level of agreement between matched beams of two ONCOR Impression plus linear accelerators from Siemens. Total scatter factors, collimator scatter factors, wedge transmission factors were measured in water for 6-MV photon. All these factors for ONCOR2 were within ±1% of those values for ONCOR1. Along with these point dose measurements we have essentially used γ-index to compare the planar dose distribution from two beam-matched accelerators. For this study a set of ready packed EDR2 films was exposed on both accelerators. The set consisted of films for percentage depth dose, beam profiles, a pyramid shape, multileaf collimator's positional and dose delivery accuracy, and a film to compare head scatter at tray level. To include treatment planning system calculations, a film kept in axial plane was exposed to 3DCRT and IMRT plans with actual gantry angles and monitor units. These films were analyzed for γ in OmniPro IMRT software using different combinations of Δdose and - Δdistances. All these films have shown good agreement for - Δdistance of 3 mm and Δdose of 3 %. Medknow Publications Pvt Ltd 2011 /pmc/articles/PMC3159224/ /pubmed/21897563 http://dx.doi.org/10.4103/0971-6203.83497 Text en © Journal of Medical Physics http://creativecommons.org/licenses/by-nc-sa/3.0 This is an open-access article distributed under the terms of the Creative Commons Attribution-Noncommercial-Share Alike 3.0 Unported, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited. |
spellingShingle | Technical Note Bhangle, Janhavi R. Narayanan, V. K. Sathiya Kumar, Namitha K. Vaitheeswaran, R. Dosimetric analysis of beam-matching procedure of two similar linear accelerators |
title | Dosimetric analysis of beam-matching procedure of two similar linear accelerators |
title_full | Dosimetric analysis of beam-matching procedure of two similar linear accelerators |
title_fullStr | Dosimetric analysis of beam-matching procedure of two similar linear accelerators |
title_full_unstemmed | Dosimetric analysis of beam-matching procedure of two similar linear accelerators |
title_short | Dosimetric analysis of beam-matching procedure of two similar linear accelerators |
title_sort | dosimetric analysis of beam-matching procedure of two similar linear accelerators |
topic | Technical Note |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3159224/ https://www.ncbi.nlm.nih.gov/pubmed/21897563 http://dx.doi.org/10.4103/0971-6203.83497 |
work_keys_str_mv | AT bhanglejanhavir dosimetricanalysisofbeammatchingprocedureoftwosimilarlinearaccelerators AT narayananvksathiya dosimetricanalysisofbeammatchingprocedureoftwosimilarlinearaccelerators AT kumarnamithak dosimetricanalysisofbeammatchingprocedureoftwosimilarlinearaccelerators AT vaitheeswaranr dosimetricanalysisofbeammatchingprocedureoftwosimilarlinearaccelerators |