Investigations on antibody binding to a micro-cantilever coated with a BAM pesticide residue

The attachment of an antibody to an antigen-coated cantilever has been investigated by repeated experiments, using a cantilever-based detection system by Cantion A/S. The stress induced by the binding of a pesticide residue BAM (2,6 dichlorobenzamide) immobilized on a cantilever surface to anti-BAM...

Descripción completa

Detalles Bibliográficos
Autores principales: Bache, Michael, Taboryski, Rafael, Schmid, Silvan, Aamand, Jens, Jakobsen, Mogens Havsteen
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Springer 2011
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3211479/
https://www.ncbi.nlm.nih.gov/pubmed/21711907
http://dx.doi.org/10.1186/1556-276X-6-386
Descripción
Sumario:The attachment of an antibody to an antigen-coated cantilever has been investigated by repeated experiments, using a cantilever-based detection system by Cantion A/S. The stress induced by the binding of a pesticide residue BAM (2,6 dichlorobenzamide) immobilized on a cantilever surface to anti-BAM antibody is measured using the CantiLab4(© )system from Cantion A/S with four gold-coated cantilevers and piezo resistive readout. The detection mechanism is in principle label-free, but fluorescent-marked antibodies have been used to subsequently verify the binding on the cantilever surface. The bending and increase in mass of each cantilever has also been investigated using a light interferometer and a Doppler Vibrometer. The system has been analyzed during repeated measurements to investigate whether the CantiLab4(© )system is a suited platform for a pesticide assay system.