An intensity based six-degree image registration algorithm between cone-beam CT (CBCT) and planning CT has been developed for image-guided radiation therapy (IGRT). CT images of an anthropomorphic chest phantom were acquired using conventional CT scanner and corresponding CBCT was reconstructed base...

Descripción completa

Detalles Bibliográficos
Autores principales: Yan, Hui, Lei, Ren, Wu, Jackie, Di, Fu, Yin, Fang-Fang
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Libertas Academica 2008
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3161701/
https://www.ncbi.nlm.nih.gov/pubmed/21892290
_version_ 1782210726013698048
author Yan, Hui
Lei, Ren
Wu, Jackie
Di, Fu
Yin, Fang-Fang
author_facet Yan, Hui
Lei, Ren
Wu, Jackie
Di, Fu
Yin, Fang-Fang
author_sort Yan, Hui
collection PubMed
description An intensity based six-degree image registration algorithm between cone-beam CT (CBCT) and planning CT has been developed for image-guided radiation therapy (IGRT). CT images of an anthropomorphic chest phantom were acquired using conventional CT scanner and corresponding CBCT was reconstructed based on projection images acquired by an on-board imager (OBI). Both sets of images were initially registered to each other using attached fudicial markers to achieve a golden standard registration. Starting from this point, an offset was applied to one set of images, and the matching result was found by a gray-value based registration method. Finally, The registration error was evaluated by comparing the detected shifts with the known shift. Three window-level (WL) combinations commonly used for image enhancement were examined to investigate the effect of anatomical information of Bony only (B), Bone+Tissue (BT), and Bone+Tissue+Air (BTA) on the accuracy and robustness of gray-value based registration algorithm. Extensive tests were performed in searching for the attraction range of registration algorithm. The widest attraction range was achieved with the WL combination of BTA. The average attraction ranges of this combination were 73.3 mm and 81.6 degree in the translation and rotation dimensions, respectively, and the average registration errors were 0.15 mm and 0.32 degree. The WL combination of BT shows the secondary largest attraction ranges. The WL combination of B shows limited convergence property and its attraction range was the smallest among the three examined combinations (on average 33.3 mm and 25.0 degree). If two sets of 3D images in original size (512 × 512) were used, registration could be accomplished within 10~20 minutes by current algorithm, which is only acceptable for off-line reviewing purpose. As the size of image set reduced by a factor of 2~4, the registration time would be 2~4 minutes which is feasible for on-line target localization.
format Online
Article
Text
id pubmed-3161701
institution National Center for Biotechnology Information
language English
publishDate 2008
publisher Libertas Academica
record_format MEDLINE/PubMed
spelling pubmed-31617012011-09-02 Evaluation of Image Enhancement Method on Target Registration Using Cone Beam CT in Radiation Therapy Yan, Hui Lei, Ren Wu, Jackie Di, Fu Yin, Fang-Fang Clin Med Oncol Technical Advance An intensity based six-degree image registration algorithm between cone-beam CT (CBCT) and planning CT has been developed for image-guided radiation therapy (IGRT). CT images of an anthropomorphic chest phantom were acquired using conventional CT scanner and corresponding CBCT was reconstructed based on projection images acquired by an on-board imager (OBI). Both sets of images were initially registered to each other using attached fudicial markers to achieve a golden standard registration. Starting from this point, an offset was applied to one set of images, and the matching result was found by a gray-value based registration method. Finally, The registration error was evaluated by comparing the detected shifts with the known shift. Three window-level (WL) combinations commonly used for image enhancement were examined to investigate the effect of anatomical information of Bony only (B), Bone+Tissue (BT), and Bone+Tissue+Air (BTA) on the accuracy and robustness of gray-value based registration algorithm. Extensive tests were performed in searching for the attraction range of registration algorithm. The widest attraction range was achieved with the WL combination of BTA. The average attraction ranges of this combination were 73.3 mm and 81.6 degree in the translation and rotation dimensions, respectively, and the average registration errors were 0.15 mm and 0.32 degree. The WL combination of BT shows the secondary largest attraction ranges. The WL combination of B shows limited convergence property and its attraction range was the smallest among the three examined combinations (on average 33.3 mm and 25.0 degree). If two sets of 3D images in original size (512 × 512) were used, registration could be accomplished within 10~20 minutes by current algorithm, which is only acceptable for off-line reviewing purpose. As the size of image set reduced by a factor of 2~4, the registration time would be 2~4 minutes which is feasible for on-line target localization. Libertas Academica 2008-03-28 /pmc/articles/PMC3161701/ /pubmed/21892290 Text en © 2008 the author(s), publisher and licensee Libertas Academica Ltd. http://creativecommons.org/licenses/by/3.0 This article is an open-access article distributed under the terms and conditions of the Creative Commons Attribution license http://creativecommons.org/licenses/by/3.0/).
spellingShingle Technical Advance
Yan, Hui
Lei, Ren
Wu, Jackie
Di, Fu
Yin, Fang-Fang
Evaluation of Image Enhancement Method on Target Registration Using Cone Beam CT in Radiation Therapy
title Evaluation of Image Enhancement Method on Target Registration Using Cone Beam CT in Radiation Therapy
title_full Evaluation of Image Enhancement Method on Target Registration Using Cone Beam CT in Radiation Therapy
title_fullStr Evaluation of Image Enhancement Method on Target Registration Using Cone Beam CT in Radiation Therapy
title_full_unstemmed Evaluation of Image Enhancement Method on Target Registration Using Cone Beam CT in Radiation Therapy
title_short Evaluation of Image Enhancement Method on Target Registration Using Cone Beam CT in Radiation Therapy
title_sort evaluation of image enhancement method on target registration using cone beam ct in radiation therapy
topic Technical Advance
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3161701/
https://www.ncbi.nlm.nih.gov/pubmed/21892290
work_keys_str_mv AT yanhui evaluationofimageenhancementmethodontargetregistrationusingconebeamctinradiationtherapy
AT leiren evaluationofimageenhancementmethodontargetregistrationusingconebeamctinradiationtherapy
AT wujackie evaluationofimageenhancementmethodontargetregistrationusingconebeamctinradiationtherapy
AT difu evaluationofimageenhancementmethodontargetregistrationusingconebeamctinradiationtherapy
AT yinfangfang evaluationofimageenhancementmethodontargetregistrationusingconebeamctinradiationtherapy