Cargando…

Quantification and visualization of cardiovascular 4D velocity mapping accelerated with parallel imaging or k-t BLAST: head to head comparison and validation at 1.5 T and 3 T

BACKGROUND: Three-dimensional time-resolved (4D) phase-contrast (PC) CMR can visualize and quantify cardiovascular flow but is hampered by long acquisition times. Acceleration with SENSE or k-t BLAST are two possibilities but results on validation are lacking, especially at 3 T. The aim of this stud...

Descripción completa

Detalles Bibliográficos
Autores principales: Carlsson, Marcus, Töger, Johannes, Kanski, Mikael, Bloch, Karin Markenroth, Ståhlberg, Freddy, Heiberg, Einar, Arheden, Håkan
Formato: Online Artículo Texto
Lenguaje:English
Publicado: BioMed Central 2011
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3213199/
https://www.ncbi.nlm.nih.gov/pubmed/21970399
http://dx.doi.org/10.1186/1532-429X-13-55
_version_ 1782216101077188608
author Carlsson, Marcus
Töger, Johannes
Kanski, Mikael
Bloch, Karin Markenroth
Ståhlberg, Freddy
Heiberg, Einar
Arheden, Håkan
author_facet Carlsson, Marcus
Töger, Johannes
Kanski, Mikael
Bloch, Karin Markenroth
Ståhlberg, Freddy
Heiberg, Einar
Arheden, Håkan
author_sort Carlsson, Marcus
collection PubMed
description BACKGROUND: Three-dimensional time-resolved (4D) phase-contrast (PC) CMR can visualize and quantify cardiovascular flow but is hampered by long acquisition times. Acceleration with SENSE or k-t BLAST are two possibilities but results on validation are lacking, especially at 3 T. The aim of this study was therefore to validate quantitative in vivo cardiac 4D-acquisitions accelerated with parallel imaging and k-t BLAST at 1.5 T and 3 T with 2D-flow as the reference and to investigate if field strengths and type of acceleration have major effects on intracardiac flow visualization. METHODS: The local ethical committee approved the study. 13 healthy volunteers were scanned at both 1.5 T and 3 T in random order with 2D-flow of the aorta and main pulmonary artery and two 4D-flow sequences of the heart accelerated with SENSE and k-t BLAST respectively. 2D-image planes were reconstructed at the aortic and pulmonary outflow. Flow curves were calculated and peak flows and stroke volumes (SV) compared to the results from 2D-flow acquisitions. Intra-cardiac flow was visualized using particle tracing and image quality based on the flow patterns of the particles was graded using a four-point scale. RESULTS: Good accuracy of SV quantification was found using 3 T 4D-SENSE (r(2 )= 0.86, -0.7 ± 7.6%) and although a larger bias was found on 1.5 T (r(2 )= 0.71, -3.6 ± 14.8%), the difference was not significant (p = 0.46). Accuracy of 4D k-t BLAST for SV was lower (p < 0.01) on 1.5 T (r(2 )= 0.65, -15.6 ± 13.7%) compared to 3 T (r(2 )= 0.64, -4.6 ± 10.0%). Peak flow was lower with 4D-SENSE at both 3 T and 1.5 T compared to 2D-flow (p < 0.01) and even lower with 4D k-t BLAST at both scanners (p < 0.01). Intracardiac flow visualization did not differ between 1.5 T and 3 T (p = 0.09) or between 4D-SENSE or 4D k-t BLAST (p = 0.85). CONCLUSIONS: The present study showed that quantitative 4D flow accelerated with SENSE has good accuracy at 3 T and compares favourably to 1.5 T. 4D flow accelerated with k-t BLAST underestimate flow velocities and thereby yield too high bias for intra-cardiac quantitative in vivo use at the present time. For intra-cardiac 4D-flow visualization, however, 1.5 T and 3 T as well as SENSE or k-t BLAST can be used with similar quality.
format Online
Article
Text
id pubmed-3213199
institution National Center for Biotechnology Information
language English
publishDate 2011
publisher BioMed Central
record_format MEDLINE/PubMed
spelling pubmed-32131992011-11-11 Quantification and visualization of cardiovascular 4D velocity mapping accelerated with parallel imaging or k-t BLAST: head to head comparison and validation at 1.5 T and 3 T Carlsson, Marcus Töger, Johannes Kanski, Mikael Bloch, Karin Markenroth Ståhlberg, Freddy Heiberg, Einar Arheden, Håkan J Cardiovasc Magn Reson Research BACKGROUND: Three-dimensional time-resolved (4D) phase-contrast (PC) CMR can visualize and quantify cardiovascular flow but is hampered by long acquisition times. Acceleration with SENSE or k-t BLAST are two possibilities but results on validation are lacking, especially at 3 T. The aim of this study was therefore to validate quantitative in vivo cardiac 4D-acquisitions accelerated with parallel imaging and k-t BLAST at 1.5 T and 3 T with 2D-flow as the reference and to investigate if field strengths and type of acceleration have major effects on intracardiac flow visualization. METHODS: The local ethical committee approved the study. 13 healthy volunteers were scanned at both 1.5 T and 3 T in random order with 2D-flow of the aorta and main pulmonary artery and two 4D-flow sequences of the heart accelerated with SENSE and k-t BLAST respectively. 2D-image planes were reconstructed at the aortic and pulmonary outflow. Flow curves were calculated and peak flows and stroke volumes (SV) compared to the results from 2D-flow acquisitions. Intra-cardiac flow was visualized using particle tracing and image quality based on the flow patterns of the particles was graded using a four-point scale. RESULTS: Good accuracy of SV quantification was found using 3 T 4D-SENSE (r(2 )= 0.86, -0.7 ± 7.6%) and although a larger bias was found on 1.5 T (r(2 )= 0.71, -3.6 ± 14.8%), the difference was not significant (p = 0.46). Accuracy of 4D k-t BLAST for SV was lower (p < 0.01) on 1.5 T (r(2 )= 0.65, -15.6 ± 13.7%) compared to 3 T (r(2 )= 0.64, -4.6 ± 10.0%). Peak flow was lower with 4D-SENSE at both 3 T and 1.5 T compared to 2D-flow (p < 0.01) and even lower with 4D k-t BLAST at both scanners (p < 0.01). Intracardiac flow visualization did not differ between 1.5 T and 3 T (p = 0.09) or between 4D-SENSE or 4D k-t BLAST (p = 0.85). CONCLUSIONS: The present study showed that quantitative 4D flow accelerated with SENSE has good accuracy at 3 T and compares favourably to 1.5 T. 4D flow accelerated with k-t BLAST underestimate flow velocities and thereby yield too high bias for intra-cardiac quantitative in vivo use at the present time. For intra-cardiac 4D-flow visualization, however, 1.5 T and 3 T as well as SENSE or k-t BLAST can be used with similar quality. BioMed Central 2011-10-04 /pmc/articles/PMC3213199/ /pubmed/21970399 http://dx.doi.org/10.1186/1532-429X-13-55 Text en Copyright ©2011 Carlsson et al; licensee BioMed Central Ltd. http://creativecommons.org/licenses/by/2.0 This is an Open Access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/2.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
spellingShingle Research
Carlsson, Marcus
Töger, Johannes
Kanski, Mikael
Bloch, Karin Markenroth
Ståhlberg, Freddy
Heiberg, Einar
Arheden, Håkan
Quantification and visualization of cardiovascular 4D velocity mapping accelerated with parallel imaging or k-t BLAST: head to head comparison and validation at 1.5 T and 3 T
title Quantification and visualization of cardiovascular 4D velocity mapping accelerated with parallel imaging or k-t BLAST: head to head comparison and validation at 1.5 T and 3 T
title_full Quantification and visualization of cardiovascular 4D velocity mapping accelerated with parallel imaging or k-t BLAST: head to head comparison and validation at 1.5 T and 3 T
title_fullStr Quantification and visualization of cardiovascular 4D velocity mapping accelerated with parallel imaging or k-t BLAST: head to head comparison and validation at 1.5 T and 3 T
title_full_unstemmed Quantification and visualization of cardiovascular 4D velocity mapping accelerated with parallel imaging or k-t BLAST: head to head comparison and validation at 1.5 T and 3 T
title_short Quantification and visualization of cardiovascular 4D velocity mapping accelerated with parallel imaging or k-t BLAST: head to head comparison and validation at 1.5 T and 3 T
title_sort quantification and visualization of cardiovascular 4d velocity mapping accelerated with parallel imaging or k-t blast: head to head comparison and validation at 1.5 t and 3 t
topic Research
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3213199/
https://www.ncbi.nlm.nih.gov/pubmed/21970399
http://dx.doi.org/10.1186/1532-429X-13-55
work_keys_str_mv AT carlssonmarcus quantificationandvisualizationofcardiovascular4dvelocitymappingacceleratedwithparallelimagingorktblastheadtoheadcomparisonandvalidationat15tand3t
AT togerjohannes quantificationandvisualizationofcardiovascular4dvelocitymappingacceleratedwithparallelimagingorktblastheadtoheadcomparisonandvalidationat15tand3t
AT kanskimikael quantificationandvisualizationofcardiovascular4dvelocitymappingacceleratedwithparallelimagingorktblastheadtoheadcomparisonandvalidationat15tand3t
AT blochkarinmarkenroth quantificationandvisualizationofcardiovascular4dvelocitymappingacceleratedwithparallelimagingorktblastheadtoheadcomparisonandvalidationat15tand3t
AT stahlbergfreddy quantificationandvisualizationofcardiovascular4dvelocitymappingacceleratedwithparallelimagingorktblastheadtoheadcomparisonandvalidationat15tand3t
AT heibergeinar quantificationandvisualizationofcardiovascular4dvelocitymappingacceleratedwithparallelimagingorktblastheadtoheadcomparisonandvalidationat15tand3t
AT arhedenhakan quantificationandvisualizationofcardiovascular4dvelocitymappingacceleratedwithparallelimagingorktblastheadtoheadcomparisonandvalidationat15tand3t