The baseline pressure of intracranial pressure (ICP) sensors can be altered by electrostatic discharges

BACKGROUND: The monitoring of intracranial pressure (ICP) has a crucial role in the surveillance of patients with brain injury. During long-term monitoring of ICP, we have seen spontaneous shifts in baseline pressure (ICP sensor zero point), which are of technical and not physiological origin. The a...

Descripción completa

Detalles Bibliográficos
Autores principales: Eide, Per K, Bakken, André
Formato: Online Artículo Texto
Lenguaje:English
Publicado: BioMed Central 2011
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3180435/
https://www.ncbi.nlm.nih.gov/pubmed/21859487
http://dx.doi.org/10.1186/1475-925X-10-75
_version_ 1782212636081913856
author Eide, Per K
Bakken, André
author_facet Eide, Per K
Bakken, André
author_sort Eide, Per K
collection PubMed
description BACKGROUND: The monitoring of intracranial pressure (ICP) has a crucial role in the surveillance of patients with brain injury. During long-term monitoring of ICP, we have seen spontaneous shifts in baseline pressure (ICP sensor zero point), which are of technical and not physiological origin. The aim of the present study was to explore whether or not baseline pressures of ICP sensors can be affected by electrostatics discharges (ESD's), when ESD's are delivered at clinically relevant magnitudes. METHODS: We performed bench-testing of a set of commercial ICP sensors. In our experimental setup, the ICP sensor was placed in a container with 0.9% NaCl solution. A test person was charged 0.5 - 10 kV, and then delivered ESD's to the sensor by touching a metal rod that was located in the container. The continuous pressure signals were recorded continuously before/after the ESD's, and the pressure readings were stored digitally using a computerized system RESULTS: A total of 57 sensors were tested, including 25 Codman ICP sensors and 32 Raumedic sensors. When charging the test person in the range 0.5-10 kV, typically ESD's in the range 0.5 - 5 kV peak pulse were delivered to the ICP sensor. Alterations in baseline pressure ≥ 2 mmHg was seen in 24 of 25 (96%) Codman sensors and in 17 of 32 (53%) Raumedic sensors. Lasting changes in baseline pressure > 10 mmHg that in the clinical setting would affect patient management, were seen frequently for both sensor types. The changes in baseline pressure were either characterized by sudden shifts or gradual drifts in baseline pressure. CONCLUSIONS: The baseline pressures of commercial solid ICP sensors can be altered by ESD's at discharge magnitudes that are clinically relevant. Shifts in baseline pressure change the ICP levels visualised to the physician on the monitor screen, and thereby reveal wrong ICP values, which likely represent a severe risk to the patient.
format Online
Article
Text
id pubmed-3180435
institution National Center for Biotechnology Information
language English
publishDate 2011
publisher BioMed Central
record_format MEDLINE/PubMed
spelling pubmed-31804352011-09-27 The baseline pressure of intracranial pressure (ICP) sensors can be altered by electrostatic discharges Eide, Per K Bakken, André Biomed Eng Online Research BACKGROUND: The monitoring of intracranial pressure (ICP) has a crucial role in the surveillance of patients with brain injury. During long-term monitoring of ICP, we have seen spontaneous shifts in baseline pressure (ICP sensor zero point), which are of technical and not physiological origin. The aim of the present study was to explore whether or not baseline pressures of ICP sensors can be affected by electrostatics discharges (ESD's), when ESD's are delivered at clinically relevant magnitudes. METHODS: We performed bench-testing of a set of commercial ICP sensors. In our experimental setup, the ICP sensor was placed in a container with 0.9% NaCl solution. A test person was charged 0.5 - 10 kV, and then delivered ESD's to the sensor by touching a metal rod that was located in the container. The continuous pressure signals were recorded continuously before/after the ESD's, and the pressure readings were stored digitally using a computerized system RESULTS: A total of 57 sensors were tested, including 25 Codman ICP sensors and 32 Raumedic sensors. When charging the test person in the range 0.5-10 kV, typically ESD's in the range 0.5 - 5 kV peak pulse were delivered to the ICP sensor. Alterations in baseline pressure ≥ 2 mmHg was seen in 24 of 25 (96%) Codman sensors and in 17 of 32 (53%) Raumedic sensors. Lasting changes in baseline pressure > 10 mmHg that in the clinical setting would affect patient management, were seen frequently for both sensor types. The changes in baseline pressure were either characterized by sudden shifts or gradual drifts in baseline pressure. CONCLUSIONS: The baseline pressures of commercial solid ICP sensors can be altered by ESD's at discharge magnitudes that are clinically relevant. Shifts in baseline pressure change the ICP levels visualised to the physician on the monitor screen, and thereby reveal wrong ICP values, which likely represent a severe risk to the patient. BioMed Central 2011-08-22 /pmc/articles/PMC3180435/ /pubmed/21859487 http://dx.doi.org/10.1186/1475-925X-10-75 Text en Copyright ©2011 Eide and Bakken; licensee BioMed Central Ltd. http://creativecommons.org/licenses/by/2.0 This is an Open Access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/2.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
spellingShingle Research
Eide, Per K
Bakken, André
The baseline pressure of intracranial pressure (ICP) sensors can be altered by electrostatic discharges
title The baseline pressure of intracranial pressure (ICP) sensors can be altered by electrostatic discharges
title_full The baseline pressure of intracranial pressure (ICP) sensors can be altered by electrostatic discharges
title_fullStr The baseline pressure of intracranial pressure (ICP) sensors can be altered by electrostatic discharges
title_full_unstemmed The baseline pressure of intracranial pressure (ICP) sensors can be altered by electrostatic discharges
title_short The baseline pressure of intracranial pressure (ICP) sensors can be altered by electrostatic discharges
title_sort baseline pressure of intracranial pressure (icp) sensors can be altered by electrostatic discharges
topic Research
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3180435/
https://www.ncbi.nlm.nih.gov/pubmed/21859487
http://dx.doi.org/10.1186/1475-925X-10-75
work_keys_str_mv AT eideperk thebaselinepressureofintracranialpressureicpsensorscanbealteredbyelectrostaticdischarges
AT bakkenandre thebaselinepressureofintracranialpressureicpsensorscanbealteredbyelectrostaticdischarges
AT eideperk baselinepressureofintracranialpressureicpsensorscanbealteredbyelectrostaticdischarges
AT bakkenandre baselinepressureofintracranialpressureicpsensorscanbealteredbyelectrostaticdischarges