Electromagnetic interference in a cardiac pacemaker during cauterization with the coagulating, not cutting mode

Electromagnetic interference in pacemakers has almost always been reported in association with the cutting mode of monopolar electrocautery and rarely in association with the coagulation mode. We report a case of electrocautery-induced electromagnetic interference with a DDDR pacemaker (dual-chamber...

Descripción completa

Detalles Bibliográficos
Autores principales: Abdelmalak, Basem, Jagannathan, Narasimhan, Arain, Faisal D, Cymbor, Susan, McLain, Robert, Tetzlaff, John E
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Medknow Publications Pvt Ltd 2011
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3214561/
https://www.ncbi.nlm.nih.gov/pubmed/22096289
http://dx.doi.org/10.4103/0970-9185.86600
_version_ 1782216292449648640
author Abdelmalak, Basem
Jagannathan, Narasimhan
Arain, Faisal D
Cymbor, Susan
McLain, Robert
Tetzlaff, John E
author_facet Abdelmalak, Basem
Jagannathan, Narasimhan
Arain, Faisal D
Cymbor, Susan
McLain, Robert
Tetzlaff, John E
author_sort Abdelmalak, Basem
collection PubMed
description Electromagnetic interference in pacemakers has almost always been reported in association with the cutting mode of monopolar electrocautery and rarely in association with the coagulation mode. We report a case of electrocautery-induced electromagnetic interference with a DDDR pacemaker (dual-chamber paced, dual-chamber sensed, dual response to sensing, and rate modulated) in the coagulating and not cutting mode during a spine procedure. We also discuss the factors affecting intraoperative electromagnetic interference. A 74-year-old man experienced intraoperative electromagnetic interference that resulted in asystole caused by surgical electrocautery in the coagulation mode while the electrodispersive pad was placed at different locations and distances from the operating site (This electromagnetic interference did not occur during the use of the cutting mode). However, because of careful management, the outcome was favorable. Clinicians should be aware that the coagulation mode of electrocautery can cause electromagnetic interference and hemodynamic instability. Heightened vigilance and preparedness can ensure a favorable outcome.
format Online
Article
Text
id pubmed-3214561
institution National Center for Biotechnology Information
language English
publishDate 2011
publisher Medknow Publications Pvt Ltd
record_format MEDLINE/PubMed
spelling pubmed-32145612011-11-17 Electromagnetic interference in a cardiac pacemaker during cauterization with the coagulating, not cutting mode Abdelmalak, Basem Jagannathan, Narasimhan Arain, Faisal D Cymbor, Susan McLain, Robert Tetzlaff, John E J Anaesthesiol Clin Pharmacol Case Report Electromagnetic interference in pacemakers has almost always been reported in association with the cutting mode of monopolar electrocautery and rarely in association with the coagulation mode. We report a case of electrocautery-induced electromagnetic interference with a DDDR pacemaker (dual-chamber paced, dual-chamber sensed, dual response to sensing, and rate modulated) in the coagulating and not cutting mode during a spine procedure. We also discuss the factors affecting intraoperative electromagnetic interference. A 74-year-old man experienced intraoperative electromagnetic interference that resulted in asystole caused by surgical electrocautery in the coagulation mode while the electrodispersive pad was placed at different locations and distances from the operating site (This electromagnetic interference did not occur during the use of the cutting mode). However, because of careful management, the outcome was favorable. Clinicians should be aware that the coagulation mode of electrocautery can cause electromagnetic interference and hemodynamic instability. Heightened vigilance and preparedness can ensure a favorable outcome. Medknow Publications Pvt Ltd 2011 /pmc/articles/PMC3214561/ /pubmed/22096289 http://dx.doi.org/10.4103/0970-9185.86600 Text en Copyright: © Journal of Anaesthesiology Clinical Pharmacology http://creativecommons.org/licenses/by-nc-sa/3.0 This is an open-access article distributed under the terms of the Creative Commons Attribution-Noncommercial-Share Alike 3.0 Unported, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
spellingShingle Case Report
Abdelmalak, Basem
Jagannathan, Narasimhan
Arain, Faisal D
Cymbor, Susan
McLain, Robert
Tetzlaff, John E
Electromagnetic interference in a cardiac pacemaker during cauterization with the coagulating, not cutting mode
title Electromagnetic interference in a cardiac pacemaker during cauterization with the coagulating, not cutting mode
title_full Electromagnetic interference in a cardiac pacemaker during cauterization with the coagulating, not cutting mode
title_fullStr Electromagnetic interference in a cardiac pacemaker during cauterization with the coagulating, not cutting mode
title_full_unstemmed Electromagnetic interference in a cardiac pacemaker during cauterization with the coagulating, not cutting mode
title_short Electromagnetic interference in a cardiac pacemaker during cauterization with the coagulating, not cutting mode
title_sort electromagnetic interference in a cardiac pacemaker during cauterization with the coagulating, not cutting mode
topic Case Report
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3214561/
https://www.ncbi.nlm.nih.gov/pubmed/22096289
http://dx.doi.org/10.4103/0970-9185.86600
work_keys_str_mv AT abdelmalakbasem electromagneticinterferenceinacardiacpacemakerduringcauterizationwiththecoagulatingnotcuttingmode
AT jagannathannarasimhan electromagneticinterferenceinacardiacpacemakerduringcauterizationwiththecoagulatingnotcuttingmode
AT arainfaisald electromagneticinterferenceinacardiacpacemakerduringcauterizationwiththecoagulatingnotcuttingmode
AT cymborsusan electromagneticinterferenceinacardiacpacemakerduringcauterizationwiththecoagulatingnotcuttingmode
AT mclainrobert electromagneticinterferenceinacardiacpacemakerduringcauterizationwiththecoagulatingnotcuttingmode
AT tetzlaffjohne electromagneticinterferenceinacardiacpacemakerduringcauterizationwiththecoagulatingnotcuttingmode