Cargando…
The accuracy of dynamic predictive autofocusing for whole slide imaging
CONTEXT: Whole slide imaging (WSI) for digital pathology involves the rapid automated acquisition of multiple high-power fields from a microscope slide containing a tissue specimen. Capturing each field in the correct focal plane is essential to create high-quality digital images. Others have descri...
Autores principales: | , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Medknow Publications Pvt Ltd
2011
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3169924/ https://www.ncbi.nlm.nih.gov/pubmed/21969919 http://dx.doi.org/10.4103/2153-3539.84231 |
_version_ | 1782211549359767552 |
---|---|
author | McKay, Richard R. Baxi, Vipul A. Montalto, Michael C. |
author_facet | McKay, Richard R. Baxi, Vipul A. Montalto, Michael C. |
author_sort | McKay, Richard R. |
collection | PubMed |
description | CONTEXT: Whole slide imaging (WSI) for digital pathology involves the rapid automated acquisition of multiple high-power fields from a microscope slide containing a tissue specimen. Capturing each field in the correct focal plane is essential to create high-quality digital images. Others have described a novel focusing method which reduces the number of focal planes required to generate accurate focus. However, this method was not applied dynamically in an automated WSI system under continuous motion. AIMS: This report measures the accuracy of this method when applied in a rapid continuous scan mode using a dual sensor WSI system with interleaved acquisition of images. METHODS: We acquired over 400 tiles in a “stop and go” scan mode, surveying the entire z depth in each tile and used this as ground truth. We compared this ground truth focal height to the focal height determined using a rapid 3-point focus algorithm applied dynamically in a continuous scanning mode. RESULTS: Our data showed the average focal height error of 0.30 (±0.27) μm compared to ground truth, which is well within the system's depth of field. On a tile by tile assessment, approximately 95% of the tiles were within the system's depth of field. Further, this method was six times faster than acquiring tiles compared to the same method in a non-continuous scan mode. CONCLUSIONS: The data indicates that the method employed can yield a significant improvement in scan speed while maintaining highly accurate autofocusing. |
format | Online Article Text |
id | pubmed-3169924 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2011 |
publisher | Medknow Publications Pvt Ltd |
record_format | MEDLINE/PubMed |
spelling | pubmed-31699242011-10-03 The accuracy of dynamic predictive autofocusing for whole slide imaging McKay, Richard R. Baxi, Vipul A. Montalto, Michael C. J Pathol Inform Research Article CONTEXT: Whole slide imaging (WSI) for digital pathology involves the rapid automated acquisition of multiple high-power fields from a microscope slide containing a tissue specimen. Capturing each field in the correct focal plane is essential to create high-quality digital images. Others have described a novel focusing method which reduces the number of focal planes required to generate accurate focus. However, this method was not applied dynamically in an automated WSI system under continuous motion. AIMS: This report measures the accuracy of this method when applied in a rapid continuous scan mode using a dual sensor WSI system with interleaved acquisition of images. METHODS: We acquired over 400 tiles in a “stop and go” scan mode, surveying the entire z depth in each tile and used this as ground truth. We compared this ground truth focal height to the focal height determined using a rapid 3-point focus algorithm applied dynamically in a continuous scanning mode. RESULTS: Our data showed the average focal height error of 0.30 (±0.27) μm compared to ground truth, which is well within the system's depth of field. On a tile by tile assessment, approximately 95% of the tiles were within the system's depth of field. Further, this method was six times faster than acquiring tiles compared to the same method in a non-continuous scan mode. CONCLUSIONS: The data indicates that the method employed can yield a significant improvement in scan speed while maintaining highly accurate autofocusing. Medknow Publications Pvt Ltd 2011-08-24 /pmc/articles/PMC3169924/ /pubmed/21969919 http://dx.doi.org/10.4103/2153-3539.84231 Text en Copyright: © 2011 McKay RR http://creativecommons.org/licenses/by-nc-sa/3.0 This is an open-access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited. |
spellingShingle | Research Article McKay, Richard R. Baxi, Vipul A. Montalto, Michael C. The accuracy of dynamic predictive autofocusing for whole slide imaging |
title | The accuracy of dynamic predictive autofocusing for whole slide imaging |
title_full | The accuracy of dynamic predictive autofocusing for whole slide imaging |
title_fullStr | The accuracy of dynamic predictive autofocusing for whole slide imaging |
title_full_unstemmed | The accuracy of dynamic predictive autofocusing for whole slide imaging |
title_short | The accuracy of dynamic predictive autofocusing for whole slide imaging |
title_sort | accuracy of dynamic predictive autofocusing for whole slide imaging |
topic | Research Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3169924/ https://www.ncbi.nlm.nih.gov/pubmed/21969919 http://dx.doi.org/10.4103/2153-3539.84231 |
work_keys_str_mv | AT mckayrichardr theaccuracyofdynamicpredictiveautofocusingforwholeslideimaging AT baxivipula theaccuracyofdynamicpredictiveautofocusingforwholeslideimaging AT montaltomichaelc theaccuracyofdynamicpredictiveautofocusingforwholeslideimaging AT mckayrichardr accuracyofdynamicpredictiveautofocusingforwholeslideimaging AT baxivipula accuracyofdynamicpredictiveautofocusingforwholeslideimaging AT montaltomichaelc accuracyofdynamicpredictiveautofocusingforwholeslideimaging |