Cargando…
The baseline pressure of intracranial pressure (ICP) sensors can be altered by electrostatic discharges
BACKGROUND: The monitoring of intracranial pressure (ICP) has a crucial role in the surveillance of patients with brain injury. During long-term monitoring of ICP, we have seen spontaneous shifts in baseline pressure (ICP sensor zero point), which are of technical and not physiological origin. The a...
Autores principales: | Eide, Per K, Bakken, André |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
BioMed Central
2011
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3180435/ https://www.ncbi.nlm.nih.gov/pubmed/21859487 http://dx.doi.org/10.1186/1475-925X-10-75 |
Ejemplares similares
-
The effect of baseline pressure errors on an intracranial pressure-derived index: results of a prospective observational study
por: Eide, Per Kristian, et al.
Publicado: (2014) -
Baseline pressure errors (BPEs) extensively influence intracranial pressure scores: results of a prospective observational study
por: Eide, Per Kristian, et al.
Publicado: (2014) -
Abnormal Intracranial Pulse Pressure Amplitude Despite Normalized Static Intracranial Pressure in Idiopathic Intracranial Hypertension Refractory to Conservative Medical Therapy
por: Eide, Per Kristian
Publicado: (2021) -
Cardiac output in idiopathic normal pressure hydrocephalus: association with arterial blood pressure and intracranial pressure wave amplitudes and outcome of shunt surgery
por: Eide, Per K
Publicado: (2011) -
Correlation of transcranial doppler (TCD) parameters with intracranial pressure (ICP)
por: Vakalos, A, et al.
Publicado: (1999)