Comparison of Hanna and Hessburg-Barron trephine and punch systems using histological, anterior segment optical coherence tomography, and elliptical curve fitting models

BACKGROUND: This study analyzes the characteristics of donor and recipient tissue preparation between the Hessburg-Barron and Hanna punch and trephine systems by using elliptical curve fitting models, light microscopy, and anterior segment optical coherence tomography (AS-OCT). METHODS: Eight millim...

Descripción completa

Detalles Bibliográficos
Autores principales: Moshirfar, Majid, Calvo, Charles M, Kinard, Krista I, Williams, Lloyd B, Sikder, Shameema, Neuffer, Marcus C
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Dove Medical Press 2011
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3162291/
https://www.ncbi.nlm.nih.gov/pubmed/21887093
http://dx.doi.org/10.2147/OPTH.S23898
_version_ 1782210807229054976
author Moshirfar, Majid
Calvo, Charles M
Kinard, Krista I
Williams, Lloyd B
Sikder, Shameema
Neuffer, Marcus C
author_facet Moshirfar, Majid
Calvo, Charles M
Kinard, Krista I
Williams, Lloyd B
Sikder, Shameema
Neuffer, Marcus C
author_sort Moshirfar, Majid
collection PubMed
description BACKGROUND: This study analyzes the characteristics of donor and recipient tissue preparation between the Hessburg-Barron and Hanna punch and trephine systems by using elliptical curve fitting models, light microscopy, and anterior segment optical coherence tomography (AS-OCT). METHODS: Eight millimeter Hessburg-Barron and Hanna vacuum trephines and punches were used on six cadaver globes and six corneal-scleral rims, respectively. Eccentricity data were generated using measurements from photographs of the corneal buttons and were used to generate an elliptical curve fit to calculate properties of the corneal button. The trephination angle and punch angle were measured by digital protractor software from light microscopy and AS-OCT images to evaluate the consistency with which each device cuts the cornea. RESULTS: The Hanna trephine showed a trend towards producing a more circular recipient button than the Barron trephine (ratio of major axis to minor axis), ie, 1.059 ± 0.041 versus 1.110 ± 0.027 (P = 0.147) and the Hanna punch showed a trend towards producing a more circular donor cut than the Barron punch, ie, 1.021 ± 0.022 versus 1.046 ± 0.039 (P = 0.445). The Hanna trephine was demonstrated to have a more consistent trephination angle than the Barron trephine when assessing light microscopy images, ie, ±14.39° (95% confidence interval [CI] 111.9–157.7) versus ±19.38° (95% CI 101.9–150.2, P = 0.492) and OCT images, ie, ±8.08° (95% CI 106.2–123.3) versus ±11.16° (95% CI 109.3–132.6, P = 0.306). The angle created by the Hanna punch had less variability than the Barron punch from both the light microscopy, ie, ±4.81° (95% CI 101.6–113.9) versus ±11.28° (95% CI 84.5–120.6, P = 0.295) and AS-OCT imaging, ie, ±9.96° (95% CI 95.7–116.4) versus ±14.02° (95% CI 91.8–123.7, P = 0.825). Statistical significance was not achieved. CONCLUSION: The Hanna trephine and punch may be more accurate and consistent in cutting corneal buttons than the Hessburg-Barron trephine and punch when evaluated using elliptical curve fitting models, light microscopy, and AS-OCT.
format Online
Article
Text
id pubmed-3162291
institution National Center for Biotechnology Information
language English
publishDate 2011
publisher Dove Medical Press
record_format MEDLINE/PubMed
spelling pubmed-31622912011-09-01 Comparison of Hanna and Hessburg-Barron trephine and punch systems using histological, anterior segment optical coherence tomography, and elliptical curve fitting models Moshirfar, Majid Calvo, Charles M Kinard, Krista I Williams, Lloyd B Sikder, Shameema Neuffer, Marcus C Clin Ophthalmol Original Research BACKGROUND: This study analyzes the characteristics of donor and recipient tissue preparation between the Hessburg-Barron and Hanna punch and trephine systems by using elliptical curve fitting models, light microscopy, and anterior segment optical coherence tomography (AS-OCT). METHODS: Eight millimeter Hessburg-Barron and Hanna vacuum trephines and punches were used on six cadaver globes and six corneal-scleral rims, respectively. Eccentricity data were generated using measurements from photographs of the corneal buttons and were used to generate an elliptical curve fit to calculate properties of the corneal button. The trephination angle and punch angle were measured by digital protractor software from light microscopy and AS-OCT images to evaluate the consistency with which each device cuts the cornea. RESULTS: The Hanna trephine showed a trend towards producing a more circular recipient button than the Barron trephine (ratio of major axis to minor axis), ie, 1.059 ± 0.041 versus 1.110 ± 0.027 (P = 0.147) and the Hanna punch showed a trend towards producing a more circular donor cut than the Barron punch, ie, 1.021 ± 0.022 versus 1.046 ± 0.039 (P = 0.445). The Hanna trephine was demonstrated to have a more consistent trephination angle than the Barron trephine when assessing light microscopy images, ie, ±14.39° (95% confidence interval [CI] 111.9–157.7) versus ±19.38° (95% CI 101.9–150.2, P = 0.492) and OCT images, ie, ±8.08° (95% CI 106.2–123.3) versus ±11.16° (95% CI 109.3–132.6, P = 0.306). The angle created by the Hanna punch had less variability than the Barron punch from both the light microscopy, ie, ±4.81° (95% CI 101.6–113.9) versus ±11.28° (95% CI 84.5–120.6, P = 0.295) and AS-OCT imaging, ie, ±9.96° (95% CI 95.7–116.4) versus ±14.02° (95% CI 91.8–123.7, P = 0.825). Statistical significance was not achieved. CONCLUSION: The Hanna trephine and punch may be more accurate and consistent in cutting corneal buttons than the Hessburg-Barron trephine and punch when evaluated using elliptical curve fitting models, light microscopy, and AS-OCT. Dove Medical Press 2011 2011-08-12 /pmc/articles/PMC3162291/ /pubmed/21887093 http://dx.doi.org/10.2147/OPTH.S23898 Text en © 2011 Moshirfar et al, publisher and licensee Dove Medical Press Ltd. This is an Open Access article which permits unrestricted noncommercial use, provided the original work is properly cited.
spellingShingle Original Research
Moshirfar, Majid
Calvo, Charles M
Kinard, Krista I
Williams, Lloyd B
Sikder, Shameema
Neuffer, Marcus C
Comparison of Hanna and Hessburg-Barron trephine and punch systems using histological, anterior segment optical coherence tomography, and elliptical curve fitting models
title Comparison of Hanna and Hessburg-Barron trephine and punch systems using histological, anterior segment optical coherence tomography, and elliptical curve fitting models
title_full Comparison of Hanna and Hessburg-Barron trephine and punch systems using histological, anterior segment optical coherence tomography, and elliptical curve fitting models
title_fullStr Comparison of Hanna and Hessburg-Barron trephine and punch systems using histological, anterior segment optical coherence tomography, and elliptical curve fitting models
title_full_unstemmed Comparison of Hanna and Hessburg-Barron trephine and punch systems using histological, anterior segment optical coherence tomography, and elliptical curve fitting models
title_short Comparison of Hanna and Hessburg-Barron trephine and punch systems using histological, anterior segment optical coherence tomography, and elliptical curve fitting models
title_sort comparison of hanna and hessburg-barron trephine and punch systems using histological, anterior segment optical coherence tomography, and elliptical curve fitting models
topic Original Research
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3162291/
https://www.ncbi.nlm.nih.gov/pubmed/21887093
http://dx.doi.org/10.2147/OPTH.S23898
work_keys_str_mv AT moshirfarmajid comparisonofhannaandhessburgbarrontrephineandpunchsystemsusinghistologicalanteriorsegmentopticalcoherencetomographyandellipticalcurvefittingmodels
AT calvocharlesm comparisonofhannaandhessburgbarrontrephineandpunchsystemsusinghistologicalanteriorsegmentopticalcoherencetomographyandellipticalcurvefittingmodels
AT kinardkristai comparisonofhannaandhessburgbarrontrephineandpunchsystemsusinghistologicalanteriorsegmentopticalcoherencetomographyandellipticalcurvefittingmodels
AT williamslloydb comparisonofhannaandhessburgbarrontrephineandpunchsystemsusinghistologicalanteriorsegmentopticalcoherencetomographyandellipticalcurvefittingmodels
AT sikdershameema comparisonofhannaandhessburgbarrontrephineandpunchsystemsusinghistologicalanteriorsegmentopticalcoherencetomographyandellipticalcurvefittingmodels
AT neuffermarcusc comparisonofhannaandhessburgbarrontrephineandpunchsystemsusinghistologicalanteriorsegmentopticalcoherencetomographyandellipticalcurvefittingmodels